Download Print this page

AMALGAMATED INSTRUMENT PM5-TR Operation And Instruction Manual

Ratemeter/totaliser panel mount display/controller

Advertisement

Quick Links

Panel Mount Display/Controller
Operation and Instruction Manual
AMALGAMATED INSTRUMENT CO
Unit 5, 28 Leighton Place Hornsby
NSW 2077 Australia
Model PM5-TR
Ratemeter/Totaliser
Telephone: +61 2 9476 2244
Facsimile: +61 2 9476 2902
ABN: 80 619 963 692
e-mail: sales@aicpl.com.au
Internet: www.aicpl.com.au

Advertisement

loading
Need help?

Need help?

Do you have a question about the PM5-TR and is the answer not in the manual?

Questions and answers

Summary of Contents for AMALGAMATED INSTRUMENT PM5-TR

  • Page 1 Model PM5-TR Ratemeter/Totaliser Panel Mount Display/Controller Operation and Instruction Manual AMALGAMATED INSTRUMENT CO ABN: 80 619 963 692 Unit 5, 28 Leighton Place Hornsby Telephone: +61 2 9476 2244 e-mail: sales@aicpl.com.au NSW 2077 Australia Facsimile: +61 2 9476 2902 Internet: www.aicpl.com.au...
  • Page 2: Table Of Contents

    Table of Contents 1 Introduction 2 Mechanical Installation 3 Electrical installation 4 Function tables - summary of setup functions 5 Explanation of functions 6 Specifications 7 Guarantee and service 2 of PM5TRMAN-1.7-0...
  • Page 3: Introduction

    Introduction General description This manual contains information for the installation and operation of the the PM5-TR Monitor. The instrument may be set to operate as a ratemeter or totaliser or allow toggling between rate and total displays. A “grand total” display is also viewable and can be separately reset. The OPEr function allows selection of one of these three modes.
  • Page 4 2. USER - with this method the user enters two frequencies via the pushbuttons and applies a scale value to each frequency. The CAL IN 1 and CAL SCL.1 functions are used to set the first frequency and the CAL IN 2 and CAL SCL.2 functions are used to set the second frequency. The display takes these two points as points in a straight line for calculating and displaying inputs above, between and below these two points.
  • Page 5 Remote input mode - this uses the Easy method of access but also requires the use of a remote input switch. PIN 1 mode - this method allows a PIN to be set with access via PIN entry. PIN 2 mode - this method also requires a PIN and would generally be use to allow a higher access level than the first PIN.
  • Page 6 CodE. When this function is seen the PIN value set at the USR.1 Pin function must be entered via the ^ or v pushbuttons followed by pressing F to accept the PIN before the user can progress to the setup functions. PIN 1 If a PIN has been set the Press and...
  • Page 7 The setup functions are organised in blocks or sections e.g. all the settings for channel 1 alarms are in the AL1 section. Once access to setup functions has been gained use the ^ and v buttons to select the section required then press F to enter this section and again us the ^ and v buttons to select the required function for alteration and press F to allow alteration of this function.
  • Page 8 The example in the flowchart (for 4 digit display) below shows the method using alarm relay 1 setup function. Alarm 2 Settings for Use ^ or v to adjust setting section alarm 2 to any setting or press Press F Press F to ^ and v together to disable Followed by...
  • Page 9: Mechanical Installation

    Mechanical Installation Choose a mounting position as far away as possible from sources of electrical noise such as motors, generators, fluorescent lights, high voltage cables/bus bars etc. An IP67 access cover which may be installed on the panel and surrounds is available as an option to be used when mounting the instrument in damp/dusty positions.
  • Page 10: Electrical Installation

