1. APPLICATION
Resistance thermometers are the fundamental elements of the industrial process control. The design of the
TFP−− resistance thermometer-family measures the temperature in industrial applications. Primarily, they
are used as sensors for heated tools and bearings. The sensor tip of the TFP−5 and TFP−6 types is made from
copper, to provide faster response time. The TFP−3 and TFP−4 have process connection with a suitable mounting
bolt. The TFP−1, TFP−2, TFP−5 and TFP−6 have different types of mounting bolts that are ordered separately. The
mounting bolts allow the sensors to be screwed into various threads.
2. TECHNICAL DATA
Type
Process temperature
Sensor
Accuracy class
Measuring current
Time constant (9 / 10)
Housing material
Probe length (L)
Connection
Cable protection
Cable length
Mechanical protection
Electrical protection
Weight
2.1. A
CCESSORIES
 User's manual
 Warranty Card
 EU Declaration of Conformity
2.2. O
RDER CODES (
NOT ALL COMBINATIONS POSSIBLE
THERMOCONT
T F P –
T
C
UBE SIZE
ODE
 6
1*
 8
2*
3
M12x1.5 ( 8)
4
M8x1 ( 6)
 6, fast response
5*
 8, fast response
6*
* Mounting bolts are ordered separately
** Only with TFP-2, TFP-3, TFP-6 types
*** Only with TFP-3, TFP-4 types
Available mounting bolts (Material: DIN 1.4571 stainless steel)
Thread
TFP−5−
M10
TFP−511−0M−000−01
M12
TFP−511−0M−000−02
M12x1,5
TFP−511−0M−000−03
M14x1
TFP−511−0M−000−04
M20x1,5
TFP−511−0M−000−05
⅛"
TFP−511−0M−000−06
¼"
TFP−511−0M−000−07
⅜"
TFP−511−0M−000−08
½"
TFP−511−0M−000−09
THERMOCONT TFP−−
–50...+200 °C (–58...+392 °F)
Pt100
A or B (MSZ EN 60751), as per order codes
max. 5 mA
TFP−5, TFP−6:  35 sec.
TFP−1, TFP−2, TFP−3, TFP−4:  45 sec.
DIN 1.4571 stainless steel
max. 250 mm (9.8"), as per order codes
Teflon-coated wire, 0.35 mm
(AWG22); dual or 4-wire: 0.14 mm
2
Tinned copper-braid protective jacket
0.6...12 m (2...40 feet), as per order codes
IP54
Class III.
Max. 0.6 kg (1.3 lb) (Depending on model)
!)
S
C
ENSOR DESIGN
ODE
Class A single
1
Class B single
2
Class A dual
4**
Class B dual
5**
Class B, single, 4-wire
6
Class A, single, 4-wire
7
TFP−6−
TFP−611−0M−000−02
TFP−611−0M−000−03
TFP−611−0M−000−04
TFP−611−0M−000−05
TFP−611−0M−000−07
TFP−611−0M−000−08
TFP−611−0M−000−09
(AWG26);
2
P
C
CABLE LENGTH
ROBE LENGTH
ODE
60 mm
1
0.6 m
100 mm
2
1 m
160 mm
3
2 m
250 mm
4
3 m
10 mm***
5
6 m
30 mm***
6
12 m
C
ODE
0
1
2
3
4
5
tfp121en1804h
1 / 2

Advertisement

loading

Summary of Contents for NIVELCO THERMOCONT TFP Series

  • Page 1 Thank you for choosing a NIVELCO instrument! 1. APPLICATION Resistance thermometers are the fundamental elements of the industrial process control. The design of the TFP−− resistance thermometer-family measures the temperature in industrial applications. Primarily, they are used as sensors for heated tools and bearings. The sensor tip of the TFP−5 and TFP−6 types is made from copper, to provide faster response time.
  • Page 2 Form) must be filled and enclosed in the parcel. Download it from our website www.nivelco.com. The device must be sent back with a declaration of decontamination. A statement must be provided in the declaration that the decontamination process was successfully completed and that the device is clean from any hazardous substances.

This manual is also suitable for:

Thermocont tfp-3 seriesThermocont tfp-4 seriesThermocont tfp-5 seriesThermocont tfp-6 series