The device must never be modified without written authorization of FARO S.p.A. For maintenance only original FARO’s spare parts are allowed. Fail to comply with this warning will immediately make decay the warranty and the conformity of the product to international regulations and directives on Medical Devices.
Lubricating grease must be applied to the surface of the post before the connection is done. Faro suggests the following grease to be applied: Molykote 111 silicon Grease. With a digital level, ensure the connection element on the unit is perfectly parallel to the ground (max 2°)
MP3020 INSTALLATION MANUAL - EN Take care that the fixed arm is parallel to ground in each position. Check the light stays balanced in all positions and no drifting of the swivel arm occurs. If necessary, adjust the force on the spring by screwing or unscrewing the nut (B) with the suitable tool until the arm is balanced.
Warnings for electrical danger and suspended masses - The device has to be installed by specialized technicians only - The light has to be installed with FARO applications only. - The light is supplied with rotation limit switch between the fixed and the swivel arm.
MP3020 INSTALLATION MANUAL - EN 2.4.2 Installation procedure A After having fixed as reference point the centre of the chair (A). install at a distance of 650mm and 150mm, according to the directions given in the picture. B. Unfit the flange (7) by removing the nuts (12) and washers (11).
MP3020 INSTALLATION MANUAL - EN N. Screw (without locking) the nuts (12) O. Unscrew the three screws (19) of the column (14) and remove the bushing and the remaining (20) washers (11) on the screws (8) of the ceiling flange (1) P.
MP3020 INSTALLATION MANUAL - EN 2.4.4 Electrical drawing – ceiling mounting without transformer WALL MOUNTING Warnings for electrical danger and fall of suspended masses The device must be installed by specialized tecnicians The power supply in the room where the fitting must be installed must be always switched off.
MP3020 INSTALLATION MANUAL - EN Connect the lamp wires to the terminal Electrical connector (5) (see wiring diagram below) including the grounding wire. Connect wires coming wall terminal board, case been previously walled absence this precaution, connection it must be carried out with an external flying cable, to be introduced into the cable gland (10).
Warning for danger of suspended mass falling Only use screws supplied by FARO. Screw in the safety screws together. Before removing the nut (G) ALWAYS remove the safety screw F. NEVER unscrew the nut (G) with the screws F mounted.
MP3020 INSTALLATION MANUAL - EN 2.8.2 Electrical Requirements The requirements for correct installation of the head are as follows: Power Supply Supply cables Power Supply requirements Type Conformity Device 60601-1 17-24 Vac Transformer complies with IEC/EN 60601- Component The medical 50/60 Hz 1 third edition.