Page 1
Model P2016 Dual Channel Handheld Digital Storage Oscilloscope & Multimeter User Manual...
Page 2
The information in this manual was correct at the time of printing. However, Berkeley Nucleonics will continue to improve products and reserves the rights to change specification at any time without notice. BNC is the registered trademark of Berkeley Nucleonics. Berkeley Nucleonics Corporation 2955 Kerner Blvd, San Rafael CA 94901 Phone: 415-453-9955, Fax: 415-453-9956 Web: www.berkeleynucleonics.com...
Page 3
General Warranty BERKELEY NUCLEONICS CORPORATION warrants that the product will be free from defects in materials and workmanship for a period of 3 years from the date of purchase of the product by the original purchaser from the BERKELEY NUCLEONICS Company. The warranty period of accessories such as probe, battery, adapter is 12 months.
BNC Model P2016 Handheld DSO & DMM Contents Table of Contents 1. DECLARATION OF CONFORMITY ··································· 1 1.1 Declaration of Conformity ··········································································· 1 1.2 Package Contents ······················································································ 2 2. SAFETY INFORMATION ················································· 3 2.1 Safety Symbols and Terms ········································································· 3 2.1.1 Safety Symbols ·····················································································...
Page 5
BNC Model P2016 Handheld DSO & DMM Contents 6.11 Freezing the Screen ··············································································· 24 6.12 Using Average for Smoothing Waveforms ················································ 25 6.13 Using Persistence to Display Waveforms ·················································· 25 6.14 Using Peak Detection to Display Glitches ················································· 26 6.15 Selecting AC-coupling ············································································ 27 6.16 Reversing the Polarity of the Displayed Waveform ·····································...
Page 6
BNC Model P2016 Handheld DSO & DMM Contents 8.7.2 Persistence ························································································ 60 8.7.3 XY Mode ···························································································· 60 8.7.4 Cymometer ························································································· 61 8.8 Waveform Saving Setups ·········································································· 62 8.8.1 Waveform Saving Setups in normal mode ················································· 62 8.8.2 Waveform Saving Setups in FFT mode ···················································· 63 8.9 Function Setting Menu ·············································································...
BNC Model P2016 Handheld DSO & DMM 1-Declaration of Conformity 1. Declaration of Conformity 1.1 Declaration of Conformity Model P2016 Handheld Digital Storage Oscilloscope & Multimeter Statement of Conformity Based on results using appropriate standards, the product is in conformity with Electromagnetic Compatibility Directive 2004/108/EC...
BNC Model P2016 Handheld DSO & DMM 1-Declaration of Conformity 1.2 Package Contents See the picture below. This picture is only for your reference. If any discrepancy found between certain picture and product, please look the actual product. Description Standard Optional ●...
BNC Model P2016 Handheld DSO & DMM 2-Safety information 2. Safety Information In order to ensure the correct using and the best efficient service, please carefully read this user's manual. 2.1 Safety Symbols and Terms 2.1.1 Safety Symbols These symbols may appear in this manual or on the instrument.
BNC Model P2016 Handheld DSO & DMM 2-Safety information 2.2 General Safety Information Carefully read the following safety information in order to avoid any personal injury and damage on this product or any products connected with it. This product can only be used in the specified applications to prevent any possible dangers.
Page 11
1:1 test leads because the probe tip voltage will be directly transmitted to the Oscilloscope & Multimeter. ● Do not use exposed metal BNC or banana plug connectors. ● Do not insert metal objects into connectors. ● Always use the Oscilloscope & Multimeter only in the manner specified.
Page 12
BNC Model P2016 Handheld DSO & DMM 2-Safety information conditions. No operation is allowed under a humid environment. No operation is allowed under an explosive environment. Keep clean and dry on the product surface. Using the Oscilloscope & Multimeter in a manner not specified may impair the protection provided by the equipment.
