Instruction manual Mega Star 600/600R Centrifuge Series ® Cat. no 76533-880 Mega Star 600 without rotor 76533-884 Mega Star 600 package 76533-882 Mega Star 600R without rotor 76533-886 Mega Star 600R package Version a Issue 06/2022...
Legal Address of Manufacturer North America Avantor One Radnor Corporate Center Building One, Suite 200 100 Matsonford Road Radnor, PA 19087 1 610 386 1700 Country of origin VWR Mega Star 600R: Germany VWR Mega Star 600: China vwr.com I Instruction manual VWR Mega Star 600/600R ®...
Page 3
vwr.com I Instruction manual VWR Mega Star 600/600R ®...
Page 4
vwr.com I Instruction manual VWR Mega Star 600/600R ®...
Table of Contents Legal Address of Manufacturer..................2 Warning .......................... 7 Safety Information ............................. 7 Set Up Conditions ............................. 7 Preparation ..............................8 Hazardous Substances ..........................8 Operating ..............................9 Maintenance .............................. 9 Package Contents ....................... 10 Delivery Content ............................10 Unpacking and Installation ..................
Page 6
System Menu ....................... 27 Flowchart System Menu .......................... 27 Troubleshooting ......................29 Mechanical emergency door release ...................... 29 Troubleshooting by Guide ........................31 When to contact customer service ......................32 Repair and maintenance ..................... 33 Maintenance and care ..........................33 User replaceable accessories and spare parts ............
Warning Before starting to use the centrifuge, read through this instruction manual carefully and follow the instructions. Failure to follow the instructions and safety information in this instruction manual will result in the expiration of the seller’s warranty. Safety Information WARNING Observe the safety instructions.
Preparation WARNING It is the obligation of the operator to make sure, that the proper protective clothing is • used. Mind the “Laboratory Biosafety Manual” of the World Health Organization (WHO) and the regulations in your country. • Do not make any changes to the mechanical components of the rotor. Do not touch the electronic components of the centrifuge nor alter any electronic or •...
If centrifuging any hazardous materials mind the “Laboratory Biosafety Manual” of the World Health Organization (WHO) and any local regulations. When centrifuging microbiological samples from the Risk Group II (according to the “Laboratory Biosafety Manual” of the World Health Organization (WHO)), aerosol-tight biological seals have to be used.
Package Contents Description Quantity VWR Mega Star 600 / 600R Centrifuge Power supply cable Instruction manual If any parts are missing, please contact the nearest VWR representative. Delivery Content Part No. Package 76533-884 VWR MEGA STAR 600 PKG 76533-886 VWR MEGA STAR 600R PKG Unpacking and Installation Before setting up 1.
Transporting WARNING Always lift the centrifuge from both sides. Never lift the centrifuge by its front panel, its back panel or at its door. Always remove the rotor before moving the centrifuge. Due to its weight, the centrifuge should be carried by at least two people. Always lift the centrifuge at both sides.
Mains Connection NOTICE Plug the centrifuge into grounded electrical sockets only. Power supply switch on the backside of the Power supply switch on the right side of the VWR Mega Star 600 VWR Mega Star 600R 1. Turn off the power supply switch. 2.
Intended use This centrifuge is a laboratory product used to separate substance mixtures of different densities. This centrifuge has to be operated by trained personnel only. Symbols and conventions This symbol refers to general hazards. CAUTION means that material damage could occur. WARNING means that injuries or material damage or contamination could occur.
Product Specifications Technical data VWR Mega Star 600 Environmental Conditions -Use in interior spaces -Altitudes of up to 2000 m above sea level -Max. relative humidity 80 % up to 31 °C; decreasing linearly to 50 % relative humidity at 40 °C Environmental Conditions during Storage Temperature: -10 °C to +50 °C and Shipping...
VWR Mega Star 600R Environmental Conditions -Use in interior spaces -Altitudes of up to 3000 m above sea level -Max. relative humidity 80 % up to 31 °C; decreasing linearly to 50 % relative humidity at 40 °C Environmental Conditions during Storage Temperature: -10 °C to +50 °C and Shipping Humidity: 15 % to 85 %...
Mains supply The following table contains an overview of the electrical connection data. This data is to be taken into consideration, when selecting the mains connection socket. Unit VWR Mega Star 600 VWR Mega Star 600R Mains Voltage 120 V ±10 % 120 V ±10 % Frequency 60 Hz...
Description of Buttons and Switches Control panel The control panel contains the keys and displays of the centrifuge (only the power switch is located on the right side (VWR Mega Star 600R) or on the backside (VWR Mega Star 600) of the device). VWR Mega Star 600 Function Display Contents...
VWR Mega Star 600R Function Display Contents Status The status of the centrifuge is displayed here. Speed / RCF Value The speed (rpm) or RCF value (x g) is displayed here. Running Time The running time is displayed here. Temperature The temperature is displayed here.
Getting Started and Operation NOTE Only display pictures of the VWR Mega Star 600R centrifuge are shown. The display of the VWR Mega Star 600 is the same, only the missing details for temperature. Switch on centrifuge Turn on the power switch of the device. The device performs a self-check of its software.
Page 20
Rotor Installation CAUTION Unapproved or incorrectly combined accessories can cause serious damage to the centrifuge. The centrifuge is equipped with an Auto-Lock-locking system. This system is used to automatically lock the rotor to the centrifuge spindle. The rotor does not have to be bolted on to the centrifuge spindle.
4. If available close the rotor with the rotor lid. Check the according rotor manual for further details on requirements and methods of closure. CAUTION Be sure to check all sealings before starting any aerosol-tight applications. See the information in the rotor instruction manual. 5.
