Appendix A: Technical Specifications - Hinge Health Enso User Manual

Table of Contents

Advertisement

Appendix A: Technical Specifications

Power
Power Source: Internally powered by fixed, rechargeable
battery
Battery Type: Lithium-polymer battery, 3.7V nominal
Battery Life: >80% of initial capacity after 500 cycles
(estimated two to three years of typical use)
Input Voltage: 5V DC
Input Current: 100mA (maximum)
Charging Source: AC adapter
AC Voltage: 100–240V AC
AC Frequency: 50–60Hz
Battery Compliance: IEC 62133-2:2017
Line Current Isolation: Treatment is not possible during
charging because the Enso device must be disconnected
from the gel pad while it is charging.
22
Output characteristics
Waveform: Biphasic, asymmetrical
Regulation: Current regulated
Current: 0–80mA (±10%, into 500Ω load)
Voltage: 90V (±10%, maximum, into open circuit)
Pulse Duration: 90us +/- 4%
Pulse Frequency: 2–160Hz (±4%, modulated according to
lookup table)
Operating Mode: Continuous
Phase Charge: 16.2µC (maximum)
Maximum Current Density: 0.51mA/cm2 into
500Ω load
Maximum Average Current: 2.33mA into 500Ω load
Maximum Average Power Density: 3.6mW/cm2 into 500Ω
load

Hide quick links: