Olympus VISERA ELITE CLV-S190 Instructions Manual
Olympus VISERA ELITE CLV-S190 Instructions Manual

Olympus VISERA ELITE CLV-S190 Instructions Manual

Xenon light source
Table of Contents

Advertisement

INSTRUCTIONS
VISERA ELITE XENON LIGHT SOURCE
OLYMPUS CLV-S190
Labels and Symbols
Important Information — Please Read Before
Use
Summary of the Functions of the Instrument
Chapter 1
Checking the Package Contents
Chapter 2
Nomenclature and Functions
Chapter 3
Installation and Connection
Chapter 4
Inspection
Chapter 5
Operation
Chapter 6
Lamp Replacement
Chapter 7
Care, Storage, and Disposal
Chapter 8
Troubleshooting
Appendix
1
4
12
15
17
25
37
57
71
83
87
93

Advertisement

Table of Contents
loading

Summary of Contents for Olympus VISERA ELITE CLV-S190

  • Page 1 Labels and Symbols Important Information — Please Read Before Summary of the Functions of the Instrument VISERA ELITE XENON LIGHT SOURCE Chapter 1 Checking the Package Contents OLYMPUS CLV-S190 Chapter 2 Nomenclature and Functions Chapter 3 Installation and Connection Chapter 4...
  • Page 3: Table Of Contents

    Contents Contents Labels and Symbols ......................1 Important Information — Please Read Before Use ............4 Intended use .......................... 4 Applicability of endoscopy and endoscopic treatment ............4 Instruction manual ......................... 4 User qualifications ......................... 6 Instrument compatibility ......................6 Repair and modification ......................
  • Page 4 Contents Chapter 4 Inspection ................37 Attention to inspection ..................37 Inspection workflow ....................38 Connection of an endoscope ................39 Inspection of the power supply ................41 Checking the lamp usage indicator ..............43 Inspection of the examination light ..............44 Inspection of the brightness mode switching function ........
  • Page 5 Contents Chapter 7 Care, Storage, and Disposal ..........83 Care ......................... 83 Storage ........................85 Disposal ........................86 Examination lamp ........................ 86 The light source or any of its components ................86 Chapter 8 Troubleshooting ..............87 Troubleshooting ..................... 87 Troubleshooting guide ...................
  • Page 6 Contents CLV-S190 INSTRUCTION MANUAL...
  • Page 7: Labels And Symbols

    Labels and Symbols Labels and Symbols Safety-related labels and symbols are attached to the light source at the locations shown below. If labels or symbols are missing or illegible, contact Olympus.  Front side OPT DGTL MODE LAMP HIGH STBY...
  • Page 8 Labels and Symbols  Rear side Electrical rating POWER INPUT 100 – 240 V 50/60 Hz 500 VA Potential equalization Product name terminal Indicates to see the instruction manual for the cable to be connected. Cautions that the inside is extremely hot immediately after the lamp is extinguished.
  • Page 9 Labels and Symbols  Back cover of this instruction manual Symbol Description Manufacturer Authorized representative in the European Community CLV-S190 INSTRUCTION MANUAL...
  • Page 10: Important Information - Please Read Before Use

     Intended use The light source has been designed to be used with Olympus endoscopes, video system centers, and other ancillary equipment for endoscopic diagnosis, treatment and video observation. Do not use the light source for any purpose other than its intended use.
  • Page 11 Important Information — Please Read Before Use  Terms used in this manual Video system center: The video system center is a device that converts signals from a videoscope or camera head into monitor images. Camera head: The camera head is a device that converts endoscopic images from a fiberscope or a rigid endoscope into monitor images.
  • Page 12: User Qualifications

    Some problems that appear to be malfunctions may be correctable by referring to Chapter 8, “Troubleshooting”. If the problem cannot be resolved using the information in Chapter 8, contact Olympus. The light source is to be repaired by Olympus technicians only. CLV-S190 INSTRUCTION MANUAL...
  • Page 13: Signal Words

