OPERATING INSTRUCTION MANUAL
FOR
INDICATION/ALARM UNIT EC-592
FOR USERS

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the EC-592 and is the answer not in the manual?

Questions and answers

Summary of Contents for Riken Keiki EC-592

  • Page 1 PT1E-0881 OPERATING INSTRUCTION MANUAL INDICATION/ALARM UNIT EC-592 FOR USERS...
  • Page 2: Table Of Contents

    INDEX 1.PRODUCT OUTLINE 1-1. In the beginning …………………………………………… 1-2. Application for use ………………………………………… 1-3. Identification of each cautional marks ……………………… 2.IMPORTANT INSTRUCTION FOR THE SAFETY 2-1. Danger items………………………………………………… 2-2. Warning items………………………………………………… 4 2-3. Caution items………………………………………………… 5 3.PRODUCT FUNCTION 3-1. External drawing……………………………………………… 6 3-2....
  • Page 3 7.MAINTENANCE CHECK 7-1. Frequency of maintenance and check items 7-1-1. Daily check………………………………………… 29 7-1-2. Regular maintenance check………………………… 29 7-2. Maintenance contract for regular check……………………… 29 7-3. Replacement of sensor and parts……………………………… 30 7-4. Treatment when operation is stopped or its installation place is moved. 7-4-1....
  • Page 4: 1.Product Outline

    1.PRODUCT OUTLINE 1-1.In the beginning First of all, we wish to express our heartfelt thanks for your purchase of our indicator/alarm unit EC-592. This instruction manual is just a guide book to operate indicator/alarm unit EC-592 of RM-590 series. Your kind reading of this manual is requested not only for first user but for already experienced staff.
  • Page 5: 2.Important Instruction For The Safety

    2.IMPORTANT INSTRUCTION FOR THE SAFETY 2-1.Danger items DANGER This is not a explosion proof unit. Do not use it at hazardous area. 2-2.Warning items WARNING ・ Detector head Do not connect other detector head than designated toxic gas detector head. If connected, you can not detect gas.
  • Page 6: 2-3. Caution Items

    2-3.Cautional items CAUTION ・ Do not use walky-talky near this unit. When this unit receives the electric waves by walky-talky etc near from the cable, this will effect the reading. When use the walky-talky, operate it at the place where there is no influence of electric wave ejection. ・...
  • Page 8: 3-3.Panel Cut-Out Dimensions

    3-3.Panel cut-out dimensions Dimensions with single case 1.Mounting for 2 steps in vertical and N lines in horizontal +0.6 (36×N-3) 0 2.Mounting for 1 step in vertical and N lines in horizontal +0.6 (36×N-3) 0 Required thickness of the panel : 1.2~3.2mm Refer to the instruction manual for multi-unit case if installed the unit to multi-unit case.
  • Page 10: 3-5.Block Diagram

    3-5.Block diagram EC-592 Display part (PW/TR) SKIP (AL1) (AL2) Power supply part Operation part POWER INPUT (MODE) (UP) (DOWN) (SET) (DC24V) Control part Alarm contact control part Gas alarm contact(AL1、AL2) Trouble alarm contact (TROUBLE) Amp part ※Standard setting ・Non-exciting (Non-exciting at normal) ・A-contact 4-20mA (Contact closes at alarm)...
  • Page 11: 4.How To Use

    4.HOW TO USE 4-1.Before operation Be sure to keep cautional items of use not only for first user but for already experienced staff. If not keep these cautional items, the unit may be defective and correct gas detection may not be performed. 4-2.Mounting/dismounting procedures (1)Mounting method ・Open the front cover of this unit.
  • Page 12: 4-3.Installation Place

    4-3.Installation place CAUTION Do not install at places where direct sun drought or where the temperature is drastically changed. Keep away the system from direct sun drought or where the temperature of the system is drastically changed. Do not install it where vibration or shocks may take place. The system consists of fine electronics parts.
  • Page 13: 4-4.Caution In The System Engineering

    4-4.Caution in the system engineering CAUTION Unstable power and noise may cause error of performance and alarm. For the system to use this unit, it is required to make design based on this manual descriptions. (1)Stable power used While the system gets stable at power failure, the external output and alarm contact may be on and the care for it must be taken.
  • Page 14 -How to think alarm contacts against inductive load- The spec for alarm contact of EC-592 is described by the conditions of resistive load. When use the inductive load for alarm contacts, the very high reverse electromotive voltage may be generated and the following trouble tends to be produced.
  • Page 15 CAUTION ・The inductive load shall not be used in principle (Do not use fluorescent lamp). ・When use the inductive load, make the contact amplification outside, but the outside relay coil belong, use the relay driven by the low voltage (within AC100V) and it is protected by an appropriate surge killer.
  • Page 16: 4-5.Grounding

