Kinetik Wellbeing JPD-FR409-BT Instruction Manual

Infrared thermometer
Hide thumbs Also See for JPD-FR409-BT:

Advertisement

Quick Links

Authorized European Representative:
MedPath GmbH
Mies-van-der-Rohe-Strasse 8, 80807 Munich, Germany
Shenzhen Jumper Medical Equipment Co., Ltd
D Building, No. 71, Xintian Road, Fuyong Street,
Baoan, Shenzhen, Guangdong, P. R. China 518103
IP22
0482
Made in China
Kinetik JPD-FR409-BT EN IB 20221018
Infrared Thermometer
Instruction Manual
0482
JPD-FR409-BT

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the JPD-FR409-BT and is the answer not in the manual?

Questions and answers

Summary of Contents for Kinetik Wellbeing JPD-FR409-BT

  • Page 1 Made in China Mies-van-der-Rohe-Strasse 8, 80807 Munich, Germany Shenzhen Jumper Medical Equipment Co., Ltd D Building, No. 71, Xintian Road, Fuyong Street, Baoan, Shenzhen, Guangdong, P. R. China 518103 Infrared Thermometer IP22 Instruction Manual JPD-FR409-BT 0482 Kinetik JPD-FR409-BT EN IB 20221018...
  • Page 2 Product Information Support Product Name: Infrared Thermometer Model: JPD-FR409-BT Introduction Thank you for purchasing this Infrared Thermometer. Please read the User Manual carefully to make sure safe and proper use of this thermometer. Our Customer Care team are available from 9am-5pm, Monday to Please read and fully understand the Safety Precautions before use.
  • Page 3: Table Of Contents

    Contents Package Contents Package Contents..............Safety Precautions.
  • Page 4 Safety Precautions Process" chapter and carefully operate the thermometer when measuring children's temperature. Read the following precautions carefully before using the Do not immerse the thermometer in water or other liquid, as it is thermometer. not waterproof. Clean and disinfect the thermometer as described in the "Cleaning and Disinfection"...
  • Page 5 Warning Do not use the thermometer in an environment where flammable anesthetic mixture with air or with oxygen or nitrous oxide is Warning available. Do not force the temperature probe of the thermometer into an ear canal. Otherwise, the ear canal may get injured. Symbols Keep the thermometer out of the reach of children.
  • Page 6: Basics

    Body Temperature Basics Product Description Generally, you can take a body temperature on the forehead, in the ear 1) Overview The Smart Ear & Forehead Thermometer measures the body temperature canal, under the armpit, in the mouth, or in the anus. The temperature based on the infrared energy emitted from the eardrum or the forehead.
  • Page 7 5) Contraindications Product Structure Do not use the thermometer if the ear is infected with otitis or suppuration Features Good safety ● Passive infrared receiving technology. Easy operation ● Ergonomic design ● One button measurement Quick measurement ● 1-second measurement (1) LCD display screen High accuracy (2) Indicator...
  • Page 8: Instructions

    Sounds and indicator Color instructions Display and Operating Instructions Indicator Range Sounds Operating Instructions Sound and Screen Display color Indicator color Displayed State Forehead temperature (Adult / Child) Measuring Ear temperature (Adult / Child) 22.0°C–37.5°C / 71.6°F–99.5°F A long beep Green 37.6°C–43°C / 99.6°F–109.4°F 3 short double beeps...
  • Page 9 Operating Instructions Operating Instructions Sound and Sound and Screen Display Screen Display Indicator color Indicator color Displayed State Displayed State Power button Press and release the for 1 second to power on the Mode thermometer. Then press the ut the cover on the probe, press See the table in the button .
  • Page 10 Operating Instructions Operating Instructions Sound and Sound and Screen Display Screen Display Indicator color Indicator color Displayed State Displayed State When the memory data is queried, when there is no memory data When the power is under the current serial number, the turned back on, a display is as shown.
  • Page 11 Operating Instructions Operating Instructions Sound and Sound and Screen Display Screen Display Indicator color Indicator color Displayed State Displayed State Error information & low battery The ambient temperature is higher A long beeps,the than 40.0°C (104.0°F) or lower than indicator light is 10.0°C (50.0°F).
  • Page 12 Measuring Ear Temperature When using the thermometer for the first time, move the battery cover and install the batteries. Do not force the thermometer into the ear canal. Otherwise, the ear canal may get injured. When taking the temperature on an adult, gently pull the ear up and back to make sure the ear canal is straight, so that the temperature 2.
  • Page 13 Measuring Forehead Temperature Measuring Object Temperature Put the cover on the probe of thermometer. Put the cover on the probe of t hermometer. Power button Power button Press the to power on the thermometer. Press the to power on the thermometer. Mode button Forehead Mode button...
  • Page 14: Batteries

    After a Measurement Replacing Batteries Slide the battery cover off along the marked direction and take it off. (1) After each measurement, you can enter the recall mode and query Insert the two AAA batteries into the compartment according to the earlier temperature readings.
  • Page 15 Connecting the Instrument to a Smartphone via Bluetooth Disinfection Download the Temperature Diary by MedM app or the MedM Health app Recommended disinfectants: from the App store or Google Play. Isopropyl alcohol solution (concentration: 70%) Install the app and register an account. Medicinal alcohol (concentration: 75%) Enable Bluetooth and the MedM app on your smart device.
  • Page 16 Specification Troubleshooting Product Name Infrared Thermometer Problem Possible Cause Solution Product Model JPD-FR409-BT Power Supply Mode Internal power supply Low battery Change the batteries. Operating Voltage DC 3V Battery Model Polarities of the Make sure that the batteries...
  • Page 17 EMC Information-Guidance and Manufacture’s Environment Humidity: 15%- 95% RH, non-condensing Atmospheric pressure: 86 - 106 kPa Declaration The infrared thermometer has been tested and conforms to the standard CAUTION: ASTM E1965-98. ASTM laboratory accuracy requirements in the display The Smart Ear & Forehead Thermometer needs special precautions range of 96.8°F to102.2°F (36°C-39°C) for ear canal IR thermometers is regarding EMC and needs to be installed and put into service according to 0.4°F ( 0.2°C).
  • Page 18: Class

    ± function. Therefore, its RF emissions are IEC 61000 -4-2 8 kV air synthetic material, the very low and are not likely to cause any relative humidity should be interference in nearby electronic at least 30 %. equipment. Power Power frequency magnetic The Smart Ear &...
  • Page 19 If the output power rating of the measured field strength in the location in which the JPD-FR409-BT is used exceeds transmitter in watts (W) the applicable RF compliance level above, the JPD-FR409-BT should be observed according to the transmitter to verify normal operation.
  • Page 20: Transportation

    Warranty and After - Sale Service between portable and mobile RF communications equipment (transmitters) and Smart Ear & Forehead Thermometer as recommended below, according to the The device is under warranty for one year from the date of purchase. The batteries, the packaging, and any damage caused by improper use are maximum output power of the communications equipment.

Table of Contents