Prior Queensgate NANOSCAN SP Series Quick Start Instructions

Prior Queensgate NANOSCAN SP Series Quick Start Instructions

Advertisement

Quick Links

NANOSCAN SP SERIES QUICK START INSTRUCTIONS
ATTEMPTING TO USE THE NANOSCAN SP SERIES SYSTEM
IDENTIFYING YOUR SYSTEM COMPONENTS
The standard NanoScan SP Series system consist of:
(1) NanoScan SP Series Piezo Sample Scanner
(1) NPC-D-6110 Controller and power supply
One or more of the sample accessories listed:
QGSP301XR - extra recessed microtitre plate
QGSP302XR - extra recessed universal specimen
QGSP303XR - extra recessed slide holder
(1) Adapter plate to mount the NanoScan SP on the microscope XY stage.
No adapter plate is require for Prior XY stages H117 & HLD117 (Except HLD117NN).
(12) M4x8 Mounting screws
(1) USB Cable QGPA0096-SK-102 + (1) 1.2 metre BNC to BNC cable QGBNCCAB+
(1) RS232C QG-0494-SK-200002
(1) 9 pin gender changer for RS232C QG-0494-SK-200001
(1) USB stick containing manuals and test report.
Note: The stage is a moving platform, the height will vary when powered compared to unpowered. The stage must be mounted
to a flat surface before being powered. Do not connect the stage to a controller until properly mounted.
STEP 1:
Where an adapter plate is
included, attach it to your XY stage
using the screws provided. Do not
screw the adapter plate for the
HLD117NN in place.
Use all available thread holes to
mount the plate. Lightly nip each screw with a hex key (0.3Nm /
300cNm/ 0.2lbft torque). In the event that the screw bottoms out
before the head seats, STOP. Do not force the screw as it may
be too long; back the screw out before moving the XY stage.
Shown with base
plate attached to back
of NanoScan SP unit.
PLEASE READ THE INSTRUCTIONS BEFORE
INSTALLING YOUR SYSTEM
SP400/600 Piezo Sample Scanner
STEP 2:
Position the NanoScan SP
stage aligning the
connection points with
those on the XY stage or
adapter plate. Mount the
stage using M4 screws
using all the available
thread holes on the XY
stage. Where the attach-
ment requires screws of
nonstandard length they
should be part of the
adapter kit. Lightly
tighten each screw with a hex hey (0.3Nm / 300cNm/
0.2lbft Torque). In the event that the screw bottoms out
before the head seats, STOP. Do not force the screw as it
may be too long; back the screw out before moving the XY
stage. It is important to attach the stage using as many
mounts at possible to achieve the specified performance.
www.prior.com
Sample Holders
Gender changer for
Adapter Plate
RS232C

Advertisement

loading
Need help?

Need help?

Do you have a question about the Queensgate NANOSCAN SP Series and is the answer not in the manual?

Questions and answers

Summary of Contents for Prior Queensgate NANOSCAN SP Series

  • Page 1 Sample Holders (1) Adapter plate to mount the NanoScan SP on the microscope XY stage. No adapter plate is require for Prior XY stages H117 & HLD117 (Except HLD117NN). (12) M4x8 Mounting screws (1) USB Cable QGPA0096-SK-102 + (1) 1.2 metre BNC to BNC cable QGBNCCAB+...
  • Page 2: Installing Your System

    It is important that the controller tuning is selected for the sample load to protect the stage. If on power-on the stage resonates/ vibrates, quickly power-down the system. Check the mounting of the stage and tuning (see below). If the stage continues to vibrate, please contact Prior for assistance. www.prior.com...
  • Page 3 CONNECTING YOUR SYSTEM - FRONT OF UNIT IN POS Indicator LED COMS Indicator LED POS MON connector Indicates the status of communications with the Indicates the status of the stage position in CLOSED Analogue position monitor output loop mode. connected computer. BNC connector(s) OFF = Stage has not reached the desired position.
  • Page 4 Start Nanobench - the screen displayed will be similar to the one show below: Select the correct COM port within Nanobench 6000, then click ‘Connect’: At this point the software should connect to the controller and begin reading the live measured position. www.prior.com...
  • Page 5: External Control

    • The cable is sensitive - do not lift the stage by the cable or bend beyond the 30mm radius. External Control The QGOP-400-UP/QGOP-400-INV/QGOP-400UP-HL and QGOP-400-INV-HL can be controlled from Prior ProScan, Micromanager, Metamorph and Ni Elements. Quickstart quides are available.