SU-200 series outputs an analog voltage that is directly proportional to UV-A radiation incident on a planar surface (does not have to be horizontal) where the radiation emanates from all angles of a hemisphere. SU-221 sensors output a digital signal using SDI-12 communication protocol.
SENSOR MODELS This manual covers the SDI-12 protocol UV-A sensor model SU-221 (in bold below). Additional models are covered in their respective manuals. Model Signal SU-200 0-10 mV SU-202 0- 2.5 V SU-205 0- 5 V SU-221 SDI-12 A sensor’s model number and serial number are located on the bottom of the sensor.
SPECIFICATIONS SU-221-SS Power Supply 5.5 to 24 V DC Current Draw 1.4 mA (quiescent), 1.8 (active) Calibration Factor Custom for each sensor and stored in the firmware Calibration Uncertainty ± 10 % Measurement Range 0 to 100 W m Measurement Repeatability Less than 0.5 %...
Page 8
Cosine Response Directional, or cosine, response is defined as the measurement error at a specific angle of radiation incidence. Directional error for Apogee SU-200 series UV-A sensors is approximately ± 2 % and ± 5 % at solar zenith angles of 45°...
Page 9
Spectral Response Spectral response estimate of Apogee SU-200 UV-A sensors. Spectral response was modeled from sensitivity of the photodetector and transmittance of the diffuser.
Mount the sensor to a solid surface with the nylon mounting screw provided. To accurately measure UV-A incident on a horizontal surface, the sensor must be level. An Apogee Instruments model AL-100 Leveling Plate is recommended to level the sensor when used on a flat surface or being mounted to surfaces such as wood. To facilitate mounting on a mast or pipe, the Apogee Instruments model AL-120 Solar Mounting Bracket with Leveling Plate is recommended.
CABLE CONNECTORS Apogee sensors offer cable connectors to simplify the process of removing sensors from weather stations for calibration (the entire cable does not have to be removed from the station and shipped with the sensor). The ruggedized M8 connectors are rated IP68, made of corrosion-resistant marine-grade stainless-steel, and designed for extended use in harsh environmental conditions.
OPERATION AND MEASUREMENT The SU-221 UV-A sensor has an SDI-12 output; UV radiation is returned in digital format. Measurement of SU-221 UV-A sensors require a measurement device with SDI-12 functionality that includes the M or C command. Wiring White: Positive (signal from sensor)
Page 13
“!”. SDI-12 language supports a variety of commands. Supported commands for the Apogee Instruments SU-221 UV-A sensors are listed in the following table (“a” is the sensor address. The following ASCII Characters are valid addresses: “0-9” or “A-Z”).
Page 14
Make Measurement Command: M! The make measurement command signals a measurement sequence to be performed. Data values generated in response to this command are stored in the sensor's buffer for subsequent collection using “D” commands. Data will be retained in sensor storage until another “M”, “C”, or “V” command is executed.
Page 15
0C2! 000101<cr><lf> 0D0! +20.0 0C3! 000101<cr><lf> 0D0! 0+90.2<cr><lf> where 30.0 is watts and 1.0 is mV and 20.0 is µmol Change Sensor Address: aAn! The change sensor address command allows the sensor address to be changed. If multiple SDI-12 devices are on the same bus, each device will require a unique SDI-12 address.
Page 16
UV-A Measurements and Spectral Errors Apogee Instruments model SU-200 UV-A sensors are calibrated to measure ultraviolet radiation from the sun between 300 and 400 nm in Watts per square meter. In addition to naturally occurring UV-A radiation from the sun, there are many electric light sources that emit UV-A radiation (e.g., cool white fluorescent, metal halide, mercury arc, and germicidal lamps).
MAINTENANCE AND RECALIBRATION Blocking of the optical path between the target and detector can cause low readings. Occasionally, accumulated materials on the diffuser of the SU-200 series UV-A sensors can block the optical path in three common ways: 1. Moisture or debris build-up on the diffuser. 2.
TROUBLESHOOTING AND CUSTOMER SUPPORT Independent Verification of Functionality If the sensor does not communicate with the datalogger, use an ammeter to check the current draw. It should be near 1.4 mA when the sensor is not communicating and spike to approximately 1.8 mA when the sensor is communicating.
RETURN AND WARRANTY POLICY RETURN POLICY Apogee Instruments will accept returns within 30 days of purchase as long as the product is in new condition (to be determined by Apogee). Returns are subject to a 10 % restocking fee. WARRANTY POLICY...
Page 20
84321, USA 5. Upon receipt, Apogee Instruments will determine the cause of failure. If the product is found to be defective in terms of operation to the published specifications due to a failure of product materials or craftsmanship, Apogee Instruments will repair or replace the items free of charge.