    Electrical installation Electrical installation The PM5 Panel Meter is designed for continuous operation and no power switch is fitted to the unit. It is recommended that an external switch and fuse be provided to allow the unit to be removed for servicing. The plug in, screw type, terminal blocks allow for wires of up to 2.5mm to be fitted for power, relays and options and 1mm for sensor and other wiring.
  • Page 11 Input connections The display will accept many common input sensor types on inputs 1 and 2. Input 1 is used as the rate/total input pulse terminal. Input 2 (terminal 8) can be used to change the totaliser count direction i.e. change from count up (input 2 inactive) or to count down (input 2 active). The totaliser direction function dir must be set to UP.dn when this count up/down control is required.
  • Page 12 PULL UP Set to Lo bIAS Set to OFF Lo F Set to OFF HYSt Set to OFF for very low amplitude inputs (25mV to 250mV) or On for inputs with amplitude 250mV to 48VDC or RMS Attn Set to OFF for no attenuation or On to reduce the input amplitude by a factor of 5 dC Set to On for input frequencies less than 10 Hz, otherwise set to OFF EdgE Select Lo (trigger from low going edge) or Hi (trigger from rising edge) as required d.bnc Normally only required for contact closure e.g.
  • Page 13 Contact closure input CONTACT CLOSURE INPUT PM5 TERMINALS 9 Input 1 10 GND IN1 function settings for contact closure input PULL UP Set to High bIAS Set to On Lo F Set to On HYSt Set to On Attn Set to OFF for no attenuation or ON to reduce the input amplitude by a factor of 5 dC Set to On EdgE Select Lo (trigger from low going edge) or Hi (trigger from rising edge) as required d.bnc Normally only required for contact closure e.g.
  • Page 14 NPN 2 wire sensor input PM5 TERMINALS NPN 2 WIRE INPUT 9 Input 1 10 GND IN1 function settings for NPN 2 wire sensor input PULL UP Set to High bIAS Set to On Lo F Set to OFF HYSt Set to On Attn Set to OFF for no attenuation or ON to reduce the input amplitude by a factor of 5 dC Set to On EdgE Select Lo (trigger from low going edge) or Hi (trigger from rising edge) as required...
  • Page 15 dC Set to On EdgE Select Lo (trigger from low going edge) or Hi (trigger from rising edge) as required d.bnc Normally only required for contact closure e.g. relay or switch input. Select a debounce time which allows operation without false triggering due to contact bounce PNP 2 wire sensor input PNP 3 WIRE INPUT PM5 TERMINALS...
  • Page 16 PULL UP Set to Lo bIAS Set to On Lo F Set to OFF HYSt Set to On Attn Set to OFF for no attenuation or ON to reduce the input amplitude by a factor of 5 dC Set to On EdgE Select Lo (trigger from low going edge) or Hi (trigger from rising edge) as required d.bnc Normally only required for contact closure e.g.
  • Page 17: Function Tables - Summary Of Setup Functions

