Indications; Contraindications; Precautions - QED Shape Coil Operator's Manual

Table of Contents

Advertisement

Symbol

3.2 Indications

The Shape Coil is intended for use with Canon 1.5T or 3.0T MR systems to produce diagnostic
images of general human anatomy that can be interpreted by a trained physician.

3.3 Contraindications

None.

3.4 Precautions

Patients with increased likelihood of seizures or claustrophobia may require special
care. Consult the MRI system operation manual.
Patients who are unconscious, heavily sedated, or in a confused mental state are at
increased risk of burn injury because they may not be able to notify the operator of heat
or pain due to excessive heating and tissue damage.
8 |
P a g e
Number
Standard
N/A
EN50419
EU2012/18/EU
5.1.8
ISO 15223-1
5.1.9
ISO 15223-1
N/A
N/A
Title, Meaning
The use of this symbol indicates that this
product should not be treated as
household waste.
By ensuring that this product is disposed of
correctly, you will help prevent potential
negative consequences for the
environment and human health, which
could otherwise be caused by
inappropriate waste handling of this
product.
For more detailed information concerning
the return and recycling of this product,
please consult the supplier from whom you
purchased the product.
Importer
Distributor
Separate the RF coil from the gantry inner
wall by at least 50 mm. Burn injuries may
result due to the electric field that is
generated in the RF coil when a high-
frequency magnetic field is transmitted.
6000921 Rev. 1