BNC Model P2016 Handheld DSO & DMM 3-General Characteristics 3. General Characteristics Oscilloscope 2 in 1 (Multimeter function support); Record length of 6,000 points for each channel; Reading-out with the cursor; Twenty automatic measurement functions; ...
4. Performing a General Inspection 4.1 Performing a general inspection When you receive have a new Model P2016 oscilloscope, it is suggested that you should perform a general inspection on the instrument according to the following steps. 4.1.1 Check whether there is any damage due to transportation...
BNC Model P2016 Handheld DSO & DMM 5-Input Connections 5. Input Connections 5.1 Input Connections 5.1.1 Input connections See the following Figure 2: Figure 2:Input connections Description: 1. The power adapter is supplied for AC power supply and battery recharging.
BNC Model P2016 Handheld DSO & DMM 5-Input Connections The output terminal of 1 kHz/5 V square-wave test signal. 5.1.2 The connection of 1 kHz/5 V Square-wave test signal At the left side of the oscilloscope is a port for testing 1 kHz/5 V square-wave signal which is used to adjust the probe, shown as Figure 3.
Page 17
BNC Model P2016 Handheld DSO & DMM 5-Input Connections Figure 4:Front Panel Description: 1. AC adapter Port 2. RS-232C Port. 3. USB Port. Berkeley Nucleonics Corporation info@berkeleynucleonics.com | www.berkeleynucleonics.com 2955 Kerner Blvd, San Rafael, CA 94901 | 800-234-7858...
Page 18
BNC Model P2016 Handheld DSO & DMM 5-Input Connections 4. USB Mass Storage Port. 5. Power switch. 6. F1 - F5: Switch or Adjust options for each menu. 7. AUTO SET: Under DSO mode, automatically selects the horizontal scales, vertical scale, and trigger level according to the input signal.
Page 19
BNC Model P2016 Handheld DSO & DMM 5-Input Connections 25. ◄ (yellow): Oscilloscope left-direction adjustment key. 26. A: Selects DMM current measurement 27. V: Selects DMM voltage measurement 28. R: Selects DMM impedance, diode, continuity and capacitance measurement 29. SET: Convert AC and DC during measure current or voltage in Multi-meter; convert resistance, diode, on-off and capacitance measure during resistance measuring.
BNC Model P2016 Handheld DSO & DMM 6-Using the Oscilloscope 6. Using the Oscilloscope 6.1 About this Chapter This chapter provides a step-by-step introduction to the scope functions. The introduction does not cover all the capabilities of the scope functions but gives basic examples to show how to use the menus and perform basic operations.
Page 21
BNC Model P2016 Handheld DSO & DMM 6-Using the Oscilloscope Figure 5: Oscilloscope Operation Window Description: Battery electric quantity indicating symbols. Auto measurement window 1, in which "f "means frequency, "T" means cycle, "V" means the average value, "Vp" the peak-peak value, "Vk" the root-mean-square value, "Ma"...
BNC Model P2016 Handheld DSO & DMM 6-Using the Oscilloscope indicates AC, the graphic "-" indicates DC, the graphic " " indicates GND. 14. This reading shows the vertical Voltage Unit Scale of CH2. 15. These graphics show the coupling mode of CH1, among which the graphic "~"...
BNC Model P2016 Handheld DSO & DMM 6-Using the Oscilloscope Figure 6: The Tool's Menus 6.5 Manually Setting the Vertical System, Horizontal System and Trigger Position 6.5.1 Setting the Vertical System To set the vertical system of CH1: (1) Press VOLTS POSITION (red) key to toggle between bellows displayed at the bottom left of the screen: ▲/▼...
BNC Model P2016 Handheld DSO & DMM 6-Using the Oscilloscope (2) When "▲/▼ – CH2 Zero" is displayed, ▲(blue) and ▼(blue) can be used to adjust the vertical position of CH2, the relevant information will be displayed; (3) When "▲/▼ – CH2 Volts/Div" is displayed, ▲ (blue) and ▼ (blue) can be used to adjust the voltage scale of CH2, the relevant information will be displayed.
