Page 1
Aquarius® XT/CT/LT/CTS Service Manual LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
Page 2
Trademarks AQUARIUS is a registered trademark of Laborie Medical Technologies, Corp., registered mark applies to USA and may not apply to other geographies. T-DOC, Unisensor, Roam DX, Urocap, and i-LIST Office Reporter are trademarks of Laborie Medical Technologies, Corp.
Cleaning and Preventive Maintenance ....................37 General Cleaning and Preventive Maintenance ......................37 Caring for the Roam™ DX ..............................37 Caring for the Urocap V Uroflowmeter ..........................37 LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
Page 4
Troubleshoot and Service ........................48 6.1.1 BLUETOOTH CONNECTION ..........................50 7 Electromagnetic Compatibility (EMC) ..................... 55 8 Symbols ..............................61 9 Schematics ..............................62 10 End-User Software License Agreement ....................62 LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
Page 5
Figure 32: Urocap™ V Charging Spacer ........................42 Figure 33: Roam™ DX Charging Bay ......................... 43 Figure 34: UPP ................................43 Figure 35: EMG Chanel During Check ........................46 Figure 36: Safe movement of cart ..........................47 LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
Common abbreviations used in this manual: CHC: Central Hub Computer CFG: software configuration file 1 INTRODUCTION This manual is intended for servicing the AQUARIUS UDS system manufactured by LABORIE. It collects data from the following devices: Uroflow Transducers Water Pressure Transducers ...
Class I Type BF Applied Part Mode of Operation: Continuous Equipment not suitable for use in the presence of a Flammable Anesthetic Mixture with Air or Oxygen or Nitrous Oxide. Table 3: AQUARIUS® Classifications LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
MINI BASE – UPP ARM (UPP STAND with PULLER) CAB1000 Cable, Custom USB Type A - to - Custom Type Mini B PRI1000/PRI1001 Printer, HP, Wireless 9034K0103 Spinning Disk Flow Transducer Table 5: AQUARIUS® Equipment and Parts List LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
2.3 EQUIPMENT SETUP 2.3.1 C ETUP LABORIE recommends the AQUARIUS system setup as follows: 1. Open shipping box. 2. For carts with shelves, lift and align shelf 3. Place the plastic bottom 4. Secure the power cord with screws and slide into position.
2.3.1.3 Spinning Disk Flow Transducer Setup LABORIE recommends the Spinning Disk Flow Transducer setup as follows: 1 = Splash guard 2 = Transducer cup 3 = Rotating disk 4 = Disk holder 5 = Motor housing 6 = Jug (1.5L steel; 1L white)
Use a 4mm Hex L-Key to insert the Puller into the Clamp Assembly (Figure 4). Puller Base Assembly Puller Stand Assembly Knob Screw Clamp Assembly Pole Assembly Washer, M8 Split Lock Washer, M8 Screw Socket Head, M8 H-Base Assembly Figure 4: Puller Base and Puller Stand Assembly LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
(6). When the drive is engaged, the catheter will be pushed or pulled along the nose tube (7). For sterilization purposes, both the catheter guide and the mobile clamp may be removed. LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
Refer to Infusion Line The saline bag must not touch any other mechanical component of the system (such as transducers, etc.). Figure 9: Infusion Transducer Connections LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
Roam™ DX on a nearby table/counter at level with the patient’s waist (or higher). This will help prevent any liquid from dripping along the cables to the Roam™ DX device when the patient voids. Keep level with patient’s waist or higher Figure 11: Roam Patient Setup LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
6. Turn the stopcock “OFF” towards the Luer lock plug (See Figure 13, Option D). 7. Prime the pressure measurement tubing and the catheter. 8. Turn the stopcock “OFF” towards the syringe for pressure recording (See Figure 13, Option A). LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
ETUP The following section is a reference guide to catheter setup and connection. Ensure that only single-, dual-, and triple-lumen catheters supplied by LABORIE are used for Urodynamic tests involving the UDS System. Do NOT use Foley catheters, as these catheters are not designed for pressure measurement, bladder infusion, or pressure recording.