    Such contact may cause unintended current flow to the patient.  If fluids are spilled on or into the light source, stop operation of the light source immediately and contact Olympus.  Do not prepare, inspect, or use the light source with wet hands.
  • Page 14 Important Information — Please Read Before Use WARNING • In case of light source failure or malfunction, always keep another light source in the room ready for use. • Never insert anything into the ventilation grills of the light source. It can cause an electric shock.
  • Page 15 Important Information — Please Read Before Use WARNING • If the endoscopic image dims during use, blood, mucus, or debris may adhere to the light guide on the distal end of the endoscope. Carefully withdraw the endoscope from the patient and remove the blood or mucus to obtain optimum illumination and to ensure the safety of the examination.
  • Page 16 Important Information — Please Read Before Use CAUTION • Do not leave the examination lamp ON when an endoscope is connected to the light source. The examination light reaches the maximum intensity and the endoscope’s distal end becomes hot. In addition, smoke may also be produced if the debris attached to the distal end is heated.
  • Page 17: Cardiac Applications

    “TYPE CF applied part” requirements for leakage current limits if they are to be used for cardiac applications. NOTE • The OLYMPUS light guide cables and camera heads listed in the “ System chart” on page 93 (TYPE CF applied part) that are suitable for cardiac applications bear a mark.
  • Page 18: Summary Of The Functions Of The Instrument

    Summary of the Functions of the Instrument Summary of the Functions of the Instrument Some of the functions described below may be unavailable or restricted depending on the combined equipment. For more details, refer to the instruction manuals for the light source and the required ancillary equipment.
  • Page 19 Summary of the Functions of the Instrument  Automatic switching to the emergency lamp If the examination lamp does not light up or blows in the middle of an examination making endoscopic observation impossible, the light source switches automatically to the emergency light.
  • Page 20 Summary of the Functions of the Instrument CLV-S190 INSTRUCTION MANUAL...
  • Page 21: Chapter 1 Checking The Package Contents

    Match all items in the package with the components shown below. Inspect each item for damage. If the light source is damaged, a component is missing, or you have any questions, do not use the light source and immediately contact Olympus. ...
  • Page 22 1.1 Checking the package contents list Ch.1 CLV-S190 INSTRUCTION MANUAL...
  • Page 23: Chapter 2 Nomenclature And Functions

    2.1 Nomenclature and functions Chapter 2 Nomenclature and Functions Nomenclature and functions  Front panel Symbol Description Symbol Description Ch.2 Power ON/OFF Brightness Decrease brightness Increase brightness Exchange lamp Lamp usage indicator reset (lamp usage indicator) New lamp Emergency lamp (lamp usage indicator) Examination lamp ignition Mode select...
  • Page 24 2.1 Nomenclature and functions  Rear and side panels Symbol Description Symbol Description Refer to instructions. Warns that the lamp is hot. Potential equalization Serial number terminal Alternating current Caution Refer to instructions. Ch.2 CLV-S190 INSTRUCTION MANUAL...
  • Page 25: Front Panel

    2.2 Front panel Front panel OPT DGTL MODE LAMP HIGH STBY STBY Ch.2 BRIGHTNESS AUTO MANU COUNTER RESET 100h 300h 500h SPARE LAMP Nomenclature Description Power switch Press to turn the light source ON or OFF. Power indicator Lights up when the light source is ON. Output socket Connect the endoscope, the endoscope’s light connector, or the light guide cable to this socket.
  • Page 26 2.2 Front panel  Control panel (buttons) OPT DGTL MODE LAMP HIGH STBY STBY BRIGHTNESS Ch.2 AUTO MANU COUNTER RESET 100h 300h 500h SPARE LAMP CLV-S190 INSTRUCTION MANUAL...
  • Page 27 2.2 Front panel Nomenclature Description Observation mode Press to switch the observation mode alternately between normal light observation and button optical-digital observation modes.  Section 5.6, “Optical-digital observation” Observation mode These indicators display whether the optical-digital observation modes (ON) or normal indicators light observation mode (STBY) is activated.
  • Page 28: Rear And Side Panels

    2.3 Rear and side panels Rear and side panels  Rear panel Ch.2  Side panel CLV-S190 INSTRUCTION MANUAL...
  • Page 29 2.3 Rear and side panels Nomenclature Description System terminal Not used. OTV terminal This terminal is the receptacle for the light source cable (MAJ-1959) to use the video system center.  Section 3.4, “Connection of the video system center” Foot switch This terminal is the receptacle for the foot switch (MAJ-1391) to enable naked-eye PDD terminal observation.
  • Page 30 2.3 Rear and side panels Ch.2 CLV-S190 INSTRUCTION MANUAL...
  • Page 31: Chapter 3 Installation And Connection

