Page 2
Operation concept a0012790-en a0012791-en Fig. 1: Pressing the soft key: selecting the menu directly Fig. 2: Turning the navigator: moving the cursor in the menu a0012792-en a0012793-en Fig. 3: Pressing the navigator: launching a function Fig. 4: Turning the navigator: selecting a value (e.g. from a list) a0012794-en a0012795-en Fig.
Month (two-digit): 01 to 12 Year (four-digit): 1970 to 2106 Set time Depends on the format Editing mode: hh (hour): 00 to 23 / 0 am to 12 pm mm (minutes): 00 to 59 ss (seconds): 00 to 59 Endress+Hauser...
Diagnostics menu • Enabled Factory setting Disabled Calibration menu Factory setting Enabled Hold delay 0 to 600 s The hold is maintained for the duration of the delay time when you switch to the measuring mode. Factory setting Endress+Hauser...
Logbook name Customized text, 20 characters Source of data Options Select the input that should be the data source of the logbook • None entries. • Channel 1 • Channel 2 • Controller 1 • Controller 2 Factory setting None Endress+Hauser...
Page 8
Factory setting Line plotter Menu to define the graphic display Axes Options Should the axes (x, y) be displayed (On) or not (Off)? • Off • On Factory setting Orientation Options • Horizontal • Vertical Factory setting Horizontal Endress+Hauser...
Path: ... /Extended setup/Diagnostics settings/Diag. behavior Function Options Info List of diagnostic messages Select the message to be changed. Only then can you make the settings for this message. Endress+Hauser...
Page 10
If Multidrop is active (Bus address > 0), the current at current output 1 is always set to 4 mA. Here, it does not matter what function has been assigned to the output (measured value/controller etc.). Current simulation is no longer possible. Endress+Hauser...
Page 11
• Restoring a tried-and-tested setup, e.g. if you have changed a lot of settings and no longer know what the original settings were Insert the SD card into the controller card reader. Go to: Menu/Setup/General settings/Extended setup/Data management/Save setup. Assign a file name (Name). Then select "Save". Endress+Hauser...
Page 12
Install the hardware or load the new software. Go to: Menu/Setup/General settings/Extended setup/Data management/Activation code. Enter the activation code which you received with the hardware or software. Then select "Confirm". --> Your new hardware or software is now activated and you can now be configured. Endress+Hauser...
The amperometric oxygen sensor COS51D cannot be sterilized. For this reason, you will not be able to define limit values for sterilization in the diagnostics settings for this sensor. On the other hand, these menu items are displayed for a sterilizable amperometric sensor, such as COS22D. Endress+Hauser...
• It is also possible to save a setup and transfer it to another device (see "Data management" in the "General settings" section). This function might be a better option than performing a configuration when a sensor is not connected. Endress+Hauser...
The damping causes a floating average curve of the measured values over the time specified. Path: Menu/Setup/Inputs/<Sensor type> Function Options Info Damping <Sensor type> 0 to 300 s You specify the damping of the main measured value of the connected sensor and that of the integrated temperature Factory setting Damping temp. sensor. Endress+Hauser...
You can balance out this effect with the "Medium compensation" function. Internal buffer pH 0 to 14 Only change the value if you are using a sensor with an (only pH glass) internal buffer other than pH 7. Factory setting pH 7.00 Endress+Hauser...
Page 17
You define the cleaning programs in the "Setup/Additional functions/Cleaning" menu. 5.2.4 Calibration settings Stability criteria You define the permitted measured value fluctuation which must not be exceeded in a certain timeframe during calibration. If the permitted difference is exceeded, calibration is not permitted and is aborted automatically. Endress+Hauser...
Page 18
• Manual Automatic (only pH glass) Factory setting The device recognizes the buffer automatically. The Fixed recognition depends on the setting for "Buffer manufacturer". Manual You enter any two buffer values. These must differ in terms of their pH value. Endress+Hauser...
Page 19
Options Info Buffer manufacturer Options Temperature tables are stored internally in the unit for the • Endress+Hauser following pH values: • Ingold/Mettler • Endress+Hauser • DIN 19266 2,00 / 4,00 / 7,00 / (9,00) / 9,20 / 10,00 / 12,00 •...
