After reading, please keep this instruction manual properly so that you can refer to it at any time when you need it. Product Name : Anesthesia Machine Model: TA60V Structure This product consists of an anesthesia machine, a vaporizer, a stand a breathing circuit and accessories. Composition:...
Warranty and Repair Service The warranty period of the purchased product is subject to the sales contract. Consumables: Refers to disposable consumables that need to be replaced after each use or fragile materials that are replaced regularly. Consumables have no warranty. The warranty period starts from the "Installation date"...
Foreword Explanation This manual describes the use, function and operation of the product in detail. Before using this product, please read and understand the contents of this manual carefully to ensure the correct use of this product and the safety of animals and operators. This manual introduces this product according to the most complete configuration, so some contents may not be applicable to the purchased product.
1.0 Safety 1.1 Safety Information WARNING — Indicates a potentially hazardous or unsafe operation which, if not avoided, could result in death or serious personal injury or property damage. CAUTION — Indicates a potentially hazardous or unsafe operation, which, if not avoided, could result in minor personal injury, product failure, damage, or property damage.
Warning: Do not use the anesthesia machine when there is a leak in the respiratory system Warning: Check the passive AGSS processing system specifications as well as the specifications of the anesthesia machine to ensure compatibility and prevent mismatched processing systems. Warning:...
Page 9
Caution: Please install or transfer the machine properly to prevent the machine from falling, bumping, being damaged by strong vibration or other mechanical external forces Caution: When the trolley is configured, the anesthesia machine can be tilted 10° to maintain stability. Please do not hang objects on both sides of the anesthesia machine in order to avoid the danger of dumping.
Caution:Please place this manual near the machine so that it can be easily and promptly obtained when needed. Caution : This manual introduces this product according to the most complete configuration, and the purchased product may not have some configurations or functions. 1.2 Device Symbol Symbol Description...
Page 11
Manufacturing date avoid rain Atmospheric face up pressure limit Humidity Limit temperature limit Fragile handle Stacking layer limit Not for use in a magnetic resonance (MRI) environment This product contains some harmful substances, which can be used with confidence within the environmental protection use period. After the environmental protection use period is exceeded, it should enter the recycling system.
2.0 Overview 2.1 System Overview Please refer to the "3.0 Installation Guide" of this manual for the installation of the anesthesia machine。 2.1.1 Intended Use This anesthesia machine is used to provide inhalation anesthesia and respiratory support during animal anesthesia. Warning:...
2.2 Device View 2.2.1 Front side view of the anaesthesia machine 2-1 Front side view of main unit - 10 -...
Page 14
Serial Part Description Vaporizer A device that can effectively output the concentration of anesthetics. Each vaporizer is suitable for a specific anesthetic. Vaporizer base It is used to install the vaporizer. After the vaporizer is removed, it can also ensure the normal ventilation of the air circuit.
2.2.2 Breathing circuit view Image 2-2 Breathing circuit view Serial Part Description Inspiratory check valve The status of the inspiratory check valve can be observation window observed from outside the device. Expiratory check valve The status of the expiratory check valve can be observed viewing window from outside the device Inspiratory port...
2.2.3 Anaesthesia machine rear view(with stand) Picture 2-3 Anaesthesia machine rear view Serial Parts Description Hook For hanging breathing circuit or other accessories AGSS Port for exhaust absorber canister Used to control the installation of CO2 absorber canister. lock and unlock handle Exhaust gas recovery line Connection line for connecting the AGSS port to the exhaust gas tank...
2.2.4 Anaesthesia machine rear view(Wall) Picture 2-4 Anaesthesia machine rear view(Wall) Serial Part Description Wall hook When the main unit is wall mounted on the wall, hang on the wall bracket Wall support pad When the host is wall mounted on the wall, support the host so that the host is perpendicular to the wall Table 2-3 Anaesthesia machine rear view (wall mounted) parts list - 14 -...
3.0 Installation Guide Warning: Disposable items such as single-use breathing hoses, soda lime may be considered potentially biohazardous and should not be reused. Dispose of these items in compliance with hospital and local regulations for contaminants and biohazards This manual introduces this product according to the most complete Caution:...
Anesthesia vaporizer mounted to anesthesia machine Serial Item Vaporizer base Connector Locking knob Vaporizer (1) Align mounting hole evaporator with the connector on the base of the evaporator, and then install the evaporator on the bottom of the evaporator. (2) Rotate the locking knob on the top of the Image 3-3Vaporizer mounted to evaporator clockwise, and the rotation anesthesia machine diagram...
Serial Part Description CO2 Absorber Canister Vessel for filling soda lime, absorbing CO2 Absorber Canister Used to control the installation of CO2 absorber canister. Lock and Unlock Handle Absorber Canister When rotating the canister disassembly knob, the support Support will rise or fall 3.4.1 CO2 absorber Canister installation (1) Make sure that the CO2 absorber canister is filled with qualified soda lime, and the filling amount does not exceed the MAX marking line。...
Page 21
Image 3-5 Schematic diagram of the waste gas canister and the waste gas recycling pipe before installation Serial Part Description AGSS Port for discharging waste gas Waste gas recycling pipe Connection line for connecting the AGSS port to the waste gas canister Waste gas canister Used to absorb anesthetic gas from AGSS waste gas...