    3.1 Attention to installation and connection Chapter 3 Installation and Connection Prepare the light source and compatible equipment (shown in the “ System chart” on page 93) before each use. Install and connect the equipment according to the procedures described in this chapter and the instruction manuals for ancillary equipment.
  • Page 32: Installation Workflow

    3.2 Installation workflow Installation workflow Please see the installation workflow below. Follow each step of the workflow before using the light source and the ancillary equipment. Install the light source and the ancillary equipment on the mobile workstation or another stable location. Section 3.3 on page 27 ...
  • Page 33: Installation Of Equipment

    3.3 Installation of equipment Installation of equipment CAUTION • Do not place any equipment other than the video system center on the top of the light source. Equipment damage can result. • Keep the ventilation grills of the light source clear. The ventilation grills are located on the side and rear panels.
  • Page 34: Installation On The Mobile Workstation (Wm-Np2, Wm-Np1, Or Wm-Wp1)

    3.3 Installation of equipment  Installation on the mobile workstation (WM-NP2, WM-NP1, or WM-WP1) Place the mobile workstation on a level surface. Lock the caster brakes by pushing them down as shown in Figure 3.1. Push to release. Push to lock. Ch.3 Caster brakes Figure 3.1...
  • Page 35: Installation In Another Location

    3.3 Installation of equipment NOTE If the instrument is placed on a mobile workstation, only two foot holders are necessary.  Installation in another location Place the pattern sheet and the foot holders on the installation location. Peel the paper from the bottom of the foot holders and lightly attach the foot holders to the sheet.
  • Page 36: Connection Of The Video System Center

    3.4 Connection of the video system center Connection of the video system center Connect the video system center to the light source using the cables listed below.  Compatible video system center Model Product name OTV-S190 VISERA ELITE video system center OTV-S7Pro VISERA Pro video system center Table 3.1...
  • Page 37: Otv-S190

    3.4 Connection of the video system center  OTV-S190  Cable Model Product name MAJ-1959 LIGHT SOURCE CABLE Table 3.2  Connecting diagram Light source terminal OTV-S190 Ch.3 MAJ-1959 OTV terminal Figure 3.4 CLV-S190 INSTRUCTION MANUAL...
  • Page 38: Otv-S7Pro

    3.4 Connection of the video system center  OTV-S7Pro  Cable Model Product name MAJ-1411 LIGHT SOURCE CABLE Table 3.3  Connecting diagram OTV-S7Pro Ch.3 Light source terminal MAJ-1411 OTV terminal Figure 3.5 CLV-S190 INSTRUCTION MANUAL...
  • Page 39: Connection Of The Foot Switch

    3.5 Connection of the foot switch Connection of the foot switch Connect the foot switch (MAJ-1391, optional) to the foot switch terminal. Foot switch terminal Ch.3 MAJ-1391 Figure 3.6 NOTE • If the foot switch is connected to the light source and the video system center is ON, the function of the foot switch depends on the setting of the video system center.
  • Page 40: Connection To The Ac Mains Power Supply

    3.6 Connection to the AC mains power supply Connection to the AC mains power supply DANGER • Be sure to connect the power plug of the power cord directly to the mobile workstation (WM-NP2, WM-NP1, or WM-WP1) or a grounded wall mains outlet. If the light source is not grounded properly, it can cause an electric shock.
  • Page 41 3.6 Connection to the AC mains power supply Confirm that the power switch is not pressed in. Connect the power cord to the AC power inlet of the light source and to the mobile workstation (WM-NP2, WM-NP1, or WM-WP1) or the wall mains outlet. Power cord Figure 3.7 Ch.3...
  • Page 42 3.6 Connection to the AC mains power supply Ch.3 CLV-S190 INSTRUCTION MANUAL...
  • Page 43: Chapter 4 Inspection

    4.1 Attention to inspection Chapter 4 Inspection Inspect the light source and other equipment to be used with the light source. Refer to the respective instruction manuals for each piece of equipment. Attention to inspection WARNING • Review Chapter 3, “Installation and Connection” thoroughly, and prepare the instruments properly before inspection.
  • Page 44: Inspection Workflow

    4.2 Inspection workflow Inspection workflow Follow each step of the workflow for inspection of the light source before use. Connect an endoscope. Section 4.3 on page 39  Confirm that the light source is turned ON normally. Section 4.4 on page 41 ...
  • Page 45: Connection Of An Endoscope