Page 25
BNC Model P2016 Handheld DSO & DMM 6-Using the Oscilloscope (yellow) to adjust trigger level position, press R key to "zero" the trigger level position. Press OPTION again and left bottom side display as ◄/► – Time ▲/▼ – Trig...
Page 26
BNC Model P2016 Handheld DSO & DMM 6-Using the Oscilloscope Figure 9: Adjust Time and Trig 2 Press ◄ (yellow) or ► (yellow) to adjust time base horizontal position, press V key to "zero" the horizontal position; press ▲(yellow) or ▼(yellow) to adjust trigger level position in Channel 2 (or Channel 1), press R key to "zero"...
BNC Model P2016 Handheld DSO & DMM 6-Using the Oscilloscope Vertical scale factor: It stands for the voltage amplitude represented by a division in the vertical direction of the display area, through the adjustment of which you can amplify or attenuate the signal and thus regulate the signal amplitude into the expected measurement range.
Look at the bottom and the right of the Oscilloscope. The Oscilloscope has six signal inputs: two safety BNC jack inputs (CH1 and CH2) for scope measurements, four safety 4-mm banana jack inputs for Multimeter R, V and A measurements.
BNC Model P2016 Handheld DSO & DMM 6-Using the Oscilloscope 6.9 Automatic Zero-returning of Trigger Horizontal Position and Trigger Level Position When we adjust the trigger horizontal position and trigger level position to be maximal to make it off the screen center remotely, then we perform the following steps to make trigger horizontal position and trigger level position return to zero automatically.
BNC Model P2016 Handheld DSO & DMM 6-Using the Oscilloscope Figure 12: Automatic Scope Measurements 6.11 Freezing the Screen You can freeze the screen (all readings and waveforms) Press the RUN/STOP key to freeze the screen and STOP appears at top right side of the screen.
BNC Model P2016 Handheld DSO & DMM 6-Using the Oscilloscope 6.12 Using Average for Smoothing Waveforms Using the Average acquisition mode, you can smooth out the displayed waveform by averaging multiple data samples. The number of averaging is selectable from 4, 16, 64, and 128.
BNC Model P2016 Handheld DSO & DMM 6-Using the Oscilloscope the item OFF is selected, the Persistence function is closed. Look at the display, a screen like the following Figure 15 can be shown. Figure 15: Persistence to Observe Dynamic Signals 6.14 Using Peak Detection to Display Glitches...
BNC Model P2016 Handheld DSO & DMM 6-Using the Oscilloscope Figure 16: Peak Detection Term interpretation Collecting mode: The oscilloscope transforms the collected analog data into a digital form after they are gathered in the following three different modes, that is, sampling, peak value detection and averaging values.
BNC Model P2016 Handheld DSO & DMM 6-Using the Oscilloscope Figure 17: AC-Coupling 6.16 Reversing the Polarity of the Displayed Waveform To invert the input CH1 waveform, do the following: Press the MENU key and the function menu appears at the right side of the screen.
BNC Model P2016 Handheld DSO & DMM 6-Using the Oscilloscope Figure 18: Inverted On 6.17 Using Waveform Mathematics Functions When adding (CH1 + CH2), subtracting (CH1 – CH2, CH2 – CH1), multiplying (CH1 * CH2) or dividing (CH1 / CH2) the input waveforms of CHI and CH2, the Oscilloscope will display the mathematical result waveform M and the input waveforms of CH1 and CH2 on the screen.
WAVE1.BIN 、 WAVE2.BIN 、 WAVE3.BIN… or WAVE1.BMP 、 WAVE2.BMP 、 WAVE3.BMP……. Then connect the USB mass storage device with the computer after data is saved and open the data of vector format with BNC analysis software, or direct open the bitmap.
