QED Shape Coil Operator's Manual page 15

Table of Contents

Advertisement

(4) Secure the coil to the phantoms using the belt supplied with the system. Place the belt in a
position that is not at the center of the coil so that the center of the coil is visible for
alignment with the positioning projector.
(5) Adjust the position of the coil so that the positioning projector beam is aligned with the
center of the coil and then send the coil to the center of the gantry. Using the couch
operation button, move the couchtop to a position where the couchtop position indicator
on the gantry operating panel reads 200.
(6) Register the patient. Enter 170 cm for the height and 60 kg for the weight.
(7) Select "Typical PAS" → "Coil QA" and click [Other]. Select the following sequences in the
PAS "Other" field.
Sequence name
15 |
P a g e
Required / Not required
FE_slt
FE_map
SNR
Required
Required
Required
6000921 Rev. 1