XT/CT/LT/CTS Service Manual
LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]
1

Advertisement

Table of Contents
loading

Summary of Contents for LABORIE Aquarius XT

  • Page 1 Aquarius® XT/CT/LT/CTS Service Manual LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
  • Page 2 Trademarks AQUARIUS is a registered trademark of Laborie Medical Technologies, Corp., registered mark applies to USA and may not apply to other geographies. T-DOC, Unisensor, Roam DX, Urocap, and i-LIST Office Reporter are trademarks of Laborie Medical Technologies, Corp.
  • Page 3: Table Of Contents

    Cleaning and Preventive Maintenance ....................37      General Cleaning and Preventive Maintenance ......................37     Caring for the Roam™ DX ..............................37     Caring for the Urocap V Uroflowmeter ..........................37 LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
  • Page 4 Troubleshoot and Service ........................48      6.1.1 BLUETOOTH CONNECTION ..........................50 7  Electromagnetic Compatibility (EMC) ..................... 55  8  Symbols ..............................61  9  Schematics ..............................62  10  End-User Software License Agreement ....................62  LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
  • Page 5 Figure 32: Urocap™ V Charging Spacer ........................42  Figure 33: Roam™ DX Charging Bay ......................... 43  Figure 34: UPP ................................43  Figure 35: EMG Chanel During Check ........................46  Figure 36: Safe movement of cart ..........................47  LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
  • Page 6 Table 21: Electronic Emissions ........................... 56  Table 22: Electromagnetic Immunity ......................... 57  Table 23: Electromagnetic Immunity – Radio Frequency Communications ............59  Table 24: Separation Distances ..........................60  Table 25: Symbols Glossary ............................61  LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
  • Page 7: About This Manual

    Common abbreviations used in this manual:  CHC: Central Hub Computer  CFG: software configuration file 1 INTRODUCTION This manual is intended for servicing the AQUARIUS UDS system manufactured by LABORIE. It collects data from the following devices:  Uroflow Transducers  Water Pressure Transducers ...
  • Page 8: Technical Specifications

    Luxo: Extension 36.0” (91.5 cm) / Weight: 3.3 lbs. (1.5kg) Speed: 0 to 3 mm/s Maximum load for printer shelf and 22 Lbs. (10 kg) workstation shelf Spinning Disk Flow Transducer Width*: 16” (400 mm) LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
  • Page 9: Classifications

    Class I Type BF Applied Part Mode of Operation: Continuous Equipment not suitable for use in the presence of a Flammable Anesthetic Mixture with Air or Oxygen or Nitrous Oxide. Table 3: AQUARIUS® Classifications LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
  • Page 10: Type Bf Applied Parts

    Computer, Medical Grade All-In-One PC, 24” Widescreen, Configured AQUARIUS™ XT AQS3000 Aquarius XT Urodynamic Analyzer Tower, 120V AQS3001 Aquarius XT Urodynamic Analyzer Workstation, with Single LCD, 120V (double LCD optional) AQS3002 Aquarius XT Urodynamic Analyzer Tower, 220V AQS3003 Aquarius XT Urodynamic Analyzer Workstation, with Single LCD, 220V (double...
  • Page 11 Custom PC, Windows 7 Ultimate, 64-bit Computer, Medical Grade All-In-One PC, 24” Widescreen, Configured SOFTWARE COMPONENTS LAB1000 Laborie Software, NGM Urodynamics Software Package, LT Software LAB1001 Laborie Software, NGM Urodynamics Software Package, CT Software LAB1002 Laborie Software, NGM Urodynamics Software Package, XT Software...
  • Page 12 T-DOC® Air Charged Catheters™, Abdominal, 7 fr CAT895 T-DOC® Air Charged Catheters™, Single Sensor, 7 fr CAT880 T-DOC® Air Charged Catheters™, Dual Sensor, 7 fr TUB101 Measurement Tubing TUB500 Pump Tubing with Infusion line LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
  • Page 13: Table 5: Aquarius® Equipment And Parts List

    MINI BASE – UPP ARM (UPP STAND with PULLER) CAB1000 Cable, Custom USB Type A - to - Custom Type Mini B PRI1000/PRI1001 Printer, HP, Wireless 9034K0103 Spinning Disk Flow Transducer Table 5: AQUARIUS® Equipment and Parts List LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
  • Page 14: Equipment Setup