    Function tables - summary of setup functions Note: the order in which the functions appear on the display may not be exactly as shown below. The availability and order of functions is determined by choice of function settings and options fitted.
  • Page 18 Relay function table Display Function Range Default Your Ref/Page record RL 1 to n.o, n.c Alarm relay x action to normally 5.12 RL 7 open (de-energised) or normally 4.16 closed (energised) RL 1 to OFF or ON Relay acknowledge 5.13 RL 7 4.16 RL 1 to...
  • Page 19 RAtE NONE, FLSH, NONE Warning message for low and high 5.25 dISP - - - -, OR.UR, display Hi.Lo RAtE Lo.F, Hi.F, Hi.F Rate/frequency range 5.26 RNgE AUgE, R.AUg RAtE 1 to 3200 secs Timeout for low frequency input 5.27 t.Out RAtE 1 to 5000 secs...
  • Page 20 UP, doWn, Totaliser count up/down operation 5.43 UP.dn NonE, For, NonE Grand total operation mode 5.44 g.tot REu, PoS, NEg, AbS Total low value visual warning Any display 5.45 value or OFF dISP Total high value visual warning Any display 5.46 value or OFF dISP...
  • Page 21 Input function tables Display Function Range Default Your Ref/Page record IN 1 to OPEn, High, High Input pull up 5.57 IN 2 PULL IN 1 to OFF, On Input bias 5.58 IN 2 bIAS IN 1 to OFF, On Input low frequency 5.59 IN 2 Lo F...
  • Page 22 RO 1 Analog output 1 option low display Any display 5.81 value ( Optional) value RO 1 1000 Analog output option high display Any display 5.82 High value ( Optional) value RO 1 Analog output 1 PI control Any display 5.83 SEtP setpoint ( Optional)
  • Page 23 RO 2 0.0 to 100.0 50.0 Analog output 2 PI control bias 5.101 biAS ( Optional) 4.11 P button and remote inputs function table Display Function Range Default Your Ref/Page record R.INP NONE, P.Hi, NONE Front P button operation mode 5.102 P.but P.Lo, Hi.Lo,...
  • Page 24 SErI RAtE, totL, RAtE Serial operation channel 5.108 g.tOt ( Optional) SErI 1200, 2400, 9600 Serial baud rate ( Optional) 5.109 bAud 4800, 9600, 19.2, 38.4, 57.6, 115.2 SErI 8 N, 8 E, 8 O, 7 Serial parity ( Optional) 5.110 PrtY E, 7 O...
  • Page 25 4.15 Access control function table Display Function Range Default Your Ref/Page record ACCES NONE, 1, 2, 3, NONE Easy access mode 5.114 EASY 4, 5, 6, CAL LEUL ACCES NONE, 1, 2, 3, NONE Remote input access mode 5.115 R.INP 4, 5, 6, CAL LEUL ACCES...
  • Page 26 4.16 Relay table Record your relay settings in the table below Display Alarm 1 Alarm 2 Alarm 3 Alarm 4 Alarm 5 Alarm 6 Alarm 7 Alarm 8 High HYSt triP P.SEC P.SEC RLYS OPEr Ltch Record which relays are allocated to which alarms and other relay settings in the table below Display Relay 1 Relay 2...
  • Page 27: Explanation Of Functions

