A.H. Systems SAS-521-2 Bilogical Antenna
SAS-521-2
Bilogical Antenna
Operation Manual
© A.H. Systems inc. – December 2008
1
REV C

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the SAS-521-2 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for A.H. Systems SAS-521-2

  • Page 1 ® Advanced Test Equipment Rentals www.atecorp.com 800-404-ATEC (2832) A.H. Systems SAS-521-2 Bilogical Antenna SAS-521-2 Bilogical Antenna Operation Manual © A.H. Systems inc. – December 2008 REV C...
  • Page 2: Table Of Contents

    OPERATING INSTRUCTIONS Assembly Instructions ..................... 6 Mounting Instructions....................6 Operating Instructions....................8 SPECIFICATIONS Specifications ......................9 Typical Data......................10 Calculations......................12 Convertion Factors....................13 MAINTENANCE Maintenance Procedures ..................14 Annual Calibration....................14 Warranty........................ 15 © A.H. Systems inc. – December 2008 REV C...
  • Page 3: Introduction

    A.H. Systems SAS-521-2 Bilogical Antenna INTRODUCTION The SAS-521-2 Bilogical Antenna (also known as a hybrid log periodic, biconical antenna) is a lightweight antenna that has been designed to ensure medium gain, low VSWR and high power handling capabilities. Constructed of lightweight aluminum, the SAS-521-2 Bilogical Antenna has been manufactured to operate over a very wide bandwidth.
  • Page 4: Intended Purposes

    Radio Frequency (RF) energy. It is the responsibility of the user to assure that the device is operated in a location which will control the radiated energy such that it will not cause injury and will not violate regulatory levels of electromagnetic interference. © A.H. Systems inc. – December 2008 REV C...
  • Page 5: Optional Equipment

    A.H. Systems SAS-521-2 Bilogical Antenna OPTIONAL EQUIPMENT The following is a recommend accessory list for the SAS-521-2 Bilogical Antenna: CONNECTING ACCESSORIES: PAM-0202 This preamplifier has a broad frequency range and at least 25 dB of gain. An ideal solution for improving overall system sensitivity for various emissions testing requirements.
  • Page 6: Assembly Instructions

    A.H. Systems SAS-521-2 Bilogical Antenna OPERATING INSTRUCTIONS ASSEMBLY INSTRUCTIONS The SAS-521-2 Bilogical Antenna is shipped with the triangle and window elements removed from the boom, this allows for easy transportation. The Bilogical Antenna also comes with the LPM-510 tripod mounting adapter that is already assembled onto the antenna.
  • Page 7 The horizontal beamwidth of the Bilogical is approximately 45 degrees and it should be pointed or aimed in the direction that the received signal is coming from. © A.H. Systems inc. – December 2008 REV C...
  • Page 8 (per ANSI C63.5). IMMUNITY TESTING MEASUREMENT POINT: Measurement is made from the tip of the antenna and typically preformed at a 1 meter distance. (3 meters per IEC 61000-4-3) © A.H. Systems inc. – December 2008 REV C...
  • Page 9: Specifications

    Also, portions of the cables that are not straight down or straight back will couple to the antenna fields. © A.H. Systems inc. – December 2008 REV C...
  • Page 10: Specifications

    Maximum Radiated Field................100 V/m Connector..................N-Type (female) Mounting....................1/4-20 (female) PHYSICAL SPECIFICATIONS Weight ......................4.4 lbs. (2 kg) Size (L x W x H).................37.3" x 38.5" x 22.2" (94.7cm x 98cm x 57cm) © A.H. Systems inc. – December 2008 REV C...
  • Page 11: Typical Data

    A.H. Systems SAS-521-2 Bilogical Antenna TYPICAL DATA © A.H. Systems inc. – December 2008 REV C...
  • Page 12: Calculations

    For Immunity measurements, the generated electric field strength can be calculated by: FS = Approximate Field Strength in (V/m) P = Power in watts g = Numeric Gain d = Distance in meters © A.H. Systems inc. – December 2008 REV C...
  • Page 13: Convertion Factors

    = dBµA + 34 (50Ω) dB µ V = dB µ A + 20log(Z) dBµV to dBµA dBµA = dBµV – 34 (50Ω) dB µ A = dB µ V – 20log(Z) © A.H. Systems inc. – December 2008 REV C...
  • Page 14: Maintenance

    ANNUAL CALIBRATION To ensure reliable and repeatable long-term performance, annual re-calibration of your antennas, preamplifiers and current probes by A.H. Systems experienced technicians is recommended. Our staff can calibrate almost any type or brand of antenna. It is always up to the user to determine the appropriate interval for calibration certification based on the requirements of the end users specific test/application.
  • Page 15: Warranty

    (3) years. All other products delivered under contract will be warranted for a period of two (2) years. A.H. Systems' obligation under this warranty shall be limited to repairing or replacing, F.O.B.