Page 37
BNC Model P2016 Handheld DSO & DMM 6-Using the Oscilloscope Disk Management of computer 4. Right click 1 or 2 red mark area, choose Format. System will display a pop-up warning message, click Yes. Format the USB disk warning 5. Set File System as FAT32, Allocation unit size 4096. Check "Perform a quick format"...
Page 38
BNC Model P2016 Handheld DSO & DMM 6-Using the Oscilloscope Formatting the USB disk setting 6. Formatting process. Formatting the USB disk Check whether the USB disk is FAT32 with allocation unit size 4096 after formatting. Use MiniTool Partition Wizard to format Download URL: http://www.partitionwizard.com/free-partition-manager.html...
Page 39
BNC Model P2016 Handheld DSO & DMM 6-Using the Oscilloscope 2. Open the software MiniTool Partition Wizard. 3. Click Reload Disk on the pull-down menu at the top left or press keyboard F5, and information about the USB disk will display on the right side with red mark 1 and 2.
Page 40
BNC Model P2016 Handheld DSO & DMM 6-Using the Oscilloscope Format setting 6. Click Apply at the top left of the menu. Then click Yes on the pop-up warning to begin formatting. Apply setting 7. Formatting process Format process Format the USB disk successfully...
BNC Model P2016 Handheld DSO & DMM 7-Using the Multimeter 7. Using the Multimeter 7.1 About this Chapter This chapter provides a step-by-step introduction to the multi-meter functions of the test tool hereafter. The introduction gives basic examples to show how to use the menus and perform basic operations.
BNC Model P2016 Handheld DSO & DMM 7-Using the Multimeter in manual operation mode and AUTO refers to the measuring range in automatic operation mode. Measurement mode indicators: DCV: Direct voltage measurement ACV: Alternating voltage measurement DCA: Direct current measurement...
BNC Model P2016 Handheld DSO & DMM 7-Using the Multimeter 7.4.1 Measuring Resistance Values To measure a resistance, do the following: Press the R key and R appears at the top of the screen. Insert the black lead into the COM banana jack input and the red lead into the V/Ω/C banana jack input.
BNC Model P2016 Handheld DSO & DMM 7-Using the Multimeter Figure 22: Diode Measurement 7.4.3 On-off Test To perform an On-off test, do the following: Press the R key and R appears on the top of the screen. Press the SET key until the following is shown on the screen.
BNC Model P2016 Handheld DSO & DMM 7-Using the Multimeter Press the R key and R appears on the top of the screen Press the SET key until C appears at the top of the screen. Insert the capacitance measurer of this multimeter to COM jack and V/Ω/C jack.
BNC Model P2016 Handheld DSO & DMM 7-Using the Multimeter Figure 25: DC Voltage Measurement 7.4.6 Measuring AC Voltage To measure the AC voltage, do the following: Press the V key and DCV appears at the top of the screen.
BNC Model P2016 Handheld DSO & DMM 7-Using the Multimeter 7.4.7 Measuring DC Current To measure a DC current which is less than 400 mA, do the following: Press the A key and DCA appears at the top of the screen. The unit on the main reading screen is mA.
BNC Model P2016 Handheld DSO & DMM 7-Using the Multimeter Figure 28: DC Current Measurement for 10A 7.4.8 Measuring AC Current To measure an AC current which is less than 400 mA, do the following: Press the A key and DCA appears at the top of the screen. The unit on the main reading screen is mA.
BNC Model P2016 Handheld DSO & DMM 7-Using the Multimeter Figure 29: AC Current Measurement for 400 mA To measure an AC current which is larger than 400 mA, do the following: Press the A key, an instruction prompt you to plug the multimeter test leads in the right position.
BNC Model P2016 Handheld DSO & DMM 7-Using the Multimeter Figure 31: Freezing the Readings 7.6 Taking a Relative Measurement A currently measured result relative to the defined reference value is displayed in a relative measurement. The following example shows how to take a relative measurement. At first, it is required to acquire a reference value.
