Page 1
H04X10 Multi-position magnetic stirrer with heating -10 positions S04X10 Multi-position magnetic stirrer without heating -10 positions Please read the User Manual carefully before use, and follow all operating and safety instructions! user manual english...
User Manual S04X10 Multi-position magnetic stirrer without heating - 10 positions H04X10 Multi-position magnetic stirrer with heating - 10 positions Preface Users should read this Manual carefully, follow the instructions and procedures, and beware of all the cautions when using this instrument.
1. Safety Instructions Warning! • Read the operating instructions carefully before use. • Ensure that only trained staff works with the instrument. Risk of burn! • Caution when touching the housing parts and the heating plate. The heating plate can reach temperatures of 120 °C. •...
• Observe the minimum distances between the devices, between the device and the wall and above the assembly (min. 100 mm) >100mm >100mm Figure 1 2. Proper Use The instrument is designed for mixing liquids in schools, laboratories, or factories. This device is not suitable for using in residential areas or other constraints mentioned in Chapter 1.
Page 5
4. Control 4.1 Control elements Power LED Stirrer point Work plate Heating Knob Stirrer knob Figure 2 Power switch Figure 3 Items Descriptions -Set the rated rotary speed in the safe stirring range from 0 to 1100 rpm. -Set the rated temperature from RT to 120 ºC. Speed/temperature -The function “Stirring”, and “Heating”...
Page 6
5. Trial Run / Operation • Put the device on the stable and safe place. • Make sure that the required operating voltage and power supply voltage match. • Ensure that socket must be earthed. • Plug in the power cable, ensure the power is on. •...
7. Maintenance and Cleaning • Proper maintenance can keep instruments working properly and lengthen its lifetime. • Do not spray cleanser into the instrument when cleaning. • Unplug the power line when cleaning. • Only use recommended cleansers: Dyes Isopropyl alcohol Construction materials Water containing tenside / Isopropyl alcohol...
9. Technical data Items Parameters Voltage [VAC] 200-240/100-120 Frequency [Hz] 50/60 Power [W] Stirring point position quantity Max. stirring quantity of single stirring position O) [l] Stirring position interval [mm] Single stirring position speed error [%] Max. stirring quantity (H O) [l] Max.
Need help?
Do you have a question about the H04X10 and is the answer not in the manual?
Questions and answers