    Explanation of functions The setup and calibration functions are configured through a push button sequence. The three push buttons located at the front of the instrument are used to alter settings. The access modes available are detailed in section 1.2, starting on page 4. Display messages shown are those which would appear on a 6 digit display, these display messages may in some cases vary slightly for other display types.
  • Page 28 Alarm relay low setpoint AL 1 to AL 8 Section: Display: Any display value or OFF Range: Default Value: Default Access Level 4010 to 4017 Function number Displays and sets the low setpoint value for the designated alarm. Use this low setpoint function if a relay operation is required when the display value becomes equal to or less than the required setpoint value.
  • Page 29 Alarm hysteresis (deadband) AL 1 to AL 8 Section: HYSt Display: 0 to 50000 Range: Default Value: Default Access Level 4020 to 4027 Function number Displays and sets the alarm hysteresis limit for the designated alarm. To set a alarm hysteresis value go to the function and use the ^ or v push buttons to set the value required then press F to accept this value.
  • Page 30 Alarm relay reset time AL 1 to AL 8 Section: Display: 0 to 5000.0 secs Range: Default Value: Default Access Level 4050 to 4057 Function number Displays and sets the alarm reset delay time in seconds. The reset time is common for both alarm high and low setpoint values.
  • Page 31 Relay selection AL 1 to AL 8 Section: RLYS Display: On or OFF Range: Default Value: Default Access Level 4330 to 4337 Function number Allows a relay to be allocated to an alarm. For example if a high alarm value has been selected at the AL1 RLYS function this alarm could be allocated to relay 3 by selecting RLY 3 On at this function.
  • Page 32 In latching mode the alarm annunciator (5 digit display type only) will flash when the display goes into alarm condition. If the display goes out of alarm condition without being acknowledged the flashing period will change to give a longer “off” time. If the alarm is acknowledged by pressing the F button then the annunciator will change from flashing to solidly lit.
  • Page 33 2. If alarms 1, 2 and 3 all use relay 1 and relay 1 is set for And operation then relay 1 will activate if the display value for the selected channels for these alarms causes alarm 1 and alarm 2 and alarm 3 to go into alarm condition.
  • Page 34 5.18 Rate display scaling method RAtE Section: USER CAL Display: INPt, USER, LIUE Range: INPt Default Value: Default Access Level 4C4A Function number Allows selection of frequency scaling method required. Three methods are available: INPt - this method uses INPt and SCL functions to scale the display. USER - this method allows the user to enter two frequencies in Hz (RAtE CAL IN 1 and RAtE CAL IN 1 and enter a scaling value for each frequency (RAtE CAL SCL 1 and RAtE CAL SCL 2) to provide the rate display scaling.
  • Page 35 5.21 Ratemeter display filter RAtE Section: FLtR Display: 0, 1, 2, 3, 4, 5, 6, 7, 8 Range: Default Value: Default Access Level 43C0 Function number This function is only seen when RAtE or both mode is selected. Digital filtering uses a weighted average method of determining the display value and is used for reducing display value variation due to short term interference.
  • Page 36 5.24 Rate high value visual warning RAtE Section: Hi dISP Display: Any display value or OFF Range: Default Value: Default Access Level 4560 Function number This function is only seen when RAtE or both mode is selected. Rate display high overrange limit value - the display can be set to show an overrange message if the display value falls below the Hi dISP setting.
  • Page 37 a period of seconds set by the A.SEc function. An averaged display is particularly useful when the input is irregular. By averaging the pulses over a period of time the display will give a more stable reading for these irregular pulses. Select R.AUg for a “rolling averaged”...
  • Page 38 5.29 Counts for average rate display RAtE Section: A.cnt Display: 0 to 30 Range: Default Value: Default Access Level 4C70 Function number Only seen if R.AUg is selected at the RNgE function. Sets the number of time periods counted over which the rolling average display will be calculated.
  • Page 39 Rate USER scaling mode frequency 2 5.32 RAtE Section: CAL IN 2 Display: Range: Any display value Default Value: Default Access Level 46F0 Function number Second frequency input value for USER scaling mode. See the CAL IN 1 function for explanation. Rate USER scaling mode scale 2 5.33 RAtE...
  • Page 40 Press F, the message RAte CAL1 SCLE will be seen followed by a value. Use ^ or v to set the display value required for the first input frequency then press F. The message CAL End should be seen to indicate that the first scaling point is increased. It is necessary to enter two scaling points so repeat the process using a different frequency and scale selection at the CAL2function.
  • Page 41 5.38 Decimal point setting for totaliser Section: d.Pnt Display: 0, 0.1, 0.02, 0.003 Range: Default Value: Default Access Level 4102 Function number This function is only seen when tOtL or both mode is selected a the. Displays and sets the decimal point position for the total display.