BNC Model P2016 Handheld DSO & DMM 7-Using the Multimeter Figure 32: Relative Measurement 7.7 Selecting Automatic / Manual Range Adjustment The defaulted range mode of the instrument is an automatic range. To switch to the manual range, perform the following steps: Press F1 key and MANUAL is displayed on the top left side of the screen to enter the manual range mode.
Page 52
BNC Model P2016 Handheld DSO & DMM 7-Using the Multimeter Attention: capacitance measurement without manual range mode. Berkeley Nucleonics Corporation info@berkeleynucleonics.com | www.berkeleynucleonics.com 2955 Kerner Blvd, San Rafael, CA 94901 | 800-234-7858...
BNC Model P2016 Handheld DSO & DMM 8-Advanced Function of Oscilloscope 8. Advanced Function of Oscilloscope 8.1 About this Chapter This chapter will detail the oscilloscope function of the test tool. 8.2 Setting the Vertical CH1 and CH2 Each channel has its own independent vertical menu and each item can be set respectively based on the specific channel.
BNC Model P2016 Handheld DSO & DMM 8-Advanced Function of Oscilloscope The following Table describes the Vertical Channel menu: Function menu Setting Description The dc component in the input signal is blocked. The ac and dc components of the input signal are Coupling allowed.
BNC Model P2016 Handheld DSO & DMM 8-Advanced Function of Oscilloscope Figure 35: AC Coupling Figure 36: DC Coupling Figure 37: Ground Coupling 8.2.2 Open and Close Settings on Channel Take CH1 for example: Press F2 Channel key first, and then OFF to make a Close setting on CH1.
BNC Model P2016 Handheld DSO & DMM 8-Advanced Function of Oscilloscope After attenuating the probe level by 10:1, you also need to magnify the display level by 10 X to match the displayed amplitude with the real amplitude. Press F3 Probe to adjust the probe attenuation level.
BNC Model P2016 Handheld DSO & DMM 8-Advanced Function of Oscilloscope Press the F3 CH1+CH2 key and the obtained waveform M appears on the screen. Press OPTION until left bottom display as — CHMath Volts/Div — CHM Zero ► Press ◄ (yellow) or (yellow) to adjust the range of M waveform Press ▲...
BNC Model P2016 Handheld DSO & DMM 8-Advanced Function of Oscilloscope displayed at the bottom of the screen. Select from F1 to F5 key to make a different setting. Press OPTION and left bottom display as below during edge trigger and video trigger: —...
Page 59
BNC Model P2016 Handheld DSO & DMM 8-Advanced Function of Oscilloscope Figure 39: Edge Trigger The Edge trigger menu is described in the following table. Function Settings Description menu Rising Triggering on the rise edge of the signal. Slope Falling Triggering on the fall edge of the signal.
BNC Model P2016 Handheld DSO & DMM 8-Advanced Function of Oscilloscope 8.5.2 Video Trigger The video trigger type is designed to capture the video signal format, NTSC, PAL or SECAM. For any other signal type, use the edge trigger. Figure 40 and Figure 41 is the Video Odd Field Trigger. Figure 42 and Figure 43 is the Video Line trigger.
BNC Model P2016 Handheld DSO & DMM 8-Advanced Function of Oscilloscope setting. Make a video designed line synchronization setting. --> To next menu The Video triggering menu (Second page): 1. When the sync is Line, Field, Odd Field, Even Field, the second-page menu is shown as below.
Page 62
BNC Model P2016 Handheld DSO & DMM 8-Advanced Function of Oscilloscope Figure 44 Alternate Trigger The Alternate triggering menu is described in the following table. When the type is set as Edge trigger, Settings Description Function menu Setting trigger type and others info for Channel 1...
