Installation; Storage; Unpacking And Inspection; Environmental Conditions - Pepperl+Fuchs HiSHPTB/32/TR-AI-01 Product Installation Manual

Table of Contents

Advertisement

PRODUCT INSTALLATION MANUAL
2. I
NSTALLATION
2.1. S
TORAGE
In case of storage of instruments and accessories, proper care should be taken to protect them from any
possible damage. Always store instruments in their sealed original packaging until they are installed.
Provide adequate protection to prevent damages that may be caused by exposure to:
Rain, excessive humidity and/or temperature excursions (inadequate sheltering).
Dust (build-up of a corrosive patina that may cause oxidations and reduce isolation).
Aggressive and polluting atmospheres with consequent corrosion.
Access by insects/rodents (damage of packing or content).
Mechanical shocks or unauthorized packing opening.
Intense vibrations (loosening of fastened parts, fatigue failures, etc.).
Any other possible risk.
Make sure the storage temperature does not exceed the limits of –20 to +70°C (-4 to 158°F) for
medium/long term storage (days/months) and -25 to +75°C (-13 to 167°F) for short term storage (a few
hours) during transportation or shipment.
2.2. U
NPACKING AND
Upon receipt of the materials, you should check the integrity of the packing and the contents. In case of
damage due to shipping, you should promptly and properly report to the shipper, supplying all necessary
information.
If the instruments are not for immediate use, we recommend that you check that all characteristics shown
on the instrument's label meet order specifications (model, supply voltage and frequency, input/output
range, certification, tag etc.) as well as the actual application requirements.
If not installed, equipment should be stored following the recommendations of the previous section.
2.3. E
NVIRONMENTAL
HIS products, like most modern electronic equipment, can operate in wide temperature and humidity
ranges. However, practical consideration suggests a guideline on operating environment for best results.
Keep operating temperature below 35°C (95° F), ideally between 20°C and 25°C (68 and 77°F),
avoiding wide and rapid temperature excursions.
Control relative humidity within 40 to 60% to avoid risks of static charges or condensation.
Limit the presence of corrosive atmosphere, fumes and dust, sealing and purifying the control room
area and using air filters in the cabinet air intakes (clean cable entry path), if necessary.
Reduce vibrations (if any) to safe levels.