Page 21
The main reasons for a deteriorating electrode status are: • Glass membrane blocked or dry • Diaphragm (reference) blocked Remedial action ► Clean or regenerate the sensor. ► Replace the sensor if this does not have the desired effect. Endress+Hauser...
Page 22
Specify your limit values for monitoring the measured value. Upper warning limit Factory setting Diagnostics code and associated message text: 900 mV 942 "Process value" Lower warning limit Factory setting Diagnostics code and associated message text: -900 mV 943 "Process value" Endress+Hauser...
Page 23
Diagnostics code and associated message text: 904 "Process check" Function Options • On • Off Factory setting Duration 1 to 240 min The measured value must change during this time. Otherwise the error message is triggered. Factory setting 60 min Endress+Hauser...
Page 25
Path: Menu/Setup/Inputs/pH/Extended setup/Diagnostics settings Function Options Info Delta zero point (pH glass) pH glass Specify your limit values for monitoring the zero point or Delta operating point (pH pH 0.00 to 2.00 operating point differential. ISFET) pH ISFET 0 to 950 mV Endress+Hauser...
Page 26
Path: ... /Extended setup/Diagnostics settings/Diag. behavior Function Options Info List of diagnostic messages Select the message to be changed. Only then can you make the settings for this message. Diag. code Read only Endress+Hauser...
Page 27
5.2.6 Tag control With this function, you specify which sensors are accepted at your device Path: Menu/Setup/Inputs/<Sensor type>/Extended setup Function Options Info Tag control Additional information on the display: tag control currently used Endress+Hauser...
Page 28
Here you can restore the factory settings for the sensor input. For this purpose, simply press the navigator button and select "OK" when the prompt for the device software appears. Only the factory settings for this particular input are restored. All other settings remain unchanged. Endress+Hauser...
Factory setting Damping temp. sensor. 6.1.3 Manual hold Path: Menu/Setup/Inputs/<Sensor type> Function Options Info Manual hold Options • On You can use this function to set the channel manually to • Off "Hold". Factory setting No hold Endress+Hauser...
Page 30
If distances from the wall are smaller, the installation factor is bigger for electrically insulating pipes (f >1), and smaller for electrically conductive pipes (f < 1). It can be measured using calibration solutions, or a close approximation determined from the following diagram. Endress+Hauser...
Page 31
NaOH 0..15% Temp. comp. mode Options Only select "without temp. comp" in very small temperature (Operating • with temp. comp ranges. mode=Concentration) • without temp. comp In all other cases, select "with temp. comp". Factory setting with temp. comp Endress+Hauser...
Page 32
Since there are no options for concentration measurement, (Operating mode=Resistance) • mS/cm / MΩcm this function is not displayed for such measurements. • S/cm / kΩcm • μS/m / kΩm • mS/m / Ωm • S/m / Ωcm Factory setting Auto / Auto Endress+Hauser...
Page 33
(23.0 to 212.0 ˚F) temperature-compensated conductivity Factory setting The alpha coefficients and alpha reference temperatures of 25.0 ˚C (77.0 ˚F) Endress+Hauser calibration solutions can be found in the documentation enclosed. Temp. source Options Decide how you want to compensate the medium •...
Page 34
• Conductivity You specify the temperature, conductivity and • Coeff. Alpha uncompensated conductivity. Recommended for large measuring ranges and small measured values. Factory setting Coeff. Alpha Conductivity As the value pairs, you specify an alpha value and the related temperature. Endress+Hauser...
• A diagnostics message is present at the channel and • Cleaning 2 • A cleaning process has been specified for this message • Cleaning 3 (--> "Inputs/Diagnostics settings/Diag. behavior"). • Cleaning 4 Factory setting None You define the cleaning programs in the "Setup/Additional functions/Cleaning" menu. Endress+Hauser...
Page 36
► Switch off the controller and switch it back on again. Path: Menu/Setup/Inputs/<Sensor type>/Extended setup/Diagnostics settings Function Options Info Process Check System Diagnostics code and associated message text: 904 "Process check" Function Options • On • Off Factory setting Endress+Hauser...
Page 37
Operation > 80°C Warning limit Factory setting Diagnostics code and associated message text: 10000 h 193 "Operating time" Operation > 120°C Only conductive sensors Warning limit Factory setting Diagnostics code and associated message text: 10000 h 195 "Operating time" Endress+Hauser...