Page 63
BNC Model P2016 Handheld DSO & DMM 8-Advanced Function of Oscilloscope Setting trigger type and others info for Channel 1. CH SEL Setting trigger type and others info for Channel 2. Edge Set vertical channel trigger as edge trigger. Type...
BNC Model P2016 Handheld DSO & DMM 8-Advanced Function of Oscilloscope the Holdoff menu is described in the following table: Function Settings Description menu Off Time Set time slot before another trigger event. Increase Set the off time to increase.
BNC Model P2016 Handheld DSO & DMM 8-Advanced Function of Oscilloscope changes and sudden peaks, but the waveform becomes noisy. Multiple samples are averaged together. The average mode Average reduces the noise level, but the waveform must be repetitive. 4, 16, 64 Select the average number.
BNC Model P2016 Handheld DSO & DMM 8-Advanced Function of Oscilloscope Figure 46: Dots Style Figure 47: Vectors Style 8.7.2 Persistence With Persist function selected, the displayed saved original data gradually decay in color and the new data are bright in color; with infinite persistence mode selected, the oscilloscope keeps all past traces of the displayed waveform.
BNC Model P2016 Handheld DSO & DMM 8-Advanced Function of Oscilloscope Operations for various control keys are shown as below: ■ The CH1 Volts/Div and CH1 Zero for CH1 (see "6.5.1 Setting the Vertical System" on P17) are used to set the horizontal scale and position.
BNC Model P2016 Handheld DSO & DMM 8-Advanced Function of Oscilloscope 48 Cymometer Figure 8.8 Waveform Saving Setups The oscilloscope can save 4 waveforms, which can be displayed on the screen with the present waveform. The recalled waveform saved in the memory cannot be adjusted.
BNC Model P2016 Handheld DSO & DMM 8-Advanced Function of Oscilloscope 6. Press the F4 key to select ON for the address A. The waveform saved in address A will be displayed on the screen in green color. The display color is green, and the zero point of waveform, voltage and time is purple Now, you can see a screen that looks like the following Figure 49.
BNC Model P2016 Handheld DSO & DMM 8-Advanced Function of Oscilloscope Press the F3 key to save the waveform on CH1 in address A To display the saved waveform on the screen, do the following: Press the F4 key to select ON for the address A. The waveform saved in address A will be displayed on the screen in green color and the zero point of waveform, Vamp and Freq is purple.
BNC Model P2016 Handheld DSO & DMM 8-Advanced Function of Oscilloscope 1. Press the MENU key and select the FUNCTION menu using MENU ▲ or MENU ▼ key. 2. Press F2 (Auto calibration). A message appears on the display, asking you to remove all cables and probes from the oscilloscope.
Page 72
BNC Model P2016 Handheld DSO & DMM 8-Advanced Function of Oscilloscope Measure the Preshoot of CH2 Measure the Rise Time of CH1 Rise Time Measure the Rise Time of CH2 Measure the Fall Time of CH1 Fall Time Measure the Fall Time of CH2...
BNC Model P2016 Handheld DSO & DMM 8-Advanced Function of Oscilloscope Figure 51: Automatic Measurements 8.11 Setting the Cursor Measurements This oscilloscope allows you to make manual cursor measurements on time and voltage. The signal sources include Channel 1(CH1), Channel 2 (CH2).
Page 74
BNC Model P2016 Handheld DSO & DMM 8-Advanced Function of Oscilloscope Press the MENU ▲ or MENU ▼ key to select CURS MEAS. Five options are shown at the bottom of the screen. Press F1 key to select the measurement type Voltage. Two purple crossing dashed lines V1 and V2 are shown on the screen.
Page 75
BNC Model P2016 Handheld DSO & DMM 8-Advanced Function of Oscilloscope Figure 53 To use the cursor for a time measurement on CH1, do the following: Press the MENU key and the function menus are displayed at the right of the screen.
