Maintenance; Configuration Method; Configuration With Display - Honeywell SLN 700 User Manual

Smartline level transmitter non-contact radar
Hide thumbs Also See for SLN 700:
Table of Contents

Advertisement

6.1 Configuration method

There are three debugging methods for SLN 700
1. Display/configuration module
2. Host PC
3. HART hand-held programmer

Configuration with Display

Display modular is a pluggable display/configuration tool. The configuration can be
done through operating with 4 buttons on Display. The language for the configuration
menu is optional. After exiting from configuration mode, display is only used for display
in general, and the measurement value can be seen clearly through the glass window.
Page 40
6

Maintenance

-Enter programming mode; -Confirm programming options;
-Confirm parameter modification.
[
]
Button
Modify parameter values
-
Shortcut keys
displays the frequency spectrum -Choose programming
[
]
options; -Choose the parameter bit to edit; -Display of
parameters.
[
]Button
-
Exit programming mode;
-Return to higher level menu.
SLN 700 SmartLine NCR Level Transmitter User's Manual
;
Revision 2

Advertisement

Table of Contents
loading

Table of Contents