Page 38
This branch, along with the same functions, can be found in different "Extended setup" menus. The list of diagnostic messages displayed depends on the path selected. There are device-specific messages, and messages that depend on what sensor is connected. Endress+Hauser...
Page 39
You can define the cleaning programs under: • Cleaning 2 Menu/Setup/Additional functions/Cleaning. • Cleaning 3 • Cleaning 4 Factory setting None Detail information Read only Here you can find more information on the diagnostic message and instructions on how to resolve the problem. Endress+Hauser...
Page 40
Factory setting measuring point, and only accepts the sensors that have the EH_CM44_ same tag. Group Numerical Factory setting "Tag" stands for the name of a measuring point, and is used in many areas of process measuring technology. Endress+Hauser...
Page 41
Here you can restore the sensor factory settings. For this purpose, simply press the navigator button and select "OK" when the prompt for the device software appears. Only the factory settings for the sensor are restored. The settings for the input remain unchanged. Endress+Hauser...
The damping causes a floating average curve of the measured values over the time specified. Path: Menu/Setup/Inputs/<Sensor type> Function Options Info Damping <Sensor type> 0 to 300 s You specify the damping of the main measured value of the connected sensor and that of the integrated temperature Factory setting Damping temp. sensor. Endress+Hauser...
Path: Menu/Setup/Inputs/DO/Extended setup Function Options Info Temp. compensation Options Decide how you want to compensate the medium • Automatic temperature: • Manual • Automatically using the temperature sensor of your sensor Factory setting • Manually by entering the medium temperature Automatic Endress+Hauser...
Page 44
Specify the number of decimal places for displaying the main • #.# measured value. • #.## • #.### • # Factory setting 7.2.3 Medium compensation (in the process) Path: Menu/Setup/Inputs/DO/Extended setup Function Options Info Medium pressure Options • Process pressure • Air pressure Factory setting Air pressure Endress+Hauser...
Page 45
You define the cleaning programs in the "Setup/Additional functions/Cleaning" menu. 7.2.5 Calibration settings Stability criteria You define the permitted measured value fluctuation which must not be exceeded in a certain timeframe during calibration. If the permitted difference is exceeded, calibration is not permitted and is aborted automatically. Endress+Hauser...
Page 46
The pressure is then 500 to 1200 hPa independent of the altitude. Medium pressure="Process pressure" 500 to 9999 hPa Factory setting 1013 hPa Rel. hum. (air variable) 0 ... 100 % Factory setting 100 % Endress+Hauser...
Page 47
The associated diagnostics code is displayed for every setting. Slope The (relative) slope characterizes the sensor condition. Decreasing values indicate electrolyte exhaustion. You can control when the electrolyte should be replaced by specifying limit values and the diagnostics messages these limit values trigger. Endress+Hauser...
Page 48
Path: Menu/Setup/Inputs/DO/Extended setup/Diagnostics settings Function Options Info Zero point 0.0 to 10.0 nA Specify the limit values for zero point monitoring in your sensor. Warning limit Factory setting Diagnostics code and associated message text: 3.0 nA 513 "Zero Warn" Endress+Hauser...
Page 49
Diagnostics code and associated message text: 108 "Sterilization" Calibrations, cap (only amperometric sensors) The calibration counters in the sensor make a distinction between sensor calibrations and calibrations with the membrane cap currently used. If this cap is replaced, only the (cap) counter is reset. Endress+Hauser...
Page 50
Function Options The number depends heavily on the process and must be • On determined individually. • Off Factory setting Warning limit Factory setting Diagnostics code and associated message text: 109 "Sterilizat. cap" Endress+Hauser...
Page 51
Diagnostics code and associated message text: 904 "Process check" Function Options • On • Off Factory setting Duration 1 to 240 min The measured value must change during this time. Otherwise the error message is triggered. Factory setting 60 min Endress+Hauser...
Page 52
Operation > 25°C Only optical sensors Warning limit Factory setting Diagnostics code and associated message text: 10000 h 190 "Operating time" Operation > 40°C Warning limit Factory setting Diagnostics code and associated message text: 10000 h 192 "Operating time" Endress+Hauser...
