LORENZ MESSTECHNIK LCV Operation Instructions Manual

LORENZ MESSTECHNIK LCV Operation Instructions Manual

Dc voltage measuring amplifier

Advertisement

Quick Links

Amplifiers
Strain Gauge Sensor-Interface LCV
Lorenz Messtechnik GmbH Obere Schlossstrasse 127/129/131
090288h_LCV
73553 Alfdorf Phone +49 7172 93730-0, Fax +49 7172 93730-22
Technical changes, errors and misprints excepted.
1/14
info@lorenz-messtechnik.de,
www.lorenz-messtechnik.de

Advertisement

Table of Contents
loading

Summary of Contents for LORENZ MESSTECHNIK LCV

  • Page 1 Amplifiers Strain Gauge Sensor-Interface LCV Lorenz Messtechnik GmbH Obere Schlossstrasse 127/129/131 090288h_LCV 73553 Alfdorf Phone +49 7172 93730-0, Fax +49 7172 93730-22 Technical changes, errors and misprints excepted. 1/14 info@lorenz-messtechnik.de, www.lorenz-messtechnik.de...
  • Page 2: Table Of Contents

    5.3.4 Filter 5.3.5 Measuring Range Adjustment Description 5.4.1 Adjusted and Calibrated LCV with Sensor, integrated or as a DC Voltage Cable Measuring Amplifier 5.4.2 Pre-adjusted LCV as a Board or Kit 5.4.3 Not adjusted LCV as a Board or a Kit 5.4.4 Adjustment to the Sensor...
  • Page 3: Imprint / Copyright

    These operating instructions do not constitute an agreed quality agreement or durability guarantee within the meaning of § 443 BGB (German Civil Code). Technical changes, errors and misprints excepted. No part of this publication may be reprinted without the written permission of Lorenz Messtechnik GmbH, Alfdorf. Important Notes / Keeping Before installing and commissioning the device, these operating instructions, and in particular the corresponding safety instructions, must be read.
  • Page 4 The devices do not contain any parts that require or can be serviced by the user. Repairs may only be carried out by Lorenz Messtechnik GmbH. If it can be assumed that safe operation of the device is no longer possible, it must be put out of operation immediately and secured against unintentional operation.
  • Page 5: Intended Use

    Intended Use Devices from Lorenz Messtechnik GmbH are to be used exclusively for measuring tasks and the di- rectly associated control tasks. Any use beyond this is considered to be improper. The valid legal and safety regulations must be observed during measurement. The instrument is not a safety component in the sense of its intended use and it is transported and stored properly.
  • Page 6: Health Protection And Safety

    The device may not be modified in terms of design or safety without the express consent of Lorenz Messtechnik GmbH. Any modification excludes any liability on our part for damages resulting there- from. Repairs and modifications are prohibited. CE Marking With the CE marking, Lorenz Messtechnik GmbH guarantees that its product meets the requirements of the relevant EC directives. Lorenz Messtechnik GmbH Obere Schlossstrasse 127/129/131 090288h_LCV 73553 Alfdorf Phone +49 7172 93730-0, Fax +49 7172 93730-22 Technical changes, errors and misprints excepted.
  • Page 7: Preamble

    A current output and a voltage output are always available, however, only one signal output variant can be adjusted. Various current and voltage variations are configured via jumpers, for subsequent changes, however, a readjustment is required. The LCV has a control input in order to activate the 100 % control signal in the sensor (if available). Power Supply Mains operation Power supply: 12V to 28VDC, min.
  • Page 8: Instruction / Description Of The Operating Mode

    Instruction / Description of the Operating Mode Connection Assignment (see also chapter ‘5.4 Adjustment Description’) Evaluation DC Voltage Measuring Amplifier LCV SG-Sensor +12 ... 28V Gain Zero Point Common GND Excitation - Out 0 ... ±10V Excitation + Signal + Out 0 ...
  • Page 9: Switching The Device On

    5.3.2 Presettings of the Sensitivity By factory default, the LCV is adjustable to approx. 2 mV/V. If the sensitivity of the sensor deviates from this value, the resistance ‘Gain rough’ must be replaced. First, the sensitivity of the sensor must be determined by the data sheet, test certificate, sensor label, etc. and multiply this value by 312.
  • Page 10: Adjustment Description

    5.4.2 Pre-adjusted LCV as a Board or Kit All sensor cables must be connected to the LCV, use only shielded cables and keep the sensor lines as short as possible. Connect the supply and the pre-adjusted current or voltage output. If the option control signal is available, it can be switched on via a positive control signal or a push-button.
  • Page 11: Interfaces And Connections

    5.5.3 Control Input, Switch-on Control Signal By the control signal on the LCV, the 100 % control signal in the sensor (if available / see sensor data sheet) can be switched on. The sensor electrically simulates a 100 % value (like 100 % load). Thus, re-adjustment and system control of subsequent evaluation systems is possible, without having to use a highly precise reference weight or torque on the sensor.
  • Page 12: Troubleshooting

    Please pack the equipment suitable for transportation The equipment may not be able to move back and forth in the package Please protect the equipment against moisture Lorenz Messtechnik GmbH Obere Schlossstrasse 127/129/131 090288h_LCV 73553 Alfdorf Phone +49 7172 93730-0, Fax +49 7172 93730-22 Technical changes, errors and misprints excepted.
  • Page 13: Commissioning And Installation

    Clean the housing with a soft and slightly moisturized cloth. While cleaning, ensure that no liquids enter the device or the connections Preventive maintenance and inspection Check the plug connections Lorenz Messtechnik GmbH Obere Schlossstrasse 127/129/131 090288h_LCV 73553 Alfdorf Phone +49 7172 93730-0, Fax +49 7172 93730-22 Technical changes, errors and misprints excepted.
  • Page 14: Equipment Disposal

    The device must be disposed of in accordance with the applicable legal regulations - see also our „General Terms and Conditions of Delivery and Sale“ at www.lorenz-messtechnik.de Reference Information 080987: Data sheet of the strain gauge sensor-interface LCV. Lorenz Messtechnik GmbH Obere Schlossstrasse 127/129/131 090288h_LCV 73553 Alfdorf Phone +49 7172 93730-0, Fax +49 7172 93730-22 Technical changes, errors and misprints excepted.