    4.3 Connection of an endoscope Connection of an endoscope WARNING • Do not look directly into the distal end of the endoscope, the distal end of the light guide cable, or the output socket of the light source when they are emitting light. The intense light may cause eye injury.
  • Page 46 4.3 Connection of an endoscope Inspect the light guide cable and rigid endoscope as described in the endoscope’s instruction manuals. Connect the light guide cable to the rigid endoscope. Insert the light guide connector into the output socket on the front panel of the light source Output socket until it clicks.
  • Page 47: Inspection Of The Power Supply

    4.4 Inspection of the power supply Inspection of the power supply Confirm that the ventilation grills are not covered with dust, that the lamp cover is attached, and that the light source is turned ON. WARNING Turning ON the light source ignites the examination lamp automatically. Do not look directly into the distal end of the endoscope or the output socket of the light source when they are emitting light.
  • Page 48 If the emergency lamp indicator on the control panel blinks The emergency lamp is damaged. Stop the use and contact Olympus.  If the indicators on the control panel light up or blink The light source is not working properly. Stop the use and contact Olympus. CLV-S190 INSTRUCTION MANUAL...
  • Page 49: Checking The Lamp Usage Indicator

    4.5 Checking the lamp usage indicator Checking the lamp usage indicator Check the lamp usage indicator. When the “500 h” indicator on the lamp usage indicator lights up, replace the examination lamp with a new one as described in Section 6.1, “Replacement of the examination (xenon) lamp”. Ch.4 Lamp usage indicator “500 h”...
  • Page 50: Inspection Of The Examination Light

    4.6 Inspection of the examination light Inspection of the examination light WARNING Do not look directly into the distal end of the endoscope, the distal end of the light guide cable, or the output socket of the light source when they are emitting light. The intense light may cause eye injury.
  • Page 51: Inspection Of The Brightness Mode Switching Function

    4.7 Inspection of the brightness mode switching function Inspection of the brightness mode switching function Confirm that the brightness mode can be switched between “AUTO” and “MANU”. CAUTION When using a fiberscope or a rigid endoscope without camera head, set the brightness mode to “MANU”.
  • Page 52: Inspection Of Brightness Adjustment

    4.8 Inspection of brightness adjustment Inspection of brightness adjustment Confirm that the brightness of the examination lamp can be adjusted. The inspection method differs according to the endoscope to be used. Select either the “AUTO” or “MANU” mode as shown in Table 4.1. Brightness Connected endoscopes Reference...
  • Page 53: Inspection Of Automatic Brightness Adjustment

    4.8 Inspection of brightness adjustment  Inspection of automatic brightness adjustment WARNING When disconnecting the camera head from the endoscope without turning the examination lamp OFF, make sure that the brightness mode indicator is set to “MANU” and that the brightness is set to the minimum using the brightness button. If the camera head is disconnected while the brightness mode indicator is set to “AUTO”, the intense light may cause eye injury.
  • Page 54 CAUTION Confirm that a beep is heard when pressing the brightness button. When a beep is not heard, the light source may malfunction. Contact Olympus. NOTE The brightness indication is interlocked with the brightness level indication of the connected video system center.
  • Page 55: Inspection Of Manual Brightness Adjustment

    4.8 Inspection of brightness adjustment  Inspection of manual brightness adjustment WARNING When using the manual brightness adjustment, always set the brightness to the minimum level necessary to complete the examination. If the light is too bright, eye injury or burns can result. Press the lamp button on the control panel: the Lamp button examination lamp is ignited.
  • Page 56 CAUTION Confirm that a beep is heard when pressing the brightness button. When a beep is Ch.4 not heard, the light source may malfunction. Contact Olympus. NOTE The brightness indication is interlocked with the brightness level indication of the connected video system center. When the brightness buttons on the video system center are pressed, the brightness indication on the light source varies in an interlocked operation.
  • Page 57: Inspection Of The Optical-Digital Observation Function