  • Page 42 rounding is useful for reducing the instrument resolution without loss of accuracy in applications where it is undesirable to display to a fine tolerance. Example: If set to 10 the display values will change in multiples of 10 only i.e. display moves from 10 to 20 to 30 etc. 5.42 Truncation operation Section:...
  • Page 43 Mode Up Count Down Count NonE No effect No effect The grand total will increase with each up The grand total will decrease with each down count input pulse. The grand total can show count input pulse. The grand total can show positive and negative totals.
  • Page 44 message can be used to alert operators to the presence of a total which is above the high limit. If this function is not required it should be set to OFF by pressing the ^ and v buttons simultaneously at this function.
  • Page 45 is powered up. If this function is set to OFF then the power up will have no effect on the grand total. 5.50 Total preset value Section: P.SEt Display: Range: Any display value Default Value: Default Access Level 4C60 Function number This function is only seen when tOtL or both mode is selected.
  • Page 46 5.53 Bargraph channel bArg Section: Display: RAtE, totL, g.tOt Range: RAtE Default Value: Default Access Level 43F6 Function number Seen only in bargraph display instruments. Displays and sets the channel/input type for the bargraph display to operate from. Choices are: RAtE - bargraph shows the rate value totL - bargraph shows the totalised value 5.54...
  • Page 47 display is indicating actual process units. To set bargraph low level go to the bArG Lo function and use the ^ or v push buttons to set the value required then press F to accept this value. 5.56 Bargraph high value bArG Section: Display:...
  • Page 48 5.59 Input low frequency IN 1 to IN 2 Section: Lo F Display: OFF, On Range: Default Value: Default Access Level 4C04 to 4C05 Function number Allows application of a low pass filter setting for the signal input 1 and count up/down input 2. When set to On an low pass filter will be applied to the input circuit.
  • Page 49 5.62 Input DC coupling IN 1 to IN 2 Section: Display: OFF, On Range: Default Value: Default Access Level 4C0A to 4C0b Function number Allows application of DC coupling for the signal input 1 and count up/down input 2. When set to OFF no DC coupling be applied.
  • Page 50 5.65 Output voltage selection P.Out Section: P.Out Display: 5U, 12U or 18U Range: Default Value: Default Access Level 4C0F Function number Allows selection of the output voltage where available on the input board. 5V 12V or 18VDC (25mA max.) is available as transmitter supply this function allow selection of 5V (5 U) or 12V (12U) or 18V (18U).
  • Page 51 5.68 Analog output 1 PI control on or off RO 1 Section: P.CtI Display: NO or YES Range: Default Value: Default Access Level 4600 Function number Allows selection of retransmission (No) or PI control analog output (YES). If set to No then the analog output will operate as a retransmission output using the limits set at the Lo and HiGH functions.
  • Page 52 or v push buttons to set the required value then press F to accept this selection. 5.71 Analog output 1 PI control proportional gain RO 1 Section: Display: Range: Any display value 1.000 Default Value: Default Access Level 4620 Function number Allows selection of the PI control proportional gain.
  • Page 53 To set the value go to the I.H function, press F and when you see a digit of the value flash use the ^ or v push buttons to set the required value then press F to accept this selection. 5.74 Analog output 1 PI control integral low limit RO 1...
  • Page 54 ^ or v push buttons to set the required value then press F to accept this selection. Example:If it is required to retransmit 4mA when the display indicates 0 then select 0 in this function using the ^ or v button. 5.77 Analog output option high value RO 1...
  • Page 55 5.79 Input selection for analog output 1 RO 1 Section: InPt Display: RAtE, totL, g.tOt Range: RAtE Default Value: Default Access Level 43E0 Function number Seen only when an analog retransmission option fitted. Sets the input from which the first analog output will operate.
  • Page 56 using the ^ or v button. 5.82 Analog output option high value RO 1 Section: High Display: Range: Any display value 1000 Default Value: Default Access Level 4130 Function number Seen only when analog retransmission option fitted. Refer to the separate “PM5 Meter Optional Output Addendum”...
  • Page 57 PI control output” chapter for a full description of the analog PI control functions. To set the value go to the SPAn function, press F and when you see a digit of the value flash use the ^ or v push buttons to set the required value then press F to accept this selection. 5.85 Analog output 1 PI control proportional gain RO 1...
  • Page 58 PI control output” chapter for a full description of the analog PI control functions. To set the value go to the I.H function, press F and when you see a digit of the value flash use the ^ or v push buttons to set the required value then press F to accept this selection. 5.88 Analog output 1 PI control integral low limit RO 1...
  • Page 59 0-1.0 for 0 to 1VDC output 0-10 for 0 to 10VDC output To set the selection go to the OutP function, press F and when you see a digit of the value flash use the ^ or v push buttons to set the required value then press F to accept this selection. 5.91 Input selection for analog output 2 RO 2...
