SPECIFICATIONS

Range:
–76 to 932°F / –60 to 500°C
Resolution: 0.1° (-9.9 to 199.9)
1° (otherwise)

OPERATION

Press and continue to hold down the
trigger to take temperature readings.
The unit will take temperature readings
while the trigger is pressed and held
(SCAN appearing on the display indicates
measurements are being taken).
After the trigger is released, the last
temperature measured will be displayed
for approximately 15 seconds and the
display will turn off automatically. HOLD
will appear on the display when the
trigger has been released to indicate that
the unit is no longer taking temperature
measurements.

DISTANCE SPOT/RATIO

There are other factors that may affect
measurement accuracy. The target
must completely fill the spot diameter
seen by the infrared sensor; otherwise
readings will be influenced by the
surface surrounding the target. The ratio
of the distance to the size of the spot
being measured is 11:1. For example,
an object's diameter of 100mm can be
measured from 1100mm.
GlobalTestSupply
www.

SELECTING UNITS OF MEASURE

Each press of the °C/°F button will toggle
between °C and °F.

LASER TARGETING

Laser targeting is always active when
the trigger is being pressed. Each time
the trigger is pressed and held, the laser
beam will be emitted.
CAUTION: Eye damage may result from
direct exposure to laser light.

MAXIMUM TEMPERATURE

While taking measurements, the unit
will automatically record the maximum
temperature reading that has been
measured. This value will appear on the
display to the right of MAX.

DISPLAY MESSAGES

"Er2" will appear on the display when
the unit is exposed to rapid changes in
ambient temperature.
"Er3" will appear on the display when
the ambient temperature is below 32° F
(0° C) or above 122° F (50° C).
"Er" will appear on the display when the
unit needs to be reset. To reset the unit,
remove the batteries, wait 2 minutes and
then re-install the batteries.
.com
sales@GlobalTestSupply.com

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Traceable 4470 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Cole Parmer Traceable 4470

  • Page 1 SPECIFICATIONS SELECTING UNITS OF MEASURE Range: –76 to 932°F / –60 to 500°C Each press of the °C/°F button will toggle Resolution: 0.1° (-9.9 to 199.9) between °C and °F. 1° (otherwise) LASER TARGETING OPERATION Laser targeting is always active when Press and continue to hold down the the trigger is being pressed.
  • Page 2 WARRANTY or SERVICE For warranty or service, contact: ® TRACEABLE PRODUCTS TRACEABLE ® INFRARED ® Traceable Products is ISO 9001:2018 Quality- Certi ied by DNV and ISO/IEC 17025:2017 THERMOMETER accredited as a Calibration Laboratory by A2LA. INSTRUCTIONS Note: Traceable Infrared thermometers, are NOT approved for Medical usage, and are ®...

Table of Contents