    4-5.Grounding Make grounding with terminal WARNING Be sure to make grounding before putting power ON. Terminal *The above fig shows the unit with single case. In case of multi-unit case, refer to instruction manual for multi-unit case. Be sure to make grounding for safety purpose and to keep stable operation of this unit. Do not connect each wire with gas pipe.
  • Page 17: 4-6.Wiring

    4-6.Wiring (In case of single case) ① ② DC24V±10% TD1 ③ GND (-) ④ ① ③ Note 1 ② CVVS 1.25sq-2c ④ TD2 Detector head ③ ⑤ ④ ⑥ ① ② AL1 1 alarm contact output TD3 ③ AL2 2 alarm contact output ④...
  • Page 18: 5.Operation Method

    5.OPERATION METHOD 5-1.Start-up ● Before putting power ON…………… Before putting power ON, keep following caution items. If not, danger for electric shock or damage to the unit will be happen. (1)Make grounding. (2)Check that wirings with outer units are made correctly. (3)Check that the supply voltage is within a rating.
  • Page 27: 5-6.How To Finish Operation

    5-6.How to finish operation When finish this operation, turn off power switch of this unit and turn off the main power (DC24V). WARNING ・When finish this operation, do it after making point skip with the upper system (Centralized system). ・When finish this operation, check the external output and function of outer equipment to be connected with external alarm contact.
  • Page 28: 6-1. Kind Of Alarm

    6.KINDS OF ALARM AND PERFORMANCE 6-1.Kinds of alarm There are two kinds of alarms such as gas alarm and trouble alarm. ・Gas alarm:When detection gas reaches preset alarm level or exceeds it, this starts alarm. 《Latched mode→Self reset after reset operation.》 *...
  • Page 29: 6-2-2.Counteraction At Gas Alarm

    6-2-3.Gas alarm possibility except the case of gas detection It may respond to interference gas For interference gases, contact the nearest agent or Riken Keiki. It may be caused by drifting due to the change of long time elapsed sensor Check the reading by daily check and make calibration through authorized service agent whenever necessary.
  • Page 30: 7.Maintenance Check

    To maintain the safety operation of the unit, it is recommended to keep the maintenance contract with service agent for regular maintenance, adjustment and overhaul etc including the gas sensitivity adjustment. For the detail of maintenance contract, contact our nearest service agent or Riken Keiki.
  • Page 31: 7-3.Replacement Of Sensor And Parts

    7-3.Replacement of sensor and parts. CAUTION For replacement of sensors and parts etc, contact our service agent or Riken Keiki. 7-4.Treatment when operation is stopped or its installation place is moved. 7-4-1.Stop at normal operation Make the power switch off on the front panel.
  • Page 32: 9.Treatment At Abnormal Case

    9-2.Counteraction at trouble alarm When any trouble was found on this unit and detector head, contact our nearest agent or Riken Keiki. 9-3.Before it is thought to be trouble It may be originated from disconnection and short circuit between units.
  • Page 33 ・Non-connection of sensor, contact failure of connector and disconnection of sensor cable. ・Check “whether sensor is connected” or “the sensor cable connector is connected to the sensor. If can not be recovered, contact our nearest agent or Riken Keiki. (5)Fuse disconnection <Performance>...
  • Page 34: 10.Definition Of Word

    10.DEFINITION OF WORD Calibration By using calibration (standard) gas etc, it seeks for instrument reading, display value or set value and true value. Maintenance mode When make maintenance of this unit, shut off the alarm and the output of signal showing the maintenance mode condition is displayed on the external output signal.
  • Page 35: 11-1.Specifications

    11.PRODUCT SPECIFICATIONS 11-1.Specifications Model designation: EC-592 Detection gas: Toxic gases Applicable detector head: Electrochemical sensor type detector head Detection signal: 4-20mADC Detector head cable: CVVS 2-core or equivalent Distance to the detector head: Max 1.25km with CVVS (1.25mm2) cable Output signal: 4-20mADC, non-isolation (Resistive load below 300Ω)
  • Page 36: 11-2.Product Composition

    Housing type for the case, front identification card type, non-explosion type. Dimensions & Weight: Approx 36(W) X 72(H) X 134(D) mm, Approx 0.1kg (unit only) * Specification subject to change without notice. 11-2.Product composition Standard compositions ・Indicator/alarm unit Model EC-592 ・Operating instruction manual…1 vol.