  • Page 60 5.94 Analog output option 2 high value RO 2 Section: High Display: Range: Any display value 1000 Default Value: Default Access Level 4131 Function number Seen only when dual analog retransmission option fitted. See function RO1 High for further details. 5.95 Analog output 2 PI control setpoint RO 2...
  • Page 61 5.98 Analog output 2 PI control integral gain RO 2 Section: Display: -32.768 to 32.767 Range: 0.000 Default Value: Default Access Level 4629 Function number Allows selection of the PI control integral gain. See function RO1 I.G for further details. 5.99 Analog output 2 PI control integral high limit RO 2...
  • Page 62 5.102 Front P button operation mode R.INP Section: P.but Display: NONE, P.Hi, P.Lo, Hi.Lo, AL.Ac, RSEt, g.rSt, PSEt Range: NONE Default Value: Default Access Level 4720 Function number Sets the operation mode for front P button. Functions available are identical to the same functions used in the R.IN.1 function.
  • Page 63 to that alarm to reset when the alarm condition is removed. If the alarm is set for automatic reset the acknowledgment will allow the alarm and any relays allocated to that alarm to reset even if the alarm condition still exists this could typically be used to silence a siren controlled by a relay even though the alarm condition is still present.
  • Page 64 readability in dark areas or to reduce the power consumption of the instrument. See also the duII function. To set brightness level go to the brgt function press F and when you see a digit of the value flash use the ^ or v push buttons to set the required value then press F to accept this selection.
  • Page 65 M.buS - Modbus RTU. ChAn - Operates in a similar manner to Cont mode but only outputs the value chosen at the SErI Ch function. 5.108 Serial operation channel SErI Section: Display: RAtE, totL, g.tOt Range: RAtE Default Value: Default Access Level 4485 Function number Allows selection of either rate, total or grand total for serial comms.
  • Page 66 5.111 Serial address SErI Section: Unit Addr Display: 1 to 127 Range: Default Value: Default Access Level 0430 Function number Allows selection of the unit address when the operation is set for POLL mode. The unit address is offset by 32(DECIMAL) to avoid clashing with ACSII special characters, therefore 42 (DECIMAL) or 2A (HEX) would be unit address 10.
  • Page 67 5.114 Easy access mode ACCES Section: EASY LEUL Display: NONE, 1, 2, 3, 4, 5, 6, CAL Range: NONE Default Value: S.CAL Default Access Level 0C00 Function number Allows choice of the access level available when using the easy access method. For example if this function is set to 3 then functions with levels 1, 2 and 3 can be viewed and changed when access to setup functions is made using this method.
  • Page 68 5.117 PIN code 1 access level ACCES Section: USR.1 LEUL Display: NONE, 1, 2, 3, 4, 5, 6, CAL Range: NONE Default Value: S.CAL Default Access Level 0C02 Function number This function allows choice of the access level available when using the PIN code 1 input access method. To access setup functions using the PIN code 1 input access method press and hold the F button until the message FUNC is seen followed by the first function message, this should take approximately 3 seconds.
  • Page 69 5.120 User assignable access 1 function number ACCES Section: Fn.1 CodE Display: 0000 to FFFF hex. Range: 0000 Default Value: S.CAL Default Access Level 0C10 Function number In addition to being assigned an access level each setup function is assigned an individual function number. This functions and the ones which follow (Fn.2 CodE etc.) can be used to alter the access level for particular functions.
  • Page 70 5.124 User assignable access 3 function number ACCES Section: Fn.3 CodE Display: 0000 to FFFF hex. Range: 0000 Default Value: S.CAL Default Access Level 0C12 Function number This function allows as third function access change and operates in the same manner as . Enter the function number required and then enter the new access level at the function which follows.
  • Page 71 5.128 Error Messages - - - - or -or-, or -or- or -HI- or -LO- or display value flashes These messages could indicate that a high or low rate or total visual alarm warning has been selected e.g. if the rate Hi dISP value is exceeded then the display will show one of the above warning messages as set by the rate dISP function.
  • Page 72: Specifications

    Specifications Technical specifications Input : Selectable for most sensor types. For inductive, AC and square wave inputs the maximum input voltage is 48VDC or RMS with appropriate input setting. Input 2 can be used to control up/down totalising. Ambient temperature: LED -10 to 60 C, LCD -10 to 50 C Humidity: 5 to 95% non condensing...
  • Page 73: Guarantee And Service

    Guarantee and service The product supplied with this manual is guaranteed against faulty workmanship for a period of two years from the date of dispatch. Our obligation assumed under this guarantee is limited to the replacement of parts which, by our exami- nation, are proved to be defective and have not been misused, carelessly handled, defaced or damaged due to incorrect installation.