BNC Model P2016 Handheld DSO & DMM 8-Advanced Function of Oscilloscope Figure 54: Cursor time measurement When logging out submenu, every value will show on the right bottom of the screen as Figure 55: Figure 55: Use the Cursor for a Time Measurement 8.11.2 Setting the cursor measurement in FFT mode...
Page 77
BNC Model P2016 Handheld DSO & DMM 8-Advanced Function of Oscilloscope the two channels Curs1 Display the relating measured value of Cursor 1 Curs2 Display the relating measured value of Cursor 2 To make a range measurement on CH1, doing the following: Press the MENU key and the function menus are displayed at the right of the screen.
BNC Model P2016 Handheld DSO & DMM 8-Advanced Function of Oscilloscope 8.12 Autoscale The function is applied to follow-up signals automatically even if the signals change at any time. Autoscale enables the instrument to set up trigger mode, voltage division, and timescale automatically according to the type, amplitude and frequency of the signals.
Page 79
BNC Model P2016 Handheld DSO & DMM 8-Advanced Function of Oscilloscope Figure 57: Autoscale Horizontal- Vertical multi-period waveforms Note: Things you should be ware when turning on Autoscale. 1. The symbol is flickering on the top left corner of the screen every half second.
BNC Model P2016 Handheld DSO & DMM 8-Advanced Function of Oscilloscope 8.13 Record Waveform record: The function provides you to set the time interval between the frames and record the waveform up to 1000 numbers and get a better analysis effect more than before through Playback and Save functions.
Page 81
BNC Model P2016 Handheld DSO & DMM 8-Advanced Function of Oscilloscope Loop Set repeat play mode. Play mode Single Set single time play mode. Play Press to start playback. Operate Stop Press to stop playing. Interval 1ms-20s Set up interval value between frames.
Page 82
BNC Model P2016 Handheld DSO & DMM 8-Advanced Function of Oscilloscope Press the MENU ▲ or MENU ▼ key to select RECORD and five options are displayed at the bottom of the screen. Press the F1 key to select the Record mode.
BNC Model P2016 Handheld DSO & DMM 8-Advanced Function of Oscilloscope 8.14 Using FFT An FFT breaks down signals into component frequencies, which the oscilloscope uses to display a graph of the frequency domain of a signal, as opposed to the oscilloscope's standard time domain graph.
Page 84
BNC Model P2016 Handheld DSO & DMM 8-Advanced Function of Oscilloscope available at the bottom of the screen. Press F1 to set ON. The green waveform F is shown on the screen after calculation. Press F2 to switch between channel CH1 and CH2 and the current status of channel shows on the top left.
Page 85
BNC Model P2016 Handheld DSO & DMM 8-Advanced Function of Oscilloscope screen when the format is Vrms. ▲/▼ — CH2 Volts/Div ▲/▼ — FFT Zero Press VOLTS POSITION (red) and the screen show as follows: ▲/▼ —CH1 OFF When FFT source is CH1: ▲/▼...
Page 86
BNC Model P2016 Handheld DSO & DMM 8-Advanced Function of Oscilloscope departures from the Origin for 2.00DIV. The shown frequency of M is the exact frequency of the cursor point in the middle of spectrum. Shown as Figure 60. ◄/►...
Page 87
BNC Model P2016 Handheld DSO & DMM 8-Advanced Function of Oscilloscope varying spectrum. This is a very good window for resolving frequencies that are very close to the same value with somewhat improved amplitude accuracy over the rectangle window. It has a...
Page 88
BNC Model P2016 Handheld DSO & DMM 8-Advanced Function of Oscilloscope Fig 62: Blackman window Fig.63. Hanning window window Fig.64. Hamming Quick Tips If desired, use the zoom feature to magnify the FFT waveform. Use the default dB scale to see a detailed view of multiple frequencies, even if they Berkeley Nucleonics Corporation info@berkeleynucleonics.com | www.berkeleynucleonics.com...