Page 53
Diagnostic message Options You can deactivate or reactivate a diagnostic message here. • On Deactivating means: • Off • No error message in the measuring mode Factory setting • No error current at the current output Depends on the message Endress+Hauser...
Page 54
7.2.7 Tag control With this function, you specify which sensors are accepted at your device Path: Menu/Setup/Inputs/<Sensor type>/Extended setup Function Options Info Tag control Additional information on the display: tag control currently used Endress+Hauser...
Page 55
Here you can restore the sensor factory settings. For this purpose, simply press the navigator button and select "OK" when the prompt for the device software appears. Only the factory settings for the sensor are restored. The settings for the input remain unchanged. Endress+Hauser...
The damping causes a floating average curve of the measured values over the time specified. Path: Menu/Setup/Inputs/<Sensor type> Function Options Info Damping <Sensor type> 0 to 300 s You specify the damping of the main measured value of the connected sensor and that of the integrated temperature Factory setting Damping temp. sensor. Endress+Hauser...
Select how many decimal places should be used to display • #.# the temperature. • #.## Factory setting Main value format Options Specify the number of decimal places for displaying the main • #.# measured value. • #.## • #.### • # Factory setting Endress+Hauser...
Page 58
• Automatically using the temperature sensor of your sensor Factory setting • Manually by entering the medium temperature Automatic Medium temperature -5.0 to 50.0 ˚C Enter the temperature of your medium. (Temp. compensation=Manual) (23.0 to 122.0 ˚F) Factory setting 20,0 ˚C (68 ˚F) Endress+Hauser...
Page 59
Calib. expiration date Warning limit Factory setting Diagnostics message: 105 "Calib. validity" 11 months Warning and alarm limits mutually affect each other's possible ranges for adjustment. Endress+Hauser...
Page 60
The associated diagnostics code is displayed for every setting. Slope The (relative) slope characterizes the sensor condition. Decreasing values indicate electrolyte exhaustion. You can control when the electrolyte should be replaced by specifying limit values and the diagnostics messages these limit values trigger. Endress+Hauser...
Page 61
Path: Menu/Setup/Inputs/Chlorine/Extended setup/Diagnostics settings Function Options Info Zero point 0.0 to 3.2 nA Specify the limit values for zero point monitoring in your sensor. Warning limit Factory setting Diagnostics code and associated message text: 2.0 nA 513 "Zero Warn" Endress+Hauser...
Page 62
Factory setting Warning limit Factory setting Diagnostics code and associated message text: 535 "Sensor check" Number of sensor calibrations The calibration counters in the sensor make a distinction between sensor calibrations and calibrations with the membrane cap currently used. Endress+Hauser...
Page 63
• Contaminated sensor, or sensor in air • Sensor failure • Process error (e.g. through control system) Remedial action ► Clean the sensor. ► Check the measuring chain. ► Switch off the controller and switch it back on again. Endress+Hauser...
Page 64
Factory setting No diagnostics messages. However, the time the sensor operates under extreme conditions is recorded in the sensor and can be read in the sensor information in the diagnostics menu. Endress+Hauser...
Page 65
It does not take the penetration of acids and bases into the sensor into account. Depending on the application it can be necessary to change the electrolyte before a load of 5 As is reached. Endress+Hauser...
Page 66
Options The messages are divided into different error classes in • Maintenance (M) accordance with NAMUR NE 107. • Out of specification (S) --> BA445C "Maintenance&diagnostics" • Function check (C) • Failure (F) Factory setting Depends on the message Endress+Hauser...
Page 67
Factory setting measuring point, and only accepts the sensors that have the EH_CM44_ same tag. Group Numerical Factory setting "Tag" stands for the name of a measuring point, and is used in many areas of process measuring technology. Endress+Hauser...
Page 68
Here you can restore the factory settings for the sensor input. For this purpose, simply press the navigator button and select "OK" when the prompt for the device software appears. Only the factory settings for this particular input are restored. All other settings remain unchanged. Endress+Hauser...
Select a saved calibration data record • Formazin The options and the factory setting depend on the setting in • Kaolin "Application type". • SiO2 • TiO2 • Excess sludge • Digested sludge • Activated sludge Factory setting Formazin or Excess sludge Endress+Hauser...
