Should at any point you require additional help or information on the use or installation of the system please contact STS directly at: sales@safetrainingsystems.com or +44 (0) 1344 483563 16/08/2017, STS Instruments Ltd, Service & Maintenance Manual...
+44 (0) 1344 483563 3.1 ENVIRONMENTAL PARAMETERS The Siloxane Monitor is not weatherproof and must be housed in a suitable kiosk or cabinet • if sited outside of a building If siting inside a building consideration should be given to the local conditions as regards to •...
Gas supply to the instrument should be at a minimum of 200mbar • Adequate ventilation/air passage around the instrument should be provided. • 3.2 INSTRUMENTATION SAFETY SYMBOLS Caution Electrical Hazard Earth Point Hot Surface 16/08/2017, STS Instruments Ltd, Service & Maintenance Manual...
Heated Line and temperature and measure the temperature inside the insulation using a Weatherproofing temperate probe. Check the insulation for damage and degradation. If damage is present, attempt to repair that section. 16/08/2017, STS Instruments Ltd, Service & Maintenance Manual...
Page 7
Flame Arrestor venting onto a sheet of white paper. Record any residue found. If excessive residue is found replace flame arrestor. Coalescing Filter Note any Discolouration and Moisture. Replace with New Filter. 16/08/2017, STS Instruments Ltd, Service & Maintenance Manual...
Page 8
Check whether instrument intake and exhaust fans start up upon powering the instrument. If they do not, using a multi-meter, Fans measure the voltage to the fans. If they aren't getting power, return instrument to the manufacturer. 16/08/2017, STS Instruments Ltd, Service & Maintenance Manual...
Page 9
240degC. If it doesn't, return instrument to manufacturer. To Heaters check the Permeation heater, run a calibration cycle with a temperature probe in the permeation oven. This probe should stabilise around 80degC. If it doesn't return instrument to manufacture. 16/08/2017, STS Instruments Ltd, Service & Maintenance Manual...
Page 10
Via Maintenance Menu, run a calibration cycle. Leave the first cycle blank, then use 20ul of 1mg/ml D5. Attach Calibration Calibration Report. Adjust the Calibration coefficient to achieve a reading of 20ug. Replace permeation oven seal once finished. 16/08/2017, STS Instruments Ltd, Service & Maintenance Manual...
Check the filter for obvious signs of clogging, should there be any which is not removed by the nitrogen the arrestor should be replaced and the unit returned to STS for overhaul. 6.3 BIOGAS REGULATOR PRESSURE SET 16/08/2017, STS Instruments Ltd, Service & Maintenance Manual...
7.0 MAINTENANCE MENU 7.1 ACCESS TO MENU The maintenance menu can be accessed from the main menu screen: Select: OPTIONS Hold CLEAR on the number pad for approximately 4 secs. 16/08/2017, STS Instruments Ltd, Service & Maintenance Manual...
Page 13
The Maintenance menu screen is displayed Enter the password “SILOX” and press ENTER, START to confirm. Hold 5 to see the View Stats screen 16/08/2017, STS Instruments Ltd, Service & Maintenance Manual...
Page 14
1. Cleaning 2. Cooldown 3. Sample 4. Purge 5. Background preheat 6. Background 7. Measurement preheat 8. Measurement The internal variables will be displayed on the readout and will change every second. 16/08/2017, STS Instruments Ltd, Service & Maintenance Manual...
Need help?
Do you have a question about the SILOXANE and is the answer not in the manual?
Questions and answers