Chapter 3
Signal Connections
Floating
V
s
Signal
Make this connection to
ground reference the signal.
Cold-Junction Sensor Connection
Warning: Exceeding the overvoltage protection on the CJSENSOR input can damage
!
VXI-SC-1102 User Manual
CH0 +
CH0 -
+
-
CH1 +
CH1 -
CGND
Reference to
Chassis Ground
Figure 3-3. Floating Signal Connection Referenced to Chassis Ground
Pins A3 and A4 (CJSENSOR) connect the temperature sensor located
on the National Instruments VXI-TB Series terminal block to the
VXI-SC-1102. The CJSENSOR signal is measured relative to CGND.
Pins A3 and A4 are connected within the VXI-SC-1102 so the position
of the MTEMP/DTEMP switch on the VXI-TB Series terminal block
does not matter. The input is overvoltage-protected to 15 VDC with the
power on and off.