Fluke 87V/IMSK User Manual page 56

Table of Contents

Advertisement

80 Series V
Users Manual
Function
Range
60.00 mA
mA
400.0 mA
\
6.000 A
10.00 A
(45 Hz to 2 kHz)
60.00 mA
mA
400.0 mA
6.000 A
[
10.00 A
μA B
600.0 μA
(45 Hz to 2 kHz)
6000 μA
μAF
600.0 μA
6000 μA
1.
AC conversion for Model 83 is ac coupled and calibrated to the rms value of a sine wave input.
2.
AC conversions for Model 87 are ac coupled, true rms responding, and valid from 3 % to 100 % of range, except 400 mA range (5 %
to 100 % of range) and 10 A range (15 % to 100 % or range).
3.
Model 87 is a true rms responding meter. When the input leads are shorted together in the ac functions, the Meter may display a
residual reading between 1 and 30 counts. A 30 count residual reading will cause only a 2 digit change for readings over 3 % of
range. Using REL to offset this reading may produce a much larger constant error in later measurements
W 10 A continuous up to 35 °C; < 20 minutes on, 5 minutes off at 35 °C to 55 °C. 20 A for 30 seconds maximum; > 10 A unspecified.
4.
5.
Below a reading of 200 counts, add 10 counts.
6.
400 mA continuous; 600 mA for 18 hrs maximum.
48
Table 14. Current Function Specifications
Resolution
0.01 mA
6
0.1 mA
0.001 A
4
0.01 A
0.01 mA
6
0.1 mA
0.001 A
4
0.01 A
0.1 μA
1 μA
0.1 μA
1 μA
Accuracy
1
Model 83
Model 87
5
± (1.2 % + 2)
± (1.0 % + 2)
5
± (1.2 % + 2)
± (1.0 % + 2)
5
± (1.2 % + 2)
± (1.0 % + 2)
5
± (1.2 % + 2)
± (1.0 % + 2)
± (0.4 % + 4)
± (0.2 % + 4)
± (0.4 % + 2)
± (0.2 % + 2)
± (0.4 % + 4)
± (0.2 % + 4)
± (0.4 % + 2)
± (0.2 % + 2)
5
± (1.2 % + 2)
± (1.0 % + 2)
5
± (1.2 % + 2)
± (1.0 % + 2)
± (0.4 % + 4)
± (0.2 % + 4)
± (0.4 % + 2)
± (0.2 % + 2)
Burden Voltage
2, 3
(typical)
1.8 mV/mA
1.8 mV/mA
0.03 V/A
0.03 V/A
1.8 mV/mA
1.8 mV/mA
0.03 V/A
0.03 V/A
100 μV/μA
100 μV/μA
100 μV/μA
100 μV/μA