    2.3 EQUIPMENT SETUP 2.3.1 C ETUP LABORIE recommends the AQUARIUS system setup as follows: 1. Open shipping box. 2. For carts with shelves, lift and align shelf 3. Place the plastic bottom 4. Secure the power cord with screws and slide into position.
  • Page 15: Table 7: Spinning Disk Flow Transducer Setup

    2.3.1.3 Spinning Disk Flow Transducer Setup LABORIE recommends the Spinning Disk Flow Transducer setup as follows: 1 = Splash guard 2 = Transducer cup 3 = Rotating disk 4 = Disk holder 5 = Motor housing 6 = Jug (1.5L steel; 1L white)
  • Page 16: Upp Puller (Optional)

    Use a 4mm Hex L-Key to insert the Puller into the Clamp Assembly (Figure 4). Puller Base Assembly Puller Stand Assembly Knob Screw Clamp Assembly Pole Assembly Washer, M8 Split Lock Washer, M8 Screw Socket Head, M8 H-Base Assembly Figure 4: Puller Base and Puller Stand Assembly LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
  • Page 17: Arm And Stand

    (6). When the drive is engaged, the catheter will be pushed or pulled along the nose tube (7). For sterilization purposes, both the catheter guide and the mobile clamp may be removed. LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
  • Page 18: Urocap™ V And Spinning Disk Flow Transducer Setup

    Refer to Infusion Line  The saline bag must not touch any other mechanical component of the system (such as transducers, etc.). Figure 9: Infusion Transducer Connections LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
  • Page 19: Pump Setup

    Roam™ DX on a nearby table/counter at level with the patient’s waist (or higher). This will help prevent any liquid from dripping along the cables to the Roam™ DX device when the patient voids. Keep level with patient’s waist or higher Figure 11: Roam Patient Setup LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
  • Page 20: Consumables And Transducer Setup

    6. Turn the stopcock “OFF” towards the Luer lock plug (See Figure 13, Option D). 7. Prime the pressure measurement tubing and the catheter. 8. Turn the stopcock “OFF” towards the syringe for pressure recording (See Figure 13, Option A). LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
  • Page 21: Catheter Setup

    ETUP The following section is a reference guide to catheter setup and connection. Ensure that only single-, dual-, and triple-lumen catheters supplied by LABORIE are used for Urodynamic tests involving the UDS System. Do NOT use Foley catheters, as these catheters are not designed for pressure measurement, bladder infusion, or pressure recording.
  • Page 22: Figure 14: Abdominal Single-Lumen Balloon Catheter Priming

    (A). 4. Fill the balloon with 1.5-3 cc’s of water. 5. Turn the stopcock switch OFF to the syringe (B). Figure 15: OFF to the Transducer Figure 16: OFF to the Syringe Cartridge LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
  • Page 23: T-Doc Air-Charged Catheter Setup

    Figure 17: EMG Patch Placement. NOTE: LABORIE recommends that each electrode be secured in place with an additional piece of surgical tape. If the electrodes detach, dry the skin and apply new surgical tape. LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
  • Page 24: Equipment Status And Light Signals

    2 = USB port: Connect the USB cable from the UPP or the Urocap™ V into the port. 3 = BDT: Plug the cable from the bubble detector into this port. Figure 19: Pump Connections LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
  • Page 25: Roam™ Dx

    Blinking orange = No external power. Battery is low. Connect external power immediately. OFF = Bluetooth channel is disconnected. ON = Bluetooth channel is connected. Every transmitted block forces this LED to blink. LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
  • Page 26: Urocap™ V

    = low battery; time to re-charge light off = battery empty; re-charge The Urocap™ V buttons are described below. Refer to Figure 22 for the location of buttons on the device. LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
  • Page 27: Upp

    Indicators blinking green = connected and sampling orange = System fault blinking orange = not connected light off = system off NOTE: Closely monitor the battery level of the UPP when in use. LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
  • Page 28: Calibration

    3. Listen for a `beep` from the Urocap™ V to signal it is ready for calibration. The touchscreen will display instructions for calibration of the device. 4. Follow the on-screen instructions to complete the calibration: Table 9: Calibrating the Urocap™ V LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
  • Page 29: Calibrate Flow And Volume With The Spinning Disk Flow Transducer

    After 10 seconds of no flow, spinning disk will automatically complete calibration. 2. When complete, tap the Home button to return to the Home screen: Table 10: Calibrating the Spinning Disk Flow Transducer LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
  • Page 30: How To Calibrate The Pressure And The Emg

