Page 1
In the meantime our company has grown into one of the world's leading producers of sterilization equipment. Verification of it success has been the sale of more than 450,000 MELAG units worldwide, attesting to the high quality of our sterilizers, which are exclusively made in Germany.
Page 2
Safety instructions To operate the CF Card Reader, pay attention to the safety instructions given below and in the individual chapters of this Operating Manual. ■ Do not open the housing of the CF Card Printer. Improper opening and repair could impair the electrical safety and pose danger to the user.
Scope of delivery ..............................3 Chapter 1 – Performance Description ......................... 4 Proper use ................................4 You can connect the following autoclaves to the MELAflash CF Card Printer ............4 Chapter 2 – Device Description ..........................4 MELAflash CF Card Printer Views ........................... 4 LED Status Displays ..............................
Chapter 1 – Performance Description MELAflash CF Card Printer Chapter 1 – Performance Description Proper use The MELAflash CF Card Printer serves for storage of sterilization logs on the MELAflash CF card, which can be read out over the standard accessory MELAflash Card Reader on the practice computer.
Chapter 2 – Device Description MELAflash CF-Card-Schreiber LED Status Displays Power Data Status Measures green LED red LED ● ● ● CF Card Printer Insert CF card ready for operation lights up ● ● Access to the CF card Do NOT remove the CF card...
For more detailed information about connecting the serial cable, consult the Operating Manual. Insert the MELAflash CF card Insert the delivered MELAflash CF card into the slot on the front of the MELAflash CF Card Printer. Make sure the card is facing the right direction and do not use force to insert it into the slot.
Initialise the MELAflash CF Card Printer on the autoclave In order to be able to save sterilization logs on the MELAflash CF Card via the CF Card Printer, you have to set the MELAflash CF Card Printer as output medium on the autoclave.
Page 8
If the option is set to NO, press the (P) to switch from Immediate Print NO to YES. In order to exit the menu, press the (S) twice. The MELAflash CF Card Printer is now ready for operation and sterilization logs can be saved on the CF card.
Chapter 4 – Logging Chapter 4 – Logging Follow the following instructions for using a CF card: ■ Never insert the CF card into the card slot with force. Warning! ■ Never draw the CF card out from the card slot during a write and read access procedure.
MELAview using fictive sterilization logs as examples. Please notice that the program will be deleted in case of formatting the CF card. Your MELAG dealer can also provide you with a demo version on CD ROM. Save preferred path In order to rapidly open preferred folders for sterilization logs, it is possible to save the respective path as a favourite.
Page 11
Chapter 4 – Logging Open the directory on your PC where the logfiles are stored. Double-click on the logfile. If the respective file extension is unknown, the dialog window depicted on the left opens up: Choose "Select program from a list" ...
Chapter 5 – Update Chapter 5 – Update Performing updates on the MELAflash CF card printer is possible from Update as of MELAflash version 2.000. Software version 2.000 The software version currently installed in the MELAflash CF card printer is listed on the end of the log print-out: In order to perform an update, proceed as follows: Save the update file (11352.MOT) in the main directory of an empty...
Chapter 6 – Error conditions Chapter 6 – Error conditions Before saving the data, the MELAflash CF Card Printer verifies the log data transferred from the autoclave. If a log file is not recognised, the system creates a directory UNKNOWN Unknown log file on the CF card.
MELAflash Card Reader 01048 MELAflash CF Card 01043 Further information on the documentation and release is provided in the download area of the MELAG website under the key term Laws & Ordinances. MELAG Medical Technology Geneststraße 6-10 E-Mail: info@melag.de Rev.: 1 – 11/1213 10829 Berlin, Germany www.melag.de...