• Off "Hold". Factory setting No hold Extended setup 9.2.1 Measured value formats Path: Menu/Setup/Inputs/Turbidity/Extended setup Function Options Info Temperature format Options Select how many decimal places should be used to display • #.# the temperature. • #.## Factory setting Endress+Hauser...
Page 71
Calibration timer and calibration expiration date You can specify the calibration interval for the sensor here. Once the time configured elapses, the "Calibration timer" diagnostics message appears on the display. The timer is reset automatically if you recalibrate the sensor. Endress+Hauser...
Page 72
Delta turbidity 0,1 ... 5,0 % Permitted measured value fluctuation during calibration Factory setting Delta SAC 2,0 % Delta temperature 0.10 to 2.00 ˚C Permitted temperature fluctuation during calibration 0.18 to 3.60 ˚F Factory setting 0.50 ˚C 0.90 ˚F Endress+Hauser...
Page 73
Operating time Total operating time of the sensor Warning limit Factory setting Diagnostics code and associated message text: 10000 h 199 "Operating time" (Operation below specified temperature limit) Endress+Hauser...
Page 74
942 "Process value" Filter change Only CAS51D Warning limit Factory setting Diagnostics code and associated message text: 10000 h 157 "Filter change" Lamp life Only CAS51D-SAC Warning limit Factory setting Diagnostics code and associated message text: 35040 h 171 "Lamp change" Endress+Hauser...
Page 75
Diagnostics code and associated message text: 904 "Process check" Function Options • On • Off Factory setting Duration 1 to 240 min The measured value must change during this time. Otherwise the error message is triggered. Factory setting 60 min Endress+Hauser...
Page 76
Options The messages are divided into different error classes in • Maintenance (M) accordance with NAMUR NE 107. • Out of specification (S) --> BA445C "Maintenance&diagnostics" • Function check (C) • Failure (F) Factory setting Depends on the message Endress+Hauser...
Page 77
Factory setting measuring point, and only accepts the sensors that have the EH_CM44_ same tag. Group Numerical Factory setting "Tag" stands for the name of a measuring point, and is used in many areas of process measuring technology. Endress+Hauser...
Page 78
Here you can restore the sensor factory settings. For this purpose, simply press the navigator button and select "OK" when the prompt for the device software appears. Only the factory settings for the sensor are restored. The settings for the input remain unchanged. Endress+Hauser...
Preselection for saved calibration data records • SAC • Transm. • Absorption • COD • TOC • DOC • BOD Factory setting Application Options Select a saved calibration data record • Factory calib. • 6 other data records Factory setting Factory calib. Endress+Hauser...
"Hold". Factory setting No hold 10.2 Extended setup 10.2.1 Measured value formats Path: Menu/Setup/Inputs/SAC/Extended setup Function Options Info Temperature format Options Select how many decimal places should be used to display • #.# the temperature. • #.## Factory setting Endress+Hauser...
Page 81
Calibration timer and calibration expiration date You can specify the calibration interval for the sensor here. Once the time configured elapses, the "Calibration timer" diagnostics message appears on the display. The timer is reset automatically if you recalibrate the sensor. Endress+Hauser...
Page 82
Delta turbidity 0,1 ... 5,0 % Permitted measured value fluctuation during calibration Factory setting Delta SAC 2,0 % Delta temperature 0.10 to 2.00 ˚C Permitted temperature fluctuation during calibration 0.18 to 3.60 ˚F Factory setting 0.50 ˚C 0.90 ˚F Endress+Hauser...
Page 83
Operating time Total operating time of the sensor Warning limit Factory setting Diagnostics code and associated message text: 10000 h 199 "Operating time" (Operation below specified temperature limit) Endress+Hauser...
Page 84
942 "Process value" Filter change Only CAS51D Warning limit Factory setting Diagnostics code and associated message text: 10000 h 157 "Filter change" Lamp life Only CAS51D-SAC Warning limit Factory setting Diagnostics code and associated message text: 35040 h 171 "Lamp change" Endress+Hauser...
Page 85
Diagnostics code and associated message text: 904 "Process check" Function Options • On • Off Factory setting Duration 1 to 240 min The measured value must change during this time. Otherwise the error message is triggered. Factory setting 60 min Endress+Hauser...
