Trouble-Shooting - Endress+Hauser Nivotester FTC325 PFM Operating Instructions Manual

Level limit switch with pfm input
Table of Contents

Advertisement

Fault
Reason
Does not switch
No power
(green LED does not light up)
Nivotester defect
Contacts welded together
(after short-circuit)
Measuring transmitter defective
Switches
Incorrect setting of change-over switch for
incorrectly
limit signal in FTC325
Permanent alarm
Device configuration changed unintentionally Restore original condition
signal
Device configuration changed
Interrupt or short-circuit line to measuring
transmitter
Measuring transmitter electronics defective
Nivotester defective
Fault indicator
Probe circuit
flashes
(probe, electronic insert or two-wire cable)
Calibration not
The measuring range of the electronic insert
possible when
is exceeded in the covered condition
probe is covered
Endress+Hauser
Remedy
Check power
Replace Nivotester
Replace FTC325;
connect a fuse into contactor circuit
Replace measuring transmitter
Correctly set change-over switch behind
front panel
Press correction key for 3 s
Check line
Replace measuring transmitter electronics
Replace Nivotester
Check connecting wires
Check connection from electronic insert
to the probe
Check probe for grounding
(Rnom > 200 kOhm)
Replace electronic insert and recalibrate
After the LED goes out, recalibrate and
check function
Calibrate with uncovered probe
Check probe for grounding
en - Trouble-shooting
Note!
A new calibration is necessary:
- if the bridge position of the
EC17Z was changed
- if the electronic insert of the
probe was exchanged.