Scan the Patient ID

Open a new SmartGuard® (Ex. 6).
Gently squeeze the SmartGuard side tabs to position onto the NPi-200.
There will be an audible click when the SmartGuard is properly
positioned (Ex. 7).
For the first patient use, in order to properly input the patient ID into the SmartGuard,
select either Barcode Scanner or Manual ID to indicate the patient ID entry method
used (Ex. 8).
Pairing the NPi-200 to the Barcode Scanner
Connect the Barcode Scanner and Charging Cradle to the power
supply and plug into a power outlet. Turn on the Barcode Scanner until an audible
beep is heard and a blue light on the device flashes. Position the Barcode Scanner
next to the NPi-200.
On the NPi-200, select Barcode Scanner. The NPi-200 will
display "Connecting..." on the touchscreen (Ex. 9). Once
successfully paired, the touchscreen will prompt when the
device is ready to scan the patient ID barcode (Ex. 10).
The patient ID will now appear on the NPi-200 touchscreen. Confirm the
patient information is correct and select Accept (Ex. 11).
The NPi-200 will display the patient ID number and read "Ready to scan" (Ex. 12).
Manual Entry of the Patient ID
Press Manual Entry. Using the touchscreen, press the Patient ID.
Select Shift to toggle from alpha to numeric (Ex. 13 & 14) as
required. When the patient ID number has been manually
entered, check for accuracy and press Enter.
NeurOptics® NPI®-200 Pupillometer System—Instructions for Use © 2022 NeurOptics, Inc.