Page 86
Options The messages are divided into different error classes in • Maintenance (M) accordance with NAMUR NE 107. • Out of specification (S) --> BA445C "Maintenance&diagnostics" • Function check (C) • Failure (F) Factory setting Depends on the message Endress+Hauser...
Page 87
Factory setting measuring point, and only accepts the sensors that have the EH_CM44_ same tag. Group Numerical Factory setting "Tag" stands for the name of a measuring point, and is used in many areas of process measuring technology. Endress+Hauser...
Page 88
Here you can restore the sensor factory settings. For this purpose, simply press the navigator button and select "OK" when the prompt for the device software appears. Only the factory settings for the sensor are restored. The settings for the input remain unchanged. Endress+Hauser...
The damping causes a floating average curve of the measured values over the time specified. Path: Menu/Setup/Inputs/<Sensor type> Function Options Info Damping <Sensor type> 0 to 300 s You specify the damping of the main measured value of the connected sensor and that of the integrated temperature Factory setting Damping temp. sensor. Endress+Hauser...
Specify the number of decimal places. • #.# • #.## • #.### • # Factory setting Unit Options Select the unit for the main measured value. • mg/l NO3-N • mg/l NO3 • ppm NO3-N • ppm NO3 Factory setting mg/l NO3-N Endress+Hauser...
Page 91
Calib. expiration date Warning limit Factory setting Diagnostics message: 105 "Calib. validity" 11 months Warning and alarm limits mutually affect each other's possible ranges for adjustment. Endress+Hauser...
Page 92
Each sensor has a limited life expectancy which heavily depends on the operating conditions. If you specify warning limits for operating times under extreme conditions, you can guarantee the operation of your measuring point without any downtime by performing maintenance tasks in time. Endress+Hauser...
Page 93
942 "Process value" Filter change Only CAS51D Warning limit Factory setting Diagnostics code and associated message text: 10000 h 157 "Filter change" Lamp life Only CAS51D-SAC Warning limit Factory setting Diagnostics code and associated message text: 35040 h 171 "Lamp change" Endress+Hauser...
Page 94
Diagnostics code and associated message text: 904 "Process check" Function Options • On • Off Factory setting Duration 1 to 240 min The measured value must change during this time. Otherwise the error message is triggered. Factory setting 60 min Endress+Hauser...
Page 95
Options The messages are divided into different error classes in • Maintenance (M) accordance with NAMUR NE 107. • Out of specification (S) --> BA445C "Maintenance&diagnostics" • Function check (C) • Failure (F) Factory setting Depends on the message Endress+Hauser...
Page 96
Factory setting measuring point, and only accepts the sensors that have the EH_CM44_ same tag. Group Numerical Factory setting "Tag" stands for the name of a measuring point, and is used in many areas of process measuring technology. Endress+Hauser...
Page 97
Here you can restore the sensor factory settings. For this purpose, simply press the navigator button and select "OK" when the prompt for the device software appears. Only the factory settings for the sensor are restored. The settings for the input remain unchanged. Endress+Hauser...
12.1.3 Damping of the temperature value The damping causes a floating average curve of the measured values over the time specified. Path: Menu/Setup/Inputs/ISE Function Options Info Damping temp. 0 to 300 s Specify the damping for the temperature measurement. Factory setting Endress+Hauser...
This branch, along with the same functions, can be found in different "Extended setup" menus. The list of diagnostic messages displayed depends on the path selected. There are device-specific messages, and messages that depend on what sensor is connected. Endress+Hauser...
Page 100
You can define the cleaning programs under: • Cleaning 2 Menu/Setup/Additional functions/Cleaning. • Cleaning 3 • Cleaning 4 Factory setting None Detail information Read only Here you can find more information on the diagnostic message and instructions on how to resolve the problem. Endress+Hauser...
► You can complete and assign the other 3 slots as you prefer. ► Specify the measured variable that should be output. No options can be selected for pH which is why the "Measured variable" function is not available for this parameter. Endress+Hauser...
Page 102
Path: Menu/Setup/Inputs/ISE/Extended setup Function Options Info Main value format Options Specify the number of decimal places for the measured • # variable of the electrode slot. • #.# • #.## Factory setting #.## Endress+Hauser...