    7. Tap the Return button to go back to the Calibrate Pressure Transducers screen: 8. Repeat steps 2 to 7 for the other Transducer Types. Table 11: Calibrating Roam™ DX Pressure Channels and Transducers LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
  • Page 31: Calibrate Emg

    4.3 HOW TO CALIBRATE THE PUMP Calibrate the pump by assigning different Pump sets. A maximum of 10 sets are allowed. 1. Tap the Calibration icon on the home screen: Tap to select the ‘HUB’ device: LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
  • Page 32 Select a Urocap™ V device from the dropdown menu, then tap Next to continue: 8. Select the correct catheter size from the listing, then tap Next: Select the fluid density from the combo box and then tap Next: LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
  • Page 33: Table 13: Calibrating The Pump

    NOTE: During calibration there will be 8 calibration steps. 13. Once complete, the pump calibration window will provide a confirmation message that calibration is complete: 14. Press the Home button to exit. Table 13: Calibrating the Pump LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
  • Page 34: How To Calibrate The Infusion Transducer

    Tap to select the Infusion Transducer Calibration option: Select a transducer and tap Next. Low Load Type in the value and then tap Next. Continue by following the on-screen instructions until calibration is complete: Table 14: Calibrating the Infusion Transducer LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
  • Page 35: How To Calibrate The Upp Puller

    1. Tap the Calibration icon on the home screen: 2. Tap to select the UPP device: 3. Follow the on-screen instructions to complete the calibration: 4. Tap the Home button to exit: Table 15: Calibrating the UPP Puller LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
  • Page 36: System Status Messages

    Central Hub Computer (CHC). This status message appears under the battery icon. Disconnect the Roam™ DX from the power supply then reconnect. If the message still appears, contact LABORIE technical support. Table 16: CHC - System Status Messages...
  • Page 37: Cleaning And Preventive Maintenance

     LABORIE recommends charging and using the Roam™ DX at least once or twice a year to maintain optimal performance. If it will be placed in storage, make sure it is stored with the battery fully charged.
  • Page 38: Caring For The Spinning Disk Flow Transducer

    6. After cleaning, rinse the components. 7. Reassemble the flow transducer. IMPORTANT:  Never immerse the connector into any liquid. Keep it dry at all times.  Never remove the disk holder for cleaning. LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
  • Page 39: Caring For The Upp Puller

    Sani-Cloth Plus or equivalent  Dilute Bleach wipes – do not soak  Alcohol is nearly inactive against certain organisms. 70% Isopropyl Alcohol cleaners may be used on the devices; however, higher level cleaners are recommended. LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
  • Page 40: Figure 27: Mobile Clamp Moving Toward Catheter Guide

    5. Slide the mobile clamp completely off the nose tube. (Figure 29) 6. Clean and disinfect the mobile clamp. 7. Thoroughly rinse and dry the clamp. Figure 29: Mobile clamp off nose tube LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
  • Page 41: Sterilization Of The Catheter Guide

    NOTE: The tip has a silver ball that allows you to snap the Catheter Guide back into place and keep it at the correct position. Figure 31: Catheter Guide Removal - Step 2 LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
  • Page 42: Caring For The Air-Charged Transducer/Cable

    HIV, HBV, HCV, VRE, MRSA, MRSE, Pseudomonas and Mycobacterium.  Wipe the transducer/cable until dry. 5.7 BATTERY - CHARGING AND PREVENTIVE MAINTENANCE IMPORTANT! For proper and safe charging, use only USB cables provided by LABORIE. 5.7.1 C HARGING THE ATTERY Urocap™...
  • Page 43: Battery Preventive Maintenance

    5.8 COMPUTER VIRUS PROTECTION All computers purchased from LABORIE are virus-free before shipment and are installed with an antivirus program. It is the customer’s responsibility to correctly use and maintain the anti-virus program to prevent virus problems. LABORIE is not responsible for any virus-related computer problems after point of delivery to the customer.
  • Page 44: Environmental Consideration Of Waste Disposal

    The volume channel should match the pump fill volume (+/- 10 ml). If the reading is above or below the 10ml threshold, Once calibration is complete, perform a re-check by following steps 1 to 5 listed here. then calibration is required. Contact LABORIE service for assistance if problems continue after the calibration and re-check. LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
  • Page 45: Check The Roam™ Dx

    You will see an increase in EMG activity on the EMG channel. 7. Unclench your fist and return to the relaxed position. 8. Check the EMG channel for activity. at 100% strain with full contraction LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
  • Page 46: Check The Infusion Transducer

