  • Page 1 LBX INC-R Refrigerated Incubator Please read the User Manual carefully before use, and follow all operating and safety instructions! user manual english...
  • Page 2: Warranty

    User Manual LBX INC-R Refrigerated Incubator Preface Users should read this Manual carefully, follow the instructions and procedures, and beware of all the cautions when using this instrument. Service If help is needed, you can always contact your dealer or Labbox via www.labbox.com (declare an incident).
  • Page 3: Structural Features

    Structural Features 1. The case is constructed from high-quality cold-rolled steel plate with a plastic coating on the surface. 2. Mirror-stainless working room, easy to clean, and with adjustable shelves. 3. Equipped with a Micro-Computer Temperature Controller and an Automatic defrosting system, along with a large LCD screen.
  • Page 4: Temperature Graph

    Temperature Graph Note: The specifications mentioned above have been tested in a 25 ºC environment. Working Conditions 1. Ambient temperature: 5 – 32 ºC (if the set temperature ≦ 10 ºC, ambient ≦ 28 ºC). 2. Humidity: ≦ 80 % RH. 3.
  • Page 5: Controller Specifications

    Controller Operation Instructions 1. Controller Specifications • Temperature setting range: 0.0 – 60.0 ºC • Timer setting range: 0 – 99 minutes 59 seconds • Display error: < 0.5 % • Sensor: PT100 2. Panel Instructions 3. Temperature-Time Operation Example: setting the temperature to 30 ºC. Once reached, stop the instrument 5 hours later (T = 30 ºC; timer = 5 hours).
  • Page 6: Internal Parameters

    4. Inner Chamber Specification Setting Press the set button for more than 3 seconds, and the screen will display “LC.” Change the number to “3” using the up and down keys, then click the set button again to access the parameters. By clicking the set button, you can navigate through the parameters in the following table and modify each specification.
  • Page 7 Check Ambient Temperature Press the set button for more than 3 seconds; it will display “LC.” Change it to “18” using the up and down keys. Click the set button, and the screen will show the ambient temperature. Press the set button for 3 seconds to exit this screen, and the specification will be saved automatically.
  • Page 8: Fault Analysis

    Fault Analysis Problem Cause Analysis Solution Power button is not lighting up No power Inspect the plug Broken fuse Change the fuse Temperature controller displays Broken sensor Change the sensor Broken controller Change the controller “□□□□” Evaporator frost or chamber with Frequent opening of the door If the temperature exceeds 50 frost...