Monitor Relay Programming - Agilent Technologies Turbo-V 1000HT Instruction Manual

Pcb controller
Table of Contents

Advertisement

Enter the
selection
CURRENT pushbutton, and the display shows:
F R O N T
/
R E M O T E
S E L E C T
I O N : X X X X X X
where: XXXXXX = means the word FRONT or
REMOTE depending on the last selection.
The controller is factory-set for FRONT panel
operation.
Choose the desired selection by pressing the
CYCLE pushbutton.
F R O N T
/
R E M O T E
S E L E C T
I O N :
Confirm the selection by pressing the PUMP
CURRENT pushbutton.
In this way you enter into an operating phase
named "Monitor Relay Programming" described in
the following paragraph.
by pressing
the
PUMP
F R O N T

Monitor Relay Programming

The display shows:
S P E E D
S E L E C T
where: XXKRPM = is the switch point of relay R1 at
the preset turbopump speed, adjustable from 00 to
99 KRPM.
The speed threshold will condition the R1 and R2
operation (see the following cycle diagram) and it is
factory-set to 36 KRPM.
Press the CYCLE NUMBER pushbutton to select
the first number.
S P E E D
S E L E C T
85
TECHNICAL INFORMATION
T H R E S H O L D
I O N : X X K R P M
T H R E S H O L D
I O N : X X K R P M
87-900-906-01 (E)