    4.9 Inspection of the optical-digital observation function Inspection of the optical-digital observation function Optical-digital observation is available when the light source is used in combination with the videoscope or camera head compatible with the optical-digital observation. Confirm that the optical-digital observation mode is switched. The PDD observation is available when installing an optional PDD filter (MAJ-1429) in the light source.
  • Page 58  Camera head, such as OTV-S7Pro-FD NOTE • Contact Olympus for details on installing the optional PDD filter (MAJ-1429). • When the foot switch (MAJ-1391, optional) is connected to the light source, the observation mode can be switched during naked-eye PDD observation using the foot switch instead of the observation mode button.
  • Page 59: 4.10 Inspection Of The High Intensity Mode

    4.10 Inspection of the high intensity mode 4.10 Inspection of the high intensity mode Press the brightness mode button on the Brightness mode indicators control panel and confirm that the “MANU” indicator lights up. Brightness mode button Figure 4.17 Press the brightness buttons on the control Brightness indicator panel to set the brightness level to 0.
  • Page 60 • For compatibility of the endoscope or the light guide cable with the high intensity mode, refer to the instruction manual for the endoscope or the light guide cable to be used or contact Olympus. • The intensity setting is automatically saved when the light source is turned OFF and is recalled when the light source is turned ON again.
  • Page 61: 4.11 After Inspection

    4.11 After inspection 4.11 After inspection If the light source will not be used immediately Control panel after inspection, press the lamp button on the control panel for about 1 second to turn OFF the examination lamp (refer to Section 5.8, “Extinguishing the examination lamp”).
  • Page 62 4.11 After inspection Ch.4 CLV-S190 INSTRUCTION MANUAL...
  • Page 63: Chapter 5 Operation

    5.1 Attention to operation Chapter 5 Operation The operator of the light source must be a physician or medical personnel under the supervision of a physician and must have received sufficient training in clinical endoscopic technique. This manual, therefore, does not explain or discuss clinical endoscopic procedures. It only describes basic operation and attention related to the operation of the light source.
  • Page 64 Leaving the examination lamp ON will cause the distal end of the endoscope to become hot and could cause operator and/or patient burns. • Use only Olympus high-frequency electrosurgical equipment with this unit. Non-Olympus equipment can cause instability of the automatic brightness adjustment.
  • Page 65 5.1 Attention to operation WARNING • If the endoscopic image seems to be dark in an optical-digital observation mode, change to the normal observation mode. The examination might not be performed safely. CAUTION • Do not press the observation mode button when the video system center is OFF (or is not connected with the light source) and a light guide cable incompatible with PDD is used.
  • Page 66: Operation Flow

    5.2 Operation flow Operation flow Please see the operation workflow below. Follow each step of the workflow for using the light source. Inspect the light source and the ancillary equipment. Chapter 4 on page 37  Turn the light source ON and ignite the examination lamp. Section 5.3 on page 61 ...
  • Page 67: Turning The Light Source On And Igniting The Examination Lamp

    5.3 Turning the light source ON and igniting the examination lamp Turning the light source ON and igniting the examination lamp WARNING When turning ON the light source, never allow the distal end of an endoscope or a light guide cable to come in contact with the patient and/or other flammable materials, such as operating room drapes.
  • Page 68: Setting The Brightness Mode

    5.4 Setting the brightness mode Setting the brightness mode Set the brightness mode of the light source according to the endoscope to be used. The brightness mode determines the examination light supplied to the endoscope and the method of adjusting its intensity.
  • Page 69: Brightness Adjustment

    5.5 Brightness adjustment Brightness adjustment Adjust the brightness of the examination lamp. The adjustment method varies depending on the brightness adjustment mode set in Section 5.4, “Setting the brightness mode”.  Automatic brightness adjustment WARNING When disconnecting the camera head from the endoscope without turning the lamp OFF, set the brightness mode indicator to “MANU”...
  • Page 70 5.5 Brightness adjustment NOTE • The standard brightness setting is “0”. • The standard brightness may be too bright or too dark depending on the endoscope type and observation region. In this case, adjust the brightness as required. • Pressing either of the brightness buttons (“–” or “+”) once moves the level one step; pressing the button continuously moves the level continuously.
  • Page 71: Manual Brightness Adjustment

    5.5 Brightness adjustment  Manual brightness adjustment WARNING Always adjust the examination light to the minimum required brightness for observation, and do not bring the examination light in the proximity of a mucous membrane for an extended period. Use of higher brightness than required may cause eye injury or burns to the patient.
  • Page 72: Optical-Digital Observation