(A). 4. Fill the balloon with 1.5-3 cc’s of water. 5. Turn the stopcock switch OFF to the syringe (B). Figure 15: OFF to the Transducer Figure 16: OFF to the Syringe Cartridge LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
Figure 17: EMG Patch Placement. NOTE: LABORIE recommends that each electrode be secured in place with an additional piece of surgical tape. If the electrodes detach, dry the skin and apply new surgical tape. LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
2 = USB port: Connect the USB cable from the UPP or the Urocap™ V into the port. 3 = BDT: Plug the cable from the bubble detector into this port. Figure 19: Pump Connections LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
Blinking orange = No external power. Battery is low. Connect external power immediately. OFF = Bluetooth channel is disconnected. ON = Bluetooth channel is connected. Every transmitted block forces this LED to blink. LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
= low battery; time to re-charge light off = battery empty; re-charge The Urocap™ V buttons are described below. Refer to Figure 22 for the location of buttons on the device. LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
Indicators blinking green = connected and sampling orange = System fault blinking orange = not connected light off = system off NOTE: Closely monitor the battery level of the UPP when in use. LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
3. Listen for a `beep` from the Urocap™ V to signal it is ready for calibration. The touchscreen will display instructions for calibration of the device. 4. Follow the on-screen instructions to complete the calibration: Table 9: Calibrating the Urocap™ V LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
After 10 seconds of no flow, spinning disk will automatically complete calibration. 2. When complete, tap the Home button to return to the Home screen: Table 10: Calibrating the Spinning Disk Flow Transducer LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
7. Tap the Return button to go back to the Calibrate Pressure Transducers screen: 8. Repeat steps 2 to 7 for the other Transducer Types. Table 11: Calibrating Roam™ DX Pressure Channels and Transducers LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
4.3 HOW TO CALIBRATE THE PUMP Calibrate the pump by assigning different Pump sets. A maximum of 10 sets are allowed. 1. Tap the Calibration icon on the home screen: Tap to select the ‘HUB’ device: LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
Page 32
Select a Urocap™ V device from the dropdown menu, then tap Next to continue: 8. Select the correct catheter size from the listing, then tap Next: Select the fluid density from the combo box and then tap Next: LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
NOTE: During calibration there will be 8 calibration steps. 13. Once complete, the pump calibration window will provide a confirmation message that calibration is complete: 14. Press the Home button to exit. Table 13: Calibrating the Pump LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
Tap to select the Infusion Transducer Calibration option: Select a transducer and tap Next. Low Load Type in the value and then tap Next. Continue by following the on-screen instructions until calibration is complete: Table 14: Calibrating the Infusion Transducer LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
1. Tap the Calibration icon on the home screen: 2. Tap to select the UPP device: 3. Follow the on-screen instructions to complete the calibration: 4. Tap the Home button to exit: Table 15: Calibrating the UPP Puller LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
Central Hub Computer (CHC). This status message appears under the battery icon. Disconnect the Roam™ DX from the power supply then reconnect. If the message still appears, contact LABORIE technical support. Table 16: CHC - System Status Messages...
LABORIE recommends charging and using the Roam™ DX at least once or twice a year to maintain optimal performance. If it will be placed in storage, make sure it is stored with the battery fully charged.
6. After cleaning, rinse the components. 7. Reassemble the flow transducer. IMPORTANT: Never immerse the connector into any liquid. Keep it dry at all times. Never remove the disk holder for cleaning. LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
Sani-Cloth Plus or equivalent Dilute Bleach wipes – do not soak Alcohol is nearly inactive against certain organisms. 70% Isopropyl Alcohol cleaners may be used on the devices; however, higher level cleaners are recommended. LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
5. Slide the mobile clamp completely off the nose tube. (Figure 29) 6. Clean and disinfect the mobile clamp. 7. Thoroughly rinse and dry the clamp. Figure 29: Mobile clamp off nose tube LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
NOTE: The tip has a silver ball that allows you to snap the Catheter Guide back into place and keep it at the correct position. Figure 31: Catheter Guide Removal - Step 2 LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
HIV, HBV, HCV, VRE, MRSA, MRSE, Pseudomonas and Mycobacterium. Wipe the transducer/cable until dry. 5.7 BATTERY - CHARGING AND PREVENTIVE MAINTENANCE IMPORTANT! For proper and safe charging, use only USB cables provided by LABORIE. 5.7.1 C HARGING THE ATTERY Urocap™...
5.8 COMPUTER VIRUS PROTECTION All computers purchased from LABORIE are virus-free before shipment and are installed with an antivirus program. It is the customer’s responsibility to correctly use and maintain the anti-virus program to prevent virus problems. LABORIE is not responsible for any virus-related computer problems after point of delivery to the customer.
