Download Print this page

General; Remounting - Siemens MOBILETT Plus Manual

Hide thumbs Also See for MOBILETT Plus:

Advertisement

Remounting collimator

General

Consult the document "Service Instructions", Chapter 9 for details about settings and
adjustments.

Remounting

Install the collimator. Center the rotary-flange so that the screws marked "E" are
centered in the grooves before tightening.
Connect the wires to connection terminal board "D".
cable tie left side
Fig. 1
Siemens-Elema AB
Solna, Sweden
D
E
E
SPR8-220.092.02
Rev. 01
Page 1 of 2 MOBILETT Plus/ -E/ -M/ -HP
08.01