3.5.2 Dismantling of waste gas canister and waste gas recycling pipe The disassembly of the waste gas canister and the waste gas recycling pipe also refers to the replacement of the waste gas canister. The operation steps are as follows: (1) Pull out the joint at both ends of the waste gas recycling pipe。...
Page 23
Wall-mounted anesthesia machine: Hang the upper wall hooks on the back of the anesthesia machine shown in Figure 2-4 (with the wall accessories installed on the anesthesia machine) in the installation slot of the main machine wall bracket shown in Figure 3-6。 Install the waste gas canister on the bracket as shown in Figure 3-6 Connecting the waste gas recycling pipe - 20 -...
4.0 Test before operation 4.1 Pre-Operation Testing Requirements The anesthesia machine is pre-operated according to the test intervals listed below. Please refer to the specific steps or precautions in this manual. Pre-operational testing should be performed in the following cases: ...
Check that the vaporizer contains enough anesthetic. Check whether the dosing port of the vaporizer is tightly closed。 For narcotic absorbers, check that they have gained more than the claimed weight gain. Before each administration of anesthesia, the following should be performed: Inspect the machine for damage or dangerous conditions.
8. Check the color of the soda lime in the canister. If the color change is noticeable, replace the soda lime immediately 9. Appropriate narcotic drugs and emergency drugs are available 10. The casters are not damaged or loose, and they are locked so that the anesthesia machine cannot be moved 11.
Make sure the breathing system is intact and connected correctly Make sure the check valve on the breathing system is working properly If the inspiratory check valve opens during inspiration and closes at the beginning of exhalation, the inspiratory check valve is working properly. If the expiratory check valve opens during exhalation and closes at the beginning of inspiration, the expiratory check valve is working properly.
5.0 Operation 5.1 Oxygen Flow Setting Connect the oxygen source and make sure the air source has sufficient pressure. The flow of oxygen in the input fresh gas can be controlled by the knob of the oxygen flowmeter, and the flowmeter displays the flow value of oxygen in the fresh gas. 5.2 Set Up the Vaporizer Press and turn the concentration control knob on the vaporizer to set the concentration of the anesthetic to the appropriate position.
Page 29
Warning: Only add the specified anesthetics to the vaporizer. Please check the name and color of the anesthetic on the vaporizer and anesthetic bottle before use. Isoflurane is purple, sevoflurane is yellow, and enflurane is orange. Warning: If the wrong anesthetic or other substance is added to the vaporizer, immediately stop using it to prevent health hazards.
Warning: If the vaporizer is not turned back to the "0" position, there may be a large amount of anesthetic vaporized and spilled Caution: After rotating the dial to the "0" position, please wait at least 5s before turning on the anesthesia vaporizer to balance the pressure and prevent the overflow of fresh gas and anesthetic vapors.
bottom of the dosing system Slowly turn counterclockwise to open the doser cap Drain the anaesthetic until no anaesthetic is visible in the sight glass and no anaesthetic continues to flow into the bottle. If necessary, close the dispensing knob in time and continue to discharge with another anesthetic bottle After the anesthetic is emptied, close the dispensing knob clockwise, close the cap of the anesthetic bottle, and tighten the dosing cap.
6.0 Maintenance 6.1 Maintenance Timetable This schedule is based on the minimum maintenance times specified for the typical use of 2000 hours per year. If the actual use time per year is longer than the typical situation, the maintenance work of the equipment should be more frequent Caution:...
Caution: Do not use abrasive cleaners such as steel wool, silver polish or cleaners. The pH value of the cleaning solution must be between 7.0 and 10.5 Name Category Water Detergent Soapy water (PH value is 7.0~10.5)) Detergent Alcohol (75%) Moderate disinfectant Table 6-2 Cleaners and Disinfectants 6.2.1 Wipe...
7.0 Appendix Caution: Use only the accessories specified in this chapter, using other accessories may cause equipment failure Disposable accessories can only be used once, repeated use may result in Caution: performance degradation or cross-contamination Do not use the accessory if you see signs of damage to the accessory Caution:...
Page 35
total) 38-000067-00 Waste gas recovery tank 22 Outer cone joint recovery tank - 32 -...
A.0 Product Specification A.1 Safety Specification Anti-shock rating BF Application Part Inlet protection level IPX1 Disinfection and Equipment with a method of disinfection and sterilization sterilization method recommended by the manufacturer Explosion protection No explosion protection (common equipment), no flammable class anesthetics Mobile level...
A.4 Air System Specifications Gas source Gas type Oxygen Air source pressure range 280-600kpa/40-87psi Interface Type NIST/DISS Flowmeter Flowmeter adjustment range 0-4L/min Table A-4 Air System Specifications A.5 Respiratory System Specifications Port Manual bladder Coaxial 22mm (OD)/15mm (ID) conical joint interface Inspiratory Valve Port Coaxial 22mm (OD)/15mm (ID) conical joint...
A.7 No Power AGSS No power AGSS Port Type 30mm Conical joint (external) Waste gas canister bracket Support configuration The diameter of waste gas canister is not more than 80mm Weight capacity Table A-7 AGSS - 35 -...