    O). If the reading is above or below the 2 cmH Once calibration is complete perform a re-check by following the instructions threshold, then calibration is required. provided above. Contact LABORIE service for assistance if problems continue after the calibration and re-check. 5.11.4 HECK THE NFUSION RANSDUCER Ensure the saline bag does not touch any other mechanical component of the system ( such as transducers, etc.) during the...
  • Page 47: Check The Upp Puller

    Once calibration is complete perform a re-check proximal mark in the required range of time, then calibration is required. by following the instructions provided above. Contact LABORIE service for assistance if problems continue after the calibration and re-check. 5.12 SAFELY MOVING THE CART 1.
  • Page 48: Troubleshoot And Service

    6 TROUBLESHOOT AND SERVICE If problems continue, contact LABORIE’s Service team at 1-800-333-1039 or email service@laborie.com. Follow the instructions provided in the table below for troubleshooting steps. Symptom(s) Possible Cause(s) Check/Corrective Action(s) ALL DEVICES No response from the No power on electrical Plug the system into a known working electrical outlet.
  • Page 49 5. Click Run and then click Start Pump on the control panel. 6. Click Info then click Patient information. 7. Click OK. 8. Re-select Use MPUR for pump speeds. Click OK to load the correct MPUR speeds for the test. LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
  • Page 50: Bluetooth Connection

    Reconnect devices in the touchscreen: PC via touchscreen? 1. First disconnect all Bluetooth devices via the touchscreen. 2. Restart the software and establish a connection to the 3. Reconnect all Bluetooth devices via the touchscreen. Refer LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
  • Page 51 Close all open programs. Office Reporter? Make sure i-LIST Message Queue Listener is running by clicking Start > Programs > Laborie > i-LIST MsgQListener. Restart the computer and attempt to send a report to i- LIST Office Reporter. If unsuccessful try the solution below.
  • Page 52 BUBBLE DETECTOR Red light on sensor lit Pump tubing is not Make sure the pump tubing is placed securely into the groove even though bubbles inserted securely? on the sensor. are not visible LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
  • Page 53 In the section click the Change button and select the AutoHotKey Unicode application in the resulting window and then click OK: Opens with 27. The section will display the AutoHotKey application. 28. Click OK. LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
  • Page 54: Table 19: Troubleshooting

    Table 19: Troubleshooting NOTE: The Laborie Aquarius LT, CT, CTS and XT are designed to be used as stand-alone cart systems, with the only mains power supply connection going through the LABORIE medical-grade power supply (LIT) in the base of the cart. Do not connect it to any other devices unless the whole system is outside the patient environment or the other device has its own medical grade power supply.
  • Page 55: Electromagnetic Compatibility (Emc)

    Consult the dealer or experienced technical personnel for help. WARNING! Changes or modifications not expressly approved by LABORIE could void the user’s authority to operate the equipment. This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference and (2) this device must accept any interference received, including interference that may cause undesired operation.
  • Page 56: Table 21: Electronic Emissions

    It Flicker Emissions IEC may be necessary to take mitigation measures, such as re- orienting 61000-3-3 or relocating the Aquarius Urodynamic Analyzer System or shielding the location. Table 21: Electronic Emissions LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
  • Page 57: Table 22: Electromagnetic Immunity

    Magnetic Field typical commercial or hospital environment. (50/60 Hz) IEC 61000-4-8 NOTE is the a.c. mains voltage prior to application of the test level. Table 22: Electromagnetic Immunity LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
  • Page 58 If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the Aquarius Urodynamic Analyzer System b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m. LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
  • Page 59: Table 23: Electromagnetic Immunity - Radio Frequency Communications

    Table 23: Electromagnetic Immunity – Radio Frequency Communications LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
  • Page 60: Table 24: Separation Distances

    NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. Table 24: Separation Distances LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
  • Page 61: Symbols

    IEC 60417 – Graphic Symbols for Use on Equipment. NOTE: Sterility symbols are applicable to consumbale devices only. Refer to consumable device instructions for use for complete symbol and instructional overview. Table 25: Symbols Glossary LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
  • Page 62: Schematics

    9 SCHEMATICS UPON REQUEST, LABORIE will make available to the customer any circuit diagrams, component part lists, and other technical documentation directly related to the AQUARIUS UDS system. 10 END-USER SOFTWARE LICENSE AGREEMENT Find licenses and terms for LABORIE products and services in the following locations: End User License Agreement: http://www.laborie.com/eula/...