    5.6 Optical-digital observation Optical-digital observation Optical-digital observation is available when the light source is used in combination with the videoscope compatible with the optical-digital observation. WARNING • Do not rely on the optical-digital method alone for primary detection of lesions or for a decision regarding any potential diagnostic or therapeutic intervention.
  • Page 73  Camera head, such as OTV-S7Pro-FD NOTE • Contact Olympus for details on installing the optional PDD filter (MAJ-1429). • When the foot switch (MAJ-1391, optional) is connected to the light source, the observation mode can be switched during naked-eye PDD observation using the foot switch instead of the observation mode button.
  • Page 74: High Intensity Mode

    5.7 High intensity mode High intensity mode CAUTION When switching the normal intensity mode to high intensity mode, be sure to set the brightness level at or below 0. Otherwise, the brightness will exceed the necessary level. It may result in operator and/or patient injury. NOTE The high intensity mode is not available when no endoscope is connected to the light source.
  • Page 75: Extinguishing The Examination Lamp

    5.8 Extinguishing the examination lamp Extinguishing the examination lamp Press and hold the lamp button for about Lamp indicator “ON” 1 second : the examination lamp turns OFF, and the indicator “STBY” lights up. Lamp button Figure 5.8 Ch.5 CLV-S190 INSTRUCTION MANUAL...
  • Page 76: Turning The Light Source Off

    WARNING When the power indicator is still lighting up after turning OFF the power, stop using the light source and disconnect the power cord. Then please contact Olympus. Disconnect the endoscope as described in the endoscope’s instruction manual. If the light source is not to be used for an extended period of time, disconnect the power cord from the hospital-grade wall main outlet.
  • Page 77: Chapter 6 Lamp Replacement

    Olympus. • Xenon lamp MD-631 WARNING • Never install a lamp that has not been approved by Olympus. The use of a nonapproved lamp can cause damage to the light source and ancillary equipment, malfunction or a fire. • Turn the light source OFF and disconnect the power cord from the wall mains outlet before replacing the lamp with a new one.
  • Page 78 6.1 Replacement of the examination (xenon) lamp CAUTION • Do not touch the glass surface (filter) inside of the lamp chamber. Natural skin moisture from your fingers can cause cracks and damage the light source. • Do not touch the glass surface of the lamp, mesh, or reflector. Natural skin moisture from your fingers can cause cracks and damage the light source.
  • Page 79: Removal Of The Lamp

    6.2 Removal of the lamp Removal of the lamp When the examination lamp is ON, press and hold the lamp button for about 1 second to extinguish it. Wait for a few minutes to allow the light source to cool the lamp chamber sufficiently. NOTE While the light source is ON and the examination lamp is extinguished, a fan built into the light source cools the lamp chamber.
  • Page 80 6.2 Removal of the lamp Remove the hexagon wrench from the rear of Hexagon wrench the lamp cover. Lamp cover Figure 6.2 Turn knob (A) counterclockwise by 90 to loosen it. Lamp chamber Knob (B) Knob (A) Ch.6 Figure 6.3 Turn knob (B) counterclockwise by 90...
  • Page 81 6.2 Removal of the lamp Holding the knobs or projections of the heat sinks, remove the examination lamp with heat sinks (A) and (B) attached. Heat sink (A) Heat sink (B) Projection Figure 6.4 Using the hexagon wrench, loosen the three bolts on heat sink (B) (on the “+” side of the examination lamp and remove heat sink (B) from the examination lamp.
  • Page 82 6.2 Removal of the lamp CAUTION Washers are attached to bolts. Retain all three of them carefully because they will be used again to clamp the new examination lamp. Bolt Washer Figure 6.6 Remove the clamp of heat sink (A) and pull out the examination lamp. (See Figure 6.5) Using a clean, lint-free cloth, wipe off any residual heat compound from the heat sink.
  • Page 83: Insertion Of The Lamp

