InTemp® CX600 Dry Ice/CX700 Cryogenic Logger Manual
(CX602 model shown)
InTemp CX600/CX700 and
Series Loggers
Models:
• CX602, 90-day logger,
single use
• CX603, 365-day logger,
multiple use
• CX702, 90-day logger,
single use
• CX703, 365-day logger,
multiple use
• CX703-UN, 365-day logger,
multiple use, without NIST
calibration
Required Items:
• InTemp app
• Device with iOS or
Android™ and Bluetooth
22025-K
InTemp CX600 Dry Ice and CX700 Cryogenic loggers are designed for cold shipment monitoring
and have a built-in external probe that can measure temperatures as low as -95°C (-139°F) for
the CX600 series or -200°C (-328°F) for the CX700 series. The loggers include a protective
sheath to prevent cutting the cable during shipment and a clip for mounting the probe.
Designed for wireless communication with a mobile device, these Bluetooth® Low Energy-
enabled loggers use the InTemp app, and InTempConnect® web-based software to form the
InTemp temperature monitoring solution. Using the InTemp app on your phone or tablet, you
can configure the loggers and then download them to share and view logger reports, which
include logged data, excursions, and alarm information. Or you can use InTempConnect to
configure and download CX series loggers via the CX Gateway. The InTempVerify™ app is also
available to easily download loggers and automatically upload reports to InTempConnect. Once
logged data is uploaded to InTempConnect, you can view logger configurations, build custom
reports, monitor trip information, and more. Both the CX600 and CX700 series loggers are
available in single-use 90-day models (CX602 and CX702) or multiple-use 365-day models
(CX603 or CX703).
Specifications
Temperature Probe
Sensor Range
Accuracy
Resolution
Drift
NIST Calibration
Cable Length
Probe Dimensions
Logger
Logger Operating Range
Radio Power
Transmission Range
Wireless Data Standard
Type
Duration
Shelf Life
Logging Rate
Time Accuracy
Battery
CX602 & CX603
-95° to 50°C (-139° to 122°F)
±1.5°C from -95° to -40°C (±2.7°F
from -139° to -40°F)
±1.0°C from -40° to -10°C (±1.8°F
from -40° to 14°F)
±0.5°C from -10° to 50°C (±0.9°F
from -14° to 122°F)
0.013°C at 25°C (0.023°F at 77°F)
<0.1°C (0.18°F) per year
Three-point NIST calibration,
probe only
1.5 m (4.92 ft)
Food-grade stainless steel probe with blunt tip, 31.75 mm (1.25 inches)
long, 4.78 mm (0.188 inches) diameter
-30° to 50°C (-22° to 122°F)
1 mW (0 dBm)
Approximately 30.5 m (100 ft) line-of-sight
Bluetooth Low Energy (Bluetooth Smart)
CX602 and CX702: Single use
CX603, CX703, and CX703-UN: Multiple use
CX602 and CX702: 90 days
CX603, CX703, and CX703-UN: Up to 1 year
Up to 1 year
1 second to 18 hours
±1 minute per month at 25°C (77°F)
CR2450 3V lithium, not replaceable. Battery for CX602 and CX702
models lasts for duration specified when logger is used within the 1-year
shelf life. Battery life for the CX603 and CX703 models is 1 year, typical
with logging interval of 1 minute. Faster logging intervals, remaining
connected with the InTemp app, excessive downloads, and paging may
impact battery life.
CX702 & CX703
-200° to 50°C (-328° to 122°F);
To ensure insulation integrity,
avoid flexing the wire during
exposure to temperatures below
-100°C (-148°F)
±1.5°C from -200° to -90°C (±2.7°F
from -328° to -130°F)
±1.0°C from -90° to -10°C (±1.8°F
from -328° to 14°F)
±0.5°C from -10° to 50°C (±0.9°F
from 14° to 122°F)
CX702 and CX703: Single-point
NIST calibration, probe only
CX703-UN: Calibration not
included
Need help?
Do you have a question about the CX600 and is the answer not in the manual?
Questions and answers