Current-Loop Receivers; Figure 2-1. Vxi-Sc-1102 Parts Locator Diagram; Figure 2-2. Bent And Trimmed Resistor - DAQ VXI-SC-1000 User Manual

32-channel multiplexer amplifier module
Table of Contents

Advertisement

Current-Loop Receivers

Warning: Before installing the resistors in your submodule, make sure that there are
© National Instruments Corporation

Figure 2-1. VXI-SC-1102 Parts Locator Diagram

The VXI-SC-1102 has pads for transforming individual channels to
current-to-voltage converters. National Instruments offers an SCXI
Process Current Resistor Kit of four 249 Ω, 0.1%, 5 ppm, 1/4 W
resistors. The reference designators for the current loop resistors have
the following format: for input channel x, the resistor is RCLx. For
example, the resistor for channel 28 is RCL28.
no signals connected to your submodule front connector.
Install the resistors by performing the following steps:
1. Follow the instructions in the VXI-SC-1000 Carrier Module
Installation Guide to remove the VXI-SC-1102 submodule from
the carrier assembly.
2. Bend and trim the resistor lead as shown in Figure 2-2. Be sure that
the resistor does not extend more than 0.65 in. above the surface of
the circuit board.
3. Insert the resistor into the appropriate pad, labeled RCLx.
4. Solder the leads to the pad on the bottom side of the module.
5. Follow the instructions in the VXI-SC-1000 Carrier Module
Installation Guide to replace the VXI-SC-1102 submodule in the
carrier assembly.

Figure 2-2. Bent and Trimmed Resistor

2-21
Chapter 2
Installation
VXI-SC-1102 User Manual

Advertisement

Table of Contents
loading

This manual is also suitable for:

Vxi-sc-1102

Table of Contents