    6.3 Insertion of the lamp Insertion of the lamp CAUTION • Do not apply the heat compound to the glass surface and the ceramic part of the examination lamp. If any compound gets on the glass surface, wipe it off with a clean, lint-free cloth.
  • Page 84 6.3 Insertion of the lamp CAUTION • When inserting the examination lamp into the heat sink, align their pin positions and tighten the bolts firmly. If the bolts are not tightened firmly, poor heat radiation may result in equipment damage, examination lamp ignition failure, and a considerable drop in the life of the examination lamp.
  • Page 85 6.3 Insertion of the lamp Place heat sink (A) and heat sink (B) so that Heat sink (A) their undersides are flat and close the heat sink Heat sink (B) clamp firmly. Clamp Figure 6.11 CAUTION Be sure to tighten the heat sink clamp firmly. Poor heat radiation may result in equipment damage, examination lamp ignition failure, and a considerable drop in the life of the examination lamp.
  • Page 86 6.3 Insertion of the lamp Turn the knob (B) clockwise by 90 from the horizontal position until it stops while pushing the knob. Figure 6.13 Turn the knob (A) clockwise by 90 from the horizontal position until it stops while pushing the knob.
  • Page 87: Lamp Usage Indicator Reset

    6.4 Lamp usage indicator reset Turn the knobs of the lamp cover and close the Lamp cover lamp cover securely. Lamp chamber Knob Figure 6.15 NOTE For safety, the light source is designed so that it cannot be turned ON if the lamp cover is not attached securely.
  • Page 88 6.4 Lamp usage indicator reset Connect the power cord and press the power switch: the light source is turned ON. Press the lamp button on the control panel for Lamp indicator “STBY” 1 second: the examination lamp is turned OFF, and the “STBY”...
  • Page 89: Chapter 7 Care, Storage, And Disposal

    7.1 Care Chapter 7 Care, Storage, and Disposal Care WARNING • After wiping with a piece of moistened lint-free cloth, dry the light source thoroughly before using it again. If it is used while still wet, there is a risk of an electric shock. •...
  • Page 90 7.1 Care Turn the light source OFF. Disconnect the power cord from the hospital-grade wall mains outlet. Power plug Figure 7.1 NOTE The plug of the power cord that follows the equipment may vary from this drawing, as in each country the commercialized cable is in accordance to the local applicable norm.
  • Page 91: Storage

    7.2 Storage Storage If the light source will not be used for a long time, store it according to the following instructions. Also, clean the light source as described in Section 7.1, “Care” before storage. CAUTION Do not store the light source in locations exposed to direct sunlight, X-rays, radioactivity, or strong electromagnetic radiation (e.g., in the vicinity of a microwave therapeutic equipment, MRI, short-wave therapeutic equipment, radio equipment, or cellular/portable phone).
  • Page 92: Disposal

    7.3 Disposal Disposal  Examination lamp WARNING • Discharge the internal gas when disposing of a used lamp. The glass could break due to the high internal pressure of the lamp. • When cutting the projection part of the electrode on the “+” side of the examination lamp, wear appropriate protection equipment at all times and make sure to cover the lamp with a cloth.
  • Page 93: Chapter 8 Troubleshooting

    Connection”, or use described in Chapter 5, “Operation”, do not use the light source and contact Olympus. Some problems that appear to be malfunctions may be correctable by referring to Section 8.2, “Troubleshooting guide”. If the problem cannot be resolved by the described remedial action, stop using the light source and contact Olympus.
  • Page 94 “Replacement of the examination (xenon) lamp”. The emergency lamp The emergency lamp has blown or Immediately turn the light source OFF, indicator blinks. failed. unplug the power cord from the wall mains outlet and the power inlet, then contact Olympus. CLV-S190 INSTRUCTION MANUAL...
  • Page 95 8.2 Troubleshooting guide Irregularity description Possible cause Solution The examination lamp does The temperature of the light source is Turn the light source OFF and confirm not ignite, the indicators on too high. that the ventilation grills are not the front panel blink, and a covered.
  • Page 96 8.2 Troubleshooting guide Irregularity description Possible cause Solution The high intensity mode An incompatible endoscope or light Connect an endoscope or light guide cannot be set. guide is connected to the output socket. cable with high intensity mode compatibility as described in Section 4.10, “Inspection of the high intensity mode”...
  • Page 97: Returning The Light Source For Repair

    8.3 Returning the light source for repair Returning the light source for repair When returning the light source for repair, contact Olympus. With the light source, include a description of the malfunction or damage and the name and telephone number of the individual at your location who is the most familiar with the problem.
  • Page 98 8.3 Returning the light source for repair Ch.8 CLV-S190 INSTRUCTION MANUAL...
  • Page 99: Appendix

