Depending on the measurement mode, some of the following settings can be changed by the user:
•
Source mode: Choose between Auto off or Always on. More details on the source behavior of
Bode 100 can be found in
•
Source level: Set the constant source level or the reference level when the shaped level function
is used.
•
Receiver bandwidth: Select the receiver bandwidth used for the measurement.
•
DUT settling time: Time between applying the measurement signal and starting to measure at the
receiver. A DUT settling time is needed if the DUT introduces a significant time delay before the
output reaches its final reaction to the input stimulus.
•
Attenuator 1 and Attenuator 2: Select the receiver input attenuation.
•
Reference switch: Receiver 1 and Receiver 2 can be connected either internally or externally,
depending on the measurement mode.
•
Termination switch: When switched on, the Channel 1 and Channel 2 inputs are terminated with
50 Ω. If the switch is open, the input impedance is 1 MΩ.
•
Probe 1 and Probe 2: Some measurement modes allow you to manually enter a probe factor. For
more details on using probes see
OMICRON Lab
9.4 Signal source settings
9.5 Using external probes
Measurement and Device settings
on page 80.
on page 83.