BNC Model P2016 Handheld DSO & DMM 8-Advanced Function of Oscilloscope Vrms have very different amplitudes. Use the scale to see an overall view of how all frequencies compare to each other. Signals that have a DC component or offset can cause incorrect FFT waveform component magnitude values.
BNC Model P2016 Handheld DSO & DMM 8-Advanced Function of Oscilloscope The screen that looks like the following figure 65 will be displayed. Figure 65: System State 8.15.1 Real time clock Function: Be synchronous with the U disk storage time.
BNC Model P2016 Handheld DSO & DMM 8-Advanced Function of Oscilloscope 6. Press OPTION ◄ or OPTION ► to switch to the year, month, date, hour, minute and second that are in need of adjustment. 7. Press OPTION to enter into normal time mode after timing.
Press F4 key, select the Bitmap or Vectors for data transmission. Use a data line to connect the oscilloscope and the PC. Open the BNC software that had been installed completely. Setting the parameter under the user's manual, and then starts data transmission.
BNC Model P2016 Handheld DSO & DMM Trouble Shooting 9. Trouble Shooting 1. The oscilloscope does not power up It may be caused by the dead battery. At this time, the oscilloscope will not start even if it is powered by the battery charger. Firstly, charge the battery and supply the oscilloscope with power through the battery charger.
Page 94
BNC Model P2016 Handheld DSO & DMM Trouble Shooting In such condition, adjust and make the trigger level in the middle or select the non-auto trigger mode. Additionally, press the AUTO SET key and complete the above setting 7. When you select the average sampling in the sampling mode or select a longer display time in the display mode under the oscilloscope, the display speed is slow.
BNC Model P2016 Handheld DSO & DMM 10-Appendix 10. Appendix 10.1 Appendix A: Specifications 10.1.1 Oscilloscope Only if other instructions are provided, are all technical specifications applicable to the probe with the 10X attenuation switch setting and the HDS-N series digital type oscilloscope.
Page 96
BNC Model P2016 Handheld DSO & DMM 10-Appendix Sampling rate range P2016 0.25 Sa/s - Sa/s (sin x) /x Waveform interpolation Record length 6k points on each channel Scanning speed range 5ns/div - 100s/div, stepping in the "1-2-5" P2016 (S/div)...
Page 97
BNC Model P2016 Handheld DSO & DMM 10-Appendix DC gain accuracy ±3% DC measurement accuracy The voltage difference (△V) between any two points on the (average value sampling waveform after averaging the captured waveforms more mode) than 16: ±(5% reading + 0.05 divisions).
BNC Model P2016 Handheld DSO & DMM 10-Appendix Bandwidth Up to 6 MHz (DC) Up to full bandwidth (DC) Attenuation rate 1: 1 10: 1 Compensation 15 pf – 35 pf range Input resistance 1 MΩ ± 2% 10 MΩ ± 2%...
Page 99
BNC Model P2016 Handheld DSO & DMM 10-Appendix Range Accuracy Resolution 40.00 mA ±1.5%±3 digits 10 uA 400.0 mA ±2%±1 digit 100 uA 10 A ±5%±3digits 10 mA Resistance Range Accuracy Resolution 400.0 Ω 0.1 Ω ±1%±3 digits 1 Ω...
BNC Model P2016 Handheld DSO & DMM 10-Appendix 10.1.3 General Specifications Basic parameter Mechanical dimension 18 cm×11.5 cm×4 cm Weight 645 g < 7 W Power consumption MODEL P2016 Display type 3.7" color liquid crystal display Display resolution 640 (horizontal) ×480 (vertical) pixels...
BNC Model P2016 Handheld DSO & DMM 10-Appendix 10.2 Appendix B: Maintenance and Cleaning 10.2.1 Maintenance Do not store or place the instrument in locations where the liquid crystal display (LCD) may be directly exposed to the sunshine for a long time.