    The recommended combinations of equipment and accessories that can be used with the light source are listed below. New products released after the introduction of the light source may also be compatible for use in combination with the light source. For further details, contact Olympus. WARNING If combinations of equipment other than those shown below are used, the full responsibility is assumed by the medical treatment facility.
  • Page 100 Flexible videoscope (ENF series, CYF series, HYF series, LTF series, LF series) App. *1 The OLYMPUS video system can be configured for cardiac applications only when TYPE CF applied part (camera heads and light guide cables) bearing the symbol are used.
  • Page 101 Combination equipment High-definition LCD monitor (OEV261H, OEV191H) High-definition monitor (OEV181H) Keyboard LCD monitor (OEV191) VISERA ELITE video system center (OTV-S190) VISERA Pro video system center (OTV-S7Pro) Mobile workstation (WM-NP2, WM-NP1, WM-WP1) VISERA ELITE xenon light source (CLV-S190) Foot switch (MAJ-1391) App.
  • Page 102: Transportation, Storage, And Operating Environments

    Transportation, storage, and operating environments Transportation, storage, and operating environments Operating Ambient temperature 10 – 40C (50 – 104F) environment Relative humidity 30 – 85% (without condensation) Atmospheric 700 – 1060 hPa pressure (0.7 – 1.1 kgf/cm (10.2 – 15.4 psia) Standard storage Ambient temperature 5 –...
  • Page 103 Specifications Illumination Examination lamp Xenon short-arc lamp (ozone-free) 300 W Average lamp life Approximately 500 hours of continuous use (with intermittent use, the lamp life may vary slightly.) Ignition method Switching regulator Brightness Light-path diaphragm control adjustment Cooling Forced-air cooling Intensity mode Normal or high intensity NBI observation...
  • Page 104 Electrical and Electronic Equipment, this symbol indicates that the product must not be disposed of as unsorted municipal waste, but should be collected separately. Refer to your local Olympus distributor for return and/or collection systems available in your country. Applied standard...
  • Page 105: Emc Information

    EMC information EMC information  Guidance and manufacturer’s declaration — Electromagnetic emissions This model is intended for use by medical personnel in hospital environments and for use in the electromagnetic environment specified below. The customer or the user of this model should assure that it is used in such an environment.
  • Page 106 The customer or the user of this model should assure that it is used in such an environment. This instrument can be used with the high-frequency electrosurgical equipment that designated by Olympus. IEC 60601-1-2 IEC 60601-1-2...
  • Page 107 EMC information IEC 60601-1-2 IEC 60601-1-2 IEC 60601-1-2 (2007, 2001) Compliance Immunity test (2014) (2007, 2001) Electromagnetic level test level test level environment — Guidance Voltage dips, short 0% U < 5% U Same as left Mains power quality should be interruptions, and (100% dip in U (>...
  • Page 108 EMC information IEC 60601-1-2 IEC 60601-1-2 IEC 60601-1-2 (2007, 2001) Immunity Compliance (2014) (2007, 2001) Electromagnetic environment — test level test level test level Guidance Conducted RF 3V (V Same as left Recommended separation distance IEC 61000-4-6 (150 kHz – (150 kHz –...
  • Page 109 EMC information NOTE • Field strength from fixed RF transmitters as determined by an electromagnetic site survey should be less than the compliance level in each frequency range Field strength from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy.
  • Page 110 Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of this instrument, including cables specified by Olympus. Otherwise, degradation of the performance of this equipment could result.
  • Page 111 ©2011 OLYMPUS MEDICAL SYSTEMS CORP. All rights reserved. No part of this publication may be reproduced or distributed without the express written permission of OLYMPUS MEDICAL SYSTEMS CORP. OLYMPUS is a registered trademark of OLYMPUS CORPORATION. Trademarks, product names, logos, or trade names used in this document are generally registered trademarks or trademarks of each company.
  • Page 112 A8F, Ping An International Financial Center, No. 1-3, Xinyuan South Road, Chaoyang District, Beijing, 100027 P.R.C. Fax: (86)10-5976-1299 Telephone: (86)10-5819-9000 Olympus Tower 9F, 446, Bongeunsa-ro, Gangnam-gu, Seoul, Korea 135-509 Fax: (02)6255-3494 Telephone: (02)6255-3210 491B, River Valley Road #12-01/04, Valley Point Office Tower, Singapore 248373...

Table of Contents