Page 22
Explanation of RCF value The relative centrifugal force (RCF) is given as a multiple of the force of gravity g. It is a unitless numerical value which is used to compare the separation or sedimentation capacity of various centrifuges, since it is RCF = 11.18 x ( independent of the type of device.
Page 23
Preselecting the Temperature You can preselect temperatures between -10 °C and +40 °C. To set the temperature, proceed as follows: Press the TEMPERATURE arrow keys. This allows to change the set temperature using the arrow keys until the desired temperature is displayed. Temperature will change in steps of single degrees Celsius. The temperature cannot be adapted until the rotor has been positively identified;...
Programs The VWR Mega Star 600 Centrifuge is able to save up to 4 programs; the VWR Mega Star 600R is able to save up to 99 programs. It is only possible to save a program if the centrifuge is in standstill. Loading or saving of programs is not possible if the centrifuge is spinning.
Page 25
Use the formula: = admissible speed = maximum speed Once the rotor has been properly installed, the main switch turned on and the centrifuge door closed, you may start centrifuging. Use of Tubes and Consumables Care should be taken to ensure that the tubes and bottles used in the centrifuge are: •...
Stopping the Centrifuge Run With Pre-Set Running Time If the run time is preset, the centrifuge will run at the selected speed until the desired run time is reached. It will then automatically decelerate and stop. Once stopped “RUN COMPLETED” will be displayed and if selected the display will flash.
Aerosol-tight rotors For your protection when using an aerosol tight lid, the rotor should only be removed with the lid closed. NOTICE Rotors supplied with a lid for aerosol-tight applications come with a mandrel, which belongs to the Auto-Lock. Be sure not to place the lid onto this mandrel to prevent it from being damaged. CAUTION Skin can be pierced by sharp mandrel tip.
Page 28
vwr.com I Instruction manual VWR Mega Star 600/600R ®...
Troubleshooting Mechanical emergency door release During a power failure, you will not be able to open the centrifuge door with the regular electric door release. A mechanical override is provided to allow sample recovery in the case of an emergency. This is only to be used in emergencies and after the rotor has come to a complete stop.
Page 30
Below front side Mechanical Emergency Door Release on the downside 4. Push the cord back into the centrifuge and mount the plug. Reconnect the centrifuge once the power has been restored. Switch on the centrifuge. Press the OPEN key to have the centrifuge door locks operative again. vwr.com I Instruction manual VWR Mega Star 600/600R ®...
Troubleshooting by Guide If problems occur other than those listed in this table, the authorized customer service representative must be contacted. Error number Error message Troubleshooting E-002; Read Manual Restart the centrifuge. E-005; If the error message appears again, inform the service. E-008;...
Error number Error message Troubleshooting E-019 Rotor Unknown Restart the centrifuge. Check, if the rotor is qualified for the VWR Mega Star 600 / 600R Centrifuge. If the error message appears again, inform the service. E-025; Read Manual Check, if the centrifuge door is blocked. E-027 Restart the centrifuge.
Repair and maintenance Maintenance and care Cleaning intervals For the sake of personal, environmental and material protection, it is your duty to clean and if necessary, disinfect the centrifuge on a regular basis. Maintenance Recommended interval Rotor Chamber (Bowl) Daily or when polluted Rotor Daily or when polluted Accessories...
Page 34
Plastic Parts Check for signs plastic crazing, fading, bruising or cracking. CAUTION Do not run any rotor or accessories with sign of damage. Ensure that the rotor, buckets and accessories are within the service life and number of cycles. It is recommended that you have rotors and accessories inspected yearly as part of your routine service to ensure safety.
Page 35
Cleaning the Filter Mat It is recommended that you clean the filter mat regularly every six weeks. Depending on the environmental conditions it may be necessary to clean it more often. How to clean the filter mat: 1. Unscrew the ventilation grid placed on the right side of the centrifuge. 2.
Page 36
Decontamination WARNING Radiation is possible when touching the contaminated rotor and centrifuge parts. Radioactive material can get into the centrifuge when a tube breaks or as a result of spills. In case of contamination, make sure that others are not put at risk. Decontaminate the affected parts immediately.
User replaceable accessories and spare parts Shipping and Disposal WARNING When removing the centrifuge and accessories from use for disposal you have to clean and if necessary, disinfect or decontaminate the entire system. In doubt contact the VWR customer service. For the disposal of the centrifuge mind the regulations in your country.
Warranty warrants that this product will be free from defects in material and workmanship for a period of two (2) years from date of delivery. If a defect is present, VWR will, at its option and cost, repair, replace, or refund the purchase price of this product to the customer, provided it is returned during the warranty period.
WEEE Conformity This product is subject to the regulations of the EU Waste Electrical & Electronic Equipment (WEEE) Directive 2012/19/EU. It is marked by the following symbol: This equipment is marked with the crossed out wheeled bin symbol to indicate that this equipment must not be disposed of with unsorted waste.
Page 44
CHEMICAL Trisodium Phosphate Tris Buffer (neutral Triton X/100™ Urea Hydrogen Peroxide (10 %) Hydrogen Peroxide (3 %) Xylene Zinc Chloride Zinc Sulfate Citric Acid (10 %) ¹ Polyethlyeneterephtalate S – Satisfactory. M – Moderate attack, may be satisfactory for use in centrifuge depending on length of exposure, speed involved, etc.;...
Page 45
North America Avantor One Radnor Corporate Center Building One, Suite 200 100 Matsonford Road Radnor, PA 19087 1 610 386 1700 vwr.com I Instruction manual VWR Mega Star 600/600R ®...