WARNING consists of the following elements: • Information about the nature of a hazardous situation • Consequences of not avoiding a hazardous situation • Methods of avoiding a hazardous situation Tomosynthesis Quality Control Manual | Operator Manual MAMMOMAT Revelation | VC20 Print No. XPW7-340G.640.67.02.24...
Performance criteria/corrective action Test 4 Test Report Test 5: Artifact detection Objective Equipment required Procedure Performance criteria/corrective action Test 5 Test Report Changes in this version Index Tomosynthesis Quality Control Manual | Operator Manual MAMMOMAT Revelation | VC20 Print No. XPW7-340G.640.67.02.24...
If not otherwise specified, measurements are performed with the tube head in 0° position. All tests shall be performed in tomo-only mode, see Operator Manual Tomosynthesis, section Creating a 2D + 3D exposure. Tomosynthesis Quality Control Manual | Operator Manual MAMMOMAT Revelation | VC20 Print No. XPW7-340G.640.67.02.24...
Weekly tests do not need to be conducted prior to the first patient examination on the day of the test. Tomosynthesis Quality Control Manual | Operator Manual MAMMOMAT Revelation | VC20 Print No. XPW7-340G.640.67.02.24...
Voltage Tube load Average Glan‐ Maximum value ≤ Maximum value Tomo thickness dular Dose [kV] [mAs] [mGy] Passed Failed [mGy] Tomo [mm] ☐ ☐ ☐ ☐ Tomosynthesis Quality Control Manual | Operator Manual MAMMOMAT Revelation | VC20 Print No. XPW7-340G.640.67.02.24...
Page 8
In case your individual clinical setting differ from the manufacturer setting you can calculate your clinical glandular dose by multiply the measured dose with your individual dose level (in %) divided by 100. Tomosynthesis Quality Control Manual | Operator Manual MAMMOMAT Revelation | VC20 Print No. XPW7-340G.640.67.02.24...
3 Select the AEC exposure technique Advanced settings: OPCOMP: Off compress with 90 N minimum Auto Decompression: Off AEC Segmentation: Off Dose level: Normal 4 Select as RPG “Standard_Phantom” Tomosynthesis Quality Control Manual | Operator Manual MAMMOMAT Revelation | VC20 Print No. XPW7-340G.640.67.02.24...
Page 10
Measure X and Y using the distance line (Tools > Distance) and record the values in the test report ( Page 13 Test 2a. Test Report X and Y accuracy) (reconstructed). Tomosynthesis Quality Control Manual | Operator Manual MAMMOMAT Revelation | VC20 Print No. XPW7-340G.640.67.02.24...
Page 11
0). This indicates that the differentiation was not resolved visually (which in turn, is indicative of a good Z-resolution). Scroll up one slice and measure the following pixel values: Tomosynthesis Quality Control Manual | Operator Manual MAMMOMAT Revelation | VC20 Print No. XPW7-340G.640.67.02.24...
The requirements pertaining to the evaluation are described in FDA regulations and in subsequently issued guidance. The issue should be addressed and documented by a Siemens customer support engineer. Tomosynthesis Quality Control Manual | Operator Manual MAMMOMAT Revelation | VC20 Print No. XPW7-340G.640.67.02.24...
→ Values for slice z Enter the maximum value of Background each the 6 specks Speck 1 Speck 2 Speck 3 Speck 4 Tomosynthesis Quality Control Manual | Operator Manual MAMMOMAT Revelation | VC20 Print No. XPW7-340G.640.67.02.24...
Page 14
→ Values for slice z Enter the maximum value of Background each the 6 specks Speck 1 Speck 2 Speck 3 Speck 4 Speck 5 Tomosynthesis Quality Control Manual | Operator Manual MAMMOMAT Revelation | VC20 Print No. XPW7-340G.640.67.02.24...
5 Mount the 40 mm PMMA phantom on the collimator. 6 Mount the compression plate and lower it all the way. 7 Perform a Tomo scan. Tomosynthesis Quality Control Manual | Operator Manual MAMMOMAT Revelation | VC20 Print No. XPW7-340G.640.67.02.24...
Test 3 Test Report Tomo Compression Plate 25 cm x 36 cm Passed Failed Edges of collimator or compression plate ☐ ☐ not visible Tomosynthesis Quality Control Manual | Operator Manual MAMMOMAT Revelation | VC20 Print No. XPW7-340G.640.67.02.24...
3 Select the AEC exposure technique 28 kV AEC settings: OPCOMP: On Auto Decompression: Off AEC Segmentation: Off Dose level: Normal 4 Select as RPG “Standard_Phantom” Tomosynthesis Quality Control Manual | Operator Manual MAMMOMAT Revelation | VC20 Print No. XPW7-340G.640.67.02.24...
Page 19
• Record the result in the test report ( Page 21 Test 4 Test Report). • Record the displayed kV, mAs and AGD in the test report ( Page 21 Test 4 Test Report). Tomosynthesis Quality Control Manual | Operator Manual MAMMOMAT Revelation | VC20 Print No. XPW7-340G.640.67.02.24...
The requirements pertaining to the evaluation are described in FDA regulations and in subsequently issued guidance. The issue should be addressed and documented by a Siemens customer support engineer. Tomosynthesis Quality Control Manual | Operator Manual MAMMOMAT Revelation | VC20 Print No. XPW7-340G.640.67.02.24...
5 Mount the 40 mm PMMA phantom on the collimator. 6 Attach the compression plate and compress until the paddle touches the breast support. 7 Perform a Tomo scan. Tomosynthesis Quality Control Manual | Operator Manual MAMMOMAT Revelation | VC20 Print No. XPW7-340G.640.67.02.24...
The issue should be addressed and documented by a Siemens customer support engineer. Test 5 Test Report Anode/Filter Combination W/Rh Passed Failed No clinically relevant artifacts found ☐ ☐ Tomosynthesis Quality Control Manual | Operator Manual MAMMOMAT Revelation | VC20 Print No. XPW7-340G.640.67.02.24...
Chapter 1.1: Test 2 is now optional, because it is an algorithm test only and needs to be performed in case the Tomosynthesis reconstruction algorithm has changed. Chapter 3.1: Objective of optional test updated. Tomosynthesis Quality Control Manual | Operator Manual MAMMOMAT Revelation | VC20 Print No. XPW7-340G.640.67.02.24...