Content Support Introduction Specification 4 - 5 Parts Warning and safety notices 7 - 8 How to use (°C displays for example) 9 - 10 Changing from Celsius to Fahrenheit Explanation of symbols on unit Product disposal Cleaning and Storage Replacing the batteries Quality warranty EMC Table...
Support Our manual should provide you with all the information you need to set up and use this product. For further assistance, why not contact our Customer Care team directly? We’re here to help! Our Customer Care team are available from 9am-5pm, Monday to Friday (excluding bank holidays). We promise to respond to all queries and will ensure to resolve any issue you may be having.
Introduction Please read the instructions carefully before using. The ear thermometer is a device capable of achieving infra-red temperature measurement when placed in the auditory canal of a user. It is a safe means of measuring human body temperature through the ear.
Specification The standard we adopted is EN12470-5:2003 Clinical thermometers -- Part 5: Performance of infrared ear thermometers (with maximum device). This device complies with the requirements of IEC 60601-1,IEC 60601-1-11 and IEC 60601-1-2. Range of displayed temperature : 34°C - 44°C (93.2°F - 111.2F) Display L°C (L°F) when the temperature is under 34°C (93.2°F) Display H°C (H°F) when the temperature is over 44°C (111.2°F) Operating ambient:...
Page 7
Specification temperature unit: centigrade or fahrenheit display of memory: last ten memories low voltage warning: the LCD display and then a beep sound is heard Power consumption: ≤30 mW Battery: two 1.5V alkalescence batteries (AAA) Battery life: 4000 uses Shelf life: 3 years Dimension: 146mm by 32mm by 37mm Net weight: Approx.
Warning and safety notices Cautions Before measuring, please condition the thermometer for at least 30 mins under operating conditions. New Disposable probe cover required for each measurement, otherwise inaccurate readings will occur. Please keep your ear canal clear, if not, the measuring result shall be inaccurate. The symbol ' ' on the LCD shows you can measure in the ear;...
Page 10
Warning and safety notices Note: • In order to avoid the spreading germs, please use a new disposable probe cover for each measurement. • Check to ensure probe cover is fitted on firmly before use (please see the diagrams below). If the probe cover is broken, discard and apply a new one immediately.
How to use (°C displays for example) Measuring human body temperature in the Ear Canal 1. Press the 'on/recall' button to turn on the thermometer, a beep sound is heard and the LCD displays as Fig A in one second. Fig A Fig B When the LCD displays as Fig B, now ready to measure.
How to use (°C displays for example) 5. Memory search: Turn the device on, then press the 'on/recall' button. The last ten memories (No.9-No.0) will take turns to display on the LCD as Fig C and Fig D. Fig C Fig D 6.
battery is empty Product disposal instructions for electronic devices The battery in this product complies with the requirements stated in European Directives 006/66/EEC Model Reference Memory Type BF equipment CE conformity marking L°C or L°F subject temperature is under 34°C H°C or H°F subject temperature is above 44°C Read IFU carefully...
Product disposal Please ensure environmental protection. Batteries do not belong in the domestic waste. Please hand them in at collection point or the municipal recycle material centre as special waste. This symbol on products and/or accompanying documents means that consumed electronic products must not be mixed with conventional domestic waste.
Replacing the batteries 1. When voltage of the batteries are low, the LCD will display ' ' symbol, please replace two new batteries in unit. The thermometer cannot work accurately under the condition of low voltage. 2. The thermometer is supplied with two 1.5V ALKALESCENCE BATTERIES (AAA). Insert new batteries when the low battery symbol appears on the LCD.
EMC Table This device is suitable for home healthcare environment and professional healthcare facility environment WARNING: Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are operating normally.
Page 17
EMC Table Guidance and manufacturer's declaration - electromagnetic emissions The device is suitable for use in the specified electromagnetic environment and it has meets the following immunity test levels. Higher immunity levels may cause the device's essential performance lost or degraded. Phenomenon Basic EMC standard Professional healthcare...
Page 18
EMC Table IEC 61000-4-4 Electric fast For input a.c. power port transients bursts d.c. power lines or signal input/output lines whose length exceeding 3m Surges IEC 61000-4-5 For 1.input a.c. power port; 2. all d.c. power ports connected permanently to cables >3m 3.
Page 19
EMC Table Recommended minimum separation distances Nowadays, many RF wireless equipments have being used in various healthcare locations where medical equipment and/or systems are used. When they are used in close proximity to medical equipment and/or systems, the medical equipment and/or systems' basic safety and essential performance may be affected. This device has been tested with the immunity test level in the below table and meet the related requirements of IEC 60601-1-2:2014.