The volume channel should match the pump fill volume (+/- 10 ml). If the reading is above or below the 10ml threshold, Once calibration is complete, perform a re-check by following steps 1 to 5 listed here. then calibration is required. Contact LABORIE service for assistance if problems continue after the calibration and re-check. LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
You will see an increase in EMG activity on the EMG channel. 7. Unclench your fist and return to the relaxed position. 8. Check the EMG channel for activity. at 100% strain with full contraction LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
O). If the reading is above or below the 2 cmH Once calibration is complete perform a re-check by following the instructions threshold, then calibration is required. provided above. Contact LABORIE service for assistance if problems continue after the calibration and re-check. 5.11.4 HECK THE NFUSION RANSDUCER Ensure the saline bag does not touch any other mechanical component of the system ( such as transducers, etc.) during the...
Once calibration is complete perform a re-check proximal mark in the required range of time, then calibration is required. by following the instructions provided above. Contact LABORIE service for assistance if problems continue after the calibration and re-check. 5.12 SAFELY MOVING THE CART 1.
6 TROUBLESHOOT AND SERVICE If problems continue, contact LABORIE’s Service team at 1-800-333-1039 or email service@laborie.com. Follow the instructions provided in the table below for troubleshooting steps. Symptom(s) Possible Cause(s) Check/Corrective Action(s) ALL DEVICES No response from the No power on electrical Plug the system into a known working electrical outlet.
Page 49
5. Click Run and then click Start Pump on the control panel. 6. Click Info then click Patient information. 7. Click OK. 8. Re-select Use MPUR for pump speeds. Click OK to load the correct MPUR speeds for the test. LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
Reconnect devices in the touchscreen: PC via touchscreen? 1. First disconnect all Bluetooth devices via the touchscreen. 2. Restart the software and establish a connection to the 3. Reconnect all Bluetooth devices via the touchscreen. Refer LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
Page 51
Close all open programs. Office Reporter? Make sure i-LIST Message Queue Listener is running by clicking Start > Programs > Laborie > i-LIST MsgQListener. Restart the computer and attempt to send a report to i- LIST Office Reporter. If unsuccessful try the solution below.
Page 52
BUBBLE DETECTOR Red light on sensor lit Pump tubing is not Make sure the pump tubing is placed securely into the groove even though bubbles inserted securely? on the sensor. are not visible LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
Page 53
In the section click the Change button and select the AutoHotKey Unicode application in the resulting window and then click OK: Opens with 27. The section will display the AutoHotKey application. 28. Click OK. LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
Table 19: Troubleshooting NOTE: The Laborie Aquarius LT, CT, CTS and XT are designed to be used as stand-alone cart systems, with the only mains power supply connection going through the LABORIE medical-grade power supply (LIT) in the base of the cart. Do not connect it to any other devices unless the whole system is outside the patient environment or the other device has its own medical grade power supply.
Consult the dealer or experienced technical personnel for help. WARNING! Changes or modifications not expressly approved by LABORIE could void the user’s authority to operate the equipment. This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference and (2) this device must accept any interference received, including interference that may cause undesired operation.
It Flicker Emissions IEC may be necessary to take mitigation measures, such as re- orienting 61000-3-3 or relocating the Aquarius Urodynamic Analyzer System or shielding the location. Table 21: Electronic Emissions LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
Magnetic Field typical commercial or hospital environment. (50/60 Hz) IEC 61000-4-8 NOTE is the a.c. mains voltage prior to application of the test level. Table 22: Electromagnetic Immunity LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
Page 58
If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the Aquarius Urodynamic Analyzer System b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m. LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. Table 24: Separation Distances LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
IEC 60417 – Graphic Symbols for Use on Equipment. NOTE: Sterility symbols are applicable to consumbale devices only. Refer to consumable device instructions for use for complete symbol and instructional overview. Table 25: Symbols Glossary LABORIE AQUARIUS XT/CT/LT/CTS Service Manual LBL-000233 [B]...
9 SCHEMATICS UPON REQUEST, LABORIE will make available to the customer any circuit diagrams, component part lists, and other technical documentation directly related to the AQUARIUS UDS system. 10 END-USER SOFTWARE LICENSE AGREEMENT Find licenses and terms for LABORIE products and services in the following locations: End User License Agreement: http://www.laborie.com/eula/...