Controller-to-Pump Connection

To connect the controller to the pump, an appropriate
interconnection cable must be prepared to connect the
P10 connector of the controller with the pump
connector.
The following figure shows the pump output connector
configuration where pins:
− A-F = pump temperature sensor
− B-C-D = 54 Vac 3-phase output to pump motor
stator
− E = ground.
P10
CONTROLLER
PHASE R
1
6
PHASE S
7
2
PHASE T
8
3
PUMP TEMP . SENSOR
4
9
PUMP TEMP . SENSOR
10 5
GND
SCREEN
Controller to pump cable
F
PUMP
75
TECHNICAL INFORMATION
G
A
TURBO-V
H
B
PUMP
E
C
CONNECTUR
D
87-900-906-01 (E)