Page 103
The coefficients are empirical values. Factory setting -2.00 (chloride) -0.85 (potassium) Mode Options The standard setting (-) corrects a measured value that is too • + high as a result of the effect of interference ions. • - Factory setting Endress+Hauser...
Page 104
You enter any two buffer values. These must differ in terms of their pH value. Buffer manufacturer Options Temperature tables are stored internally in the unit for the • Endress+Hauser following pH values: • Ingold/Mettler • Endress+Hauser • DIN 19266 2,00 / 4,00 / 7,00 / (9,00) / 9,20 / 10,00 / 12,00 •...
Page 105
102, appears on 1000 h the display. Diagnostics settings This menu branch is used for specifying warning limits, and for defining whether and how diagnostics tools should be used. The associated diagnostics code is displayed for every setting. Endress+Hauser...
Page 106
Diagnostics code and associated message text: 904 "Process check" Function Options • On • Off Factory setting Duration 1 to 240 min The measured value must change during this time. Otherwise the error message is triggered. Factory setting 60 min Endress+Hauser...
Page 107
The difference is an indicator for the condition of the sensor. The greater the change, the greater the wear experienced by the pH-sensitive glass membrane as a result of chemical corrosion or abrasion. Endress+Hauser...
Each sensor has a limited life expectancy which heavily depends on the operating conditions. If you specify warning limits for operating times under extreme conditions, you can guarantee the operation of your measuring point without any downtime by performing maintenance tasks in time. Endress+Hauser...
Type of measurement that should be displayed and • Interface level calculated: • Interface range Interface level Factory setting Distance from the basin floor to the boundary area Interface level Interface range Distance from the water line to the boundary area Endress+Hauser...
The values can be as high as 60 if the sludge/water transition is relatively "soft". If you require significantly higher gain values, this is an indication of overranging. It is then difficult or impossible to reliably evaluate the echo signal. Endress+Hauser...
5 ... 30 Restricts the gain in automatic mode in order to prevent system overload. Factory setting Gain increment 0,1 ... 0,5 Defines how quickly the gain can adapt to changing process conditions in the automatic mode. Factory setting Bottom definition Endress+Hauser...
Page 113
Speed at which the interface window opens Factory setting 13.5.3 Diagnostics settings This menu branch is used for specifying warning limits, and for defining whether and how diagnostics tools should be used. The associated diagnostics code is displayed for every setting. Endress+Hauser...
Page 114
• Relay 1 to n (depends on assign the message to an output (Menu/Setup/Outputs, the device version) assign "Diagnostics" function and set Operating mode to "as Factory setting assigned"). None One alarm relay is always available, regardless of the device version. Other relays are optional. Endress+Hauser...
Page 115
Here you can restore the factory settings for the sensor input. For this purpose, simply press the navigator button and select "OK" when the prompt for the device software appears. Only the factory settings for this particular input are restored. All other settings remain unchanged. Endress+Hauser...
You can output the entire measuring range of the "Measured settings depend on: value" or just some of it at the current output. For this Range upper value "Measured value" purpose, specify the start and end of the range in accordance with your requirements. Endress+Hauser...
Page 117
• Raw value μs (only Oxygen (opt.)) Cond i Options • Temperature Cond c • Conductivity • Resistance Chlorine Options • Temperature • Current • Concentration Options • Temperature • pH • Ammonium • Nitrate • Potassium • Chloride Endress+Hauser...
Page 118
Only the two unipolar manipulated variable parts are available for selection for outputting to modulated relays. If outputting the values via a current output, you also have the possibility of outputting the bipolar manipulated variable y to one current output only (split range). Endress+Hauser...
Page 120
PWM=pulse width modulation • PWM PFM=pulse frequency modulation • PFM Factory setting Types of modulation: PWM (pulse width modulation): The duty cycle is varied within a period T (T=t ). The cycle duration remains constant. Typical application: solenoid valve Endress+Hauser...
Page 121
You can output two categories of diagnostics messages via the relay: Diagnostics messages from one of the 4 Namur classes (--> BA445C "Maintenance&Diagnostics" for more information on the Namur classes) Diagnostics messages which you have individually assigned to the relay output Endress+Hauser...
"Outputs/Current outputs" section. Please refer to this section for more information. ► Please note that if you select Hold behavior="Freeze" , the system not only flags the status but also "freezes" the measured value. Endress+Hauser...
