Advertisement
LCD
Measurement results, prism temperature, and remaining battery charge are displayed. The displayed value is an example.
START button
Press to take measurements and hold dow n to turn off the display.
Battery compartment
Place and remove batteries from here.
Sample stage
Apply water and samples on the glass prism located in the center of the sample stage.
ZERO button
Press to perform zero-setting.
Thank you for purchasing the instrument. Carefully read and follow all instructions. Keep this manual for future reference.
Read and follow all safety instructions before operating the instrument. Failure to comply with the following instructions may result in personal injury or property damage.
<International Protection Classification IP65>
<Chemical Resistance of Body Case>
1,2-Dichloroethane/Dichloromethane
Solvents that are harmful to the plastic body case include but not limited to the above substances.
<LCD Auto Shut-off>
The instrument will turn itself off after 2 minutes of inactivity. To manually turn it off, hold down the START button for more than 2 seconds.
<For oily/fatty samples>
Try stirring the sample on the prism while measuring to improve the repeatability of oily/fatty samples.
Wipe off the sample, rinse with water, and wipe off the water to clean the sample stage thoroughly. Dry the sample stage completely with dry tissues.
For oily samples:
Clean oily residues with ethyl alcohol or mild soap, and then, rinse with water.
This unit converts refractive index into TDS (total dissolved solids) for convenient concentration measurements of cutting oil.
The relationship between the concentration of cutting oil and its refractive index differs with each type of oil.
Therefore, to determine the concentration of each type of oil, each solution should have its own custom conversion table to compensate for the value displayed by instrument.
Conversion method (in the case of water based solutions)
<How to calculate the Conversion Factor>
E.g.) For an oil solution, with a concentration of 5% (dilution ratio at 20 times), mix 95ml of water to 5ml of undiluted solution.
For this example, assume the displayed value, on the instrument, is 4.0.
Next, convert the displayed value using the following relationship to determine the conversion factor; Actual concentration÷Displayed value, Ex: (5.0÷4.0 = 1.25 ).
Once the conversion factor in known (1.25 in this example), measuring a sample with an unknown concentration is simple.
Taking an unknown sample, assume that the measured value is 3.0.
Using the simple formula, displayed value × conversion factor (3.0×1.25) the actual concentration of the sampled measured is 3.75.
<Offset feature use>
Program a conversion factor to automatically display adjusted measurement
In the example above, the measurement values multiplied by the factor of 1.25 will be displayed.
y = ax + b
y: Actual concentration
x: PAL-102S readings
a: Coefficient (multiplication)
Range: 0.01 to 10.00(Factory default 1.00)
b: Addition/subtraction number
Range: -10.00 to 10.00(Factory default 0.00)
The following messages alert the user when an operation has failed.
The battery is low.
The ZERO button was pressed with nothing or something other than water on the prism.
The START button was pressed with nothing or an insufficient amount of sample on the prism.
The sample measured outside the measurement range.
The prism temperature is below the temperature range.
The prism temperature is above the temperature range.
The instrument is faulty. (Replace the batteries. Contact ATAGO if this error persists.)
Too much light is entering the prism, and the instrument cannot measure accurately. (Shade the sample stage with your hand and take a measurement again.)
The readings are corrected, based on the temperature of the prism, within the automatic temperature compensation range.
Store the instrument in a dry place away from direct sunlight. Exposure to humidity and heat may damage the instrument.
Do not use organic solvents (paint thinner, benzene, gasoline, etc.) on the plastic body case.
Clean and dry the sample stage thoroughly, following the "Cleaning" instructions. Store the unit away from direct sunlight at a stable temperature with as little fluctuation as possible.
Measurement range | Cutting Oil 0.0 to 70.0 Temperature 10.0 to 75.0℃ | |
Resolution | Cutting Oil 0.1 | Temperature 0.1℃ |
Accuracy | Cutting Oil ±0.2 | Temperature ±1℃ |
Automatic temperature compensation range | 10 to 75℃ | |
Ambient temperature range | 10 to 40℃ |
Sample volume | At least 0.3ml |
Measurement time | Approx. 3 seconds |
Backlight | The backlight stays on for 30 seconds after any button is pressed. |
Power supply | Two (2) AAA alkaline batteries |
Battery life | Approx. 11,000 measurements (when using alkaline batteries) |
International Protection class | IP65 |
Dimensions and weight | 55 (W) x 31 (D) x 109 (H)mm, 100g (main unit only) |
Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.
Download Atago Pocket PAL-102S - Cutting Oil Refractometer Manual
Advertisement
Need help?
Do you have a question about the Pocket PAL-102S and is the answer not in the manual?
Questions and answers