Installation - Panasonic MDF-U76VA Operating Instructions Manual

Ultra-low temperature freezer
Hide thumbs Also See for MDF-U76VA:
Table of Contents

Advertisement

INSTALLATION

1. Removing the packaging materials and tapes
Remove all transportation packaging materials and tapes. Open the doors and ventilate the unit. If the
outside panels are dirty, clean them with a diluted neutral dishwashing detergent. (Undiluted detergent
can damage the plastic components. For the dilution, refer to the instruction of the detergent.) After the
cleaning with the diluted detergent, always wipe it off with a wet cloth. Then wipe off the panels with a dry
cloth.
Note:
Remove the cable tie banding the power supply cord.
Prolonged banding may cause the corrosion of the
cord coating.
2. Adjusting the leveling foot
Extend
the
leveling
counterclockwise to contact them to the floor.
Ensure the unit is level. (Fig. 1)
3. Fixing the unit
Two fixtures are attached to the rear of the frame.
Fix the frame to the wall with these fixtures and rope
or chain. (Fig. 2)
4. Power source
For stable and reliable cooling operation, check the
following items before connecting the freezer to a
power source. (Fig. 3)
■ Voltage of the power source: 115 V, 60 Hz
■ Receptacle of the power source: NEMA reference
5-20R
(The equipment has a NEMA 5-20P plug.)
If not, it is necessary to install 5-20R receptacle by
qualified engineers.
■ Current capacity of the power source: 20 amperes
■ Rating amperage of the circuit breaker or fuse used in
the power line: 20 amperes
(Use item delay type in case of fuse.)
■ Do not connect the other equipment t
4. Ground (earth)
The ground (earth) is for preventing the electric shock in the case of the electrical insulation is somehow
degraded. Always ground the unit at the time of installation.
WARNING
Use a power supply outlet with ground (earth) to prevent electric shock. If the power supply outlet is
not grounded, it is necessary to install a ground by qualified engineers.
Never ground the unit through a gas pipe, water main, telephone line or lightning rod.
grounding may cause electric shock in the case of an incomplete circuit.
5. Installing the earth leakage breakers
This product is to connect a earth leakage breaker to the power supply side of the product.
Contact our sales representative or agent at the time of installation of the
feet
by
rotating
them
13
Leveling foot
Fixture
5-20P
5-20R
Fig. 3
breaker.
earth leakage
Fig.1
Fig. 2
Such

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents