Measurement Procedure - Hanna Instruments HI98713 Instruction Manual

Iso portable turbidimeter
Hide thumbs Also See for HI98713:
Table of Contents

Advertisement

When taking any turbidity measurements several basic rules should be followed:
• Use always cuvettes without scratches or cracks because they can cause inaccurate readings.
• Cap always the cuvettes to avoid spillage of the sample into the instrument.
• Close always the lid of the instrument during measurement.
• Keep the lid of the instrument closed when it is not used to prevent dust or dirt entering.
• Put always the instrument on a flat, rugged surface when taking measurements.
• Do not operate in direct sunlight.
• Do not use too much oil to prevent contamination of the optical system.
To take turbidity measurements, follow the next steps:
• Turn the instrument ON by pressing ON/OFF.
When dashes are displayed on the LCD, the
instrument is ready. On the secondary LCD the
current time appears, if selected in SETUP menu.
• Fill a clean, dry cuvette with 10 mL of sample up
to the mark, taking care to handle the cuvette
by the top.
• Replace the cap.
• Wipe the cuvette thoroughly with a lint‑free
cloth to remove any fingerprints, dirt or water
spots.
• Apply silicone oil on the cuvette and wipe with
a lint‑free cloth to obtain an even film over the
entire surface of the cuvette.
Note: It is very important to oil the cuvette, especial‑
ly for low turbidity values (< 1 FNU) to hide
the glass imperfections that can influence
the reading.
• Place the cuvette into the instrument. Align
the mark from the cuvette with the sign on the
instrument case and close the lid.
Note: If you have a cuvette with orientation mark,
place the cuvette into the instrument with
the orientation mark aligned with the sign
on the instrument top.

MEASUREMENT PROCEDURE

19

Advertisement

Table of Contents
loading

Table of Contents