• Temperature Options • Temperature • ORP mV • ORP % Oxygen (amp.) Options • Temperature Oxygen (opt.) • Partial pressure • Concentration liquid • Saturation • Raw value nA (only Oxygen (amp.)) • Raw value μs (only Oxygen (opt.)) Endress+Hauser...
Page 125
All the mathematic functions can also be used as a data source and the calculated value can be used as the measured value. You can monitor the manipulated variable by assigning the controller manipulated variable to a limit switch (e.g. configure a dosing time alarm). Endress+Hauser...
Page 126
Depends on the measured switch to pick up, become closer or move further apart. It is value needed to ensure a stable switching behavior. Start delay 0 to 9999 s Synonyms: pick-up and drop-out delay Factory setting Switch off delay Endress+Hauser...
The neutral zone is used to suppress control deviations (e) that are too small. The control deviation thus filtered now powers the actual PID controller which is divided into three parts (top-down) based on the Endress+Hauser...
Page 128
First configure the controller and leave the switch at "Off" • Off during this time. • Automatic Once you have made all the settings, you can assign the • Manual mode controller to an output and switch it on. Factory setting Endress+Hauser...
Page 129
A mixture of both process types can often be found in practice: the semi-batch process. Depending on the ratio between the flow and the container volume, this arrangement behaves either like a batch process or an in-line process. Endress+Hauser...
Page 130
With the "Direct" effective direction, the controller acts as a "downwards controller". It becomes active when the process value (e.g. the temperature) is too high (range B). The red curve shows overlap between the curves of the two 1-sided controllers. Endress+Hauser...
Page 131
The integral action time is the time needed in a step-function response to achieve a change in the manipulated variable - as a result of the I effect - which has the same magnitude as the P-value. e = control deviation, e=w-x (setpoint controlled variable) Endress+Hauser...
Page 132
Depends on the Source of data As soon as X deviates more than X from the set point, Y balance="Symmetric" reaches 100%. Xp Low Range of adjustment and Process balance="Asymmetric" factory setting Xp High Depends on the Source of data Endress+Hauser...
Page 133
Manipulated variable is set to 0 (no dosing) State Options Internal controller status • Freeze Freeze • Reset No change Reset Factory setting After a hold, the control system starts from scratch, and Freeze settling time takes place again. Endress+Hauser...
00-00:01 to 07-00:00 The interval value can be between 1 minute and 7 days. (DD-hh:mm) Example: You have set the value "01-00:00". Each day, the Factory setting cleaning cycle starts at the same time you started the first 01-00:00 cleaning cycle. Endress+Hauser...
• Displayed in user-defined measuring menus. 15.4.1 Difference You can subtract the measured values of two sensors and use the result to detect incorrect measurements, for example. To calculate a difference, you must use two measured values with the same engineering unit. Endress+Hauser...
Page 136
Measured value Y3 (optional) Measured value Deviation control Options You can monitor the redundancy. Specify an absolute limit • Off value that may not be exceeded. • On Factory setting Deviation limit Depends on the selected measured value Endress+Hauser...
Page 137
• If carbon dioxide is regarded as non-corrosive, the live steam can be directed to the turbine far earlier during startup. • If the conductivity value increases during normal operation, users can immediately determine an ingress of coolant or air by calculating the degassed conductivity. Endress+Hauser...
Page 138
Measured value (Outlet, e.g. sensor downstream from the ion exchanger). Outlet Measured value Main value format Options • Auto • # • #.# • #.## • #.### Factory setting Auto Endress+Hauser...
Page 139
Outlet Sensor downstream from the cation exchanger, "acid conductivity" The choice of measured value is obsolete since it must always be "Conductivity". Calculated pH Read only View this value in a user-defined measuring screen or via the current output. Endress+Hauser...
Online configuration competes with onsite operation, i.e. each of the two options blocks the other one. On both sides it is possible to disallow access from the other side. A video showing an example of how to set up a Fieldcare connection to CM442 is available on the Endress+Hauser...
► Establish the connection via your HART device. ► Operate Liquiline via the HART device. To do so, follow the instructions in the related manual. All the information on HART communication with Liquiline is provided on the CD (LIT-18.doc). Endress+Hauser...