Table of Contents

Advertisement

Available languages

Available languages

Quick Links

Version: 2023.01
All the rights to this manual are reserved, especially reproduction, publication and translation rights. No portion of this manual may be
reproduced by no means (photocopy, microfilm or other methods) without previous written authorization from ECLERIS. Neither can it
be reproduced, multiplied or published through any electronic data processing form.
The information contained in this manual may be modified or broaden without previous notification, without it representing any
responsibility to ECLERIS.
ECLERIS USA
11801 NW 100th Rd
Suite 17
Medley, FL 33178, U.S.A.
Phone: +1 786 483 7490
USER MANUAL
MICROSTAR SERIE OM-200
MICROSCOPE
INFORMATION ABOUT THIS MANUAL
Year: 2023
Language: English
CMC Medical Devices & Drugs S.L.
C/ Horacio Lengo Nº18, CP29006,
Málaga, Spain
Ph. +34951214054
Email: info@cmcmedicaldevices.com

Advertisement

Chapters

Table of Contents
loading

Summary of Contents for Ecleris MICROSTAR Series

  • Page 1 All the rights to this manual are reserved, especially reproduction, publication and translation rights. No portion of this manual may be reproduced by no means (photocopy, microfilm or other methods) without previous written authorization from ECLERIS. Neither can it be reproduced, multiplied or published through any electronic data processing form.
  • Page 2: Table Of Contents

    USER MANUAL / MANUAL DE USUARIO – MICROSTAR TABLE OF CONTENT SYMBOLS ..........................4 SAFETY INFORMATION ......................5 WARNINGS .......................... 5 DESCRIPTION......................... 6 INTENDED USE ........................6 INTENDED USER ........................ 6 DESCRIPTION AND COMPONENTS .................. 6 3.3.1 FLOOR STAND MICROSCOPE (OM-200F) ..............7 3.3.2 OPTICAL HEAD ......................
  • Page 3 USER MANUAL / MANUAL DE USUARIO – MICROSTAR 7.1.1 CLEANING THE EXTERNAL SURFACES ..............26 7.1.2 DISINFECTION ......................26 7.1.3 CLEANING OF OBJECTIVE LENS AND EYEPIECES ..........26 ELECTRICAL CONNECTIONS ..................27 FUSE CHANGE ........................27 TECHNICAL SPECIFICATIONS .................... 28 TROUBLESHOOTING ......................
  • Page 4: Symbols

    USER MANUAL / MANUAL DE USUARIO – MICROSTAR 1. SYMBOLS EU Authorized Warning Representative Caution CE Mark Read the related documentation This side up Alternate current Fuse Protective earth (Ground) Video output Equipotential bonding Freeze image function Manufacturer identification Video camera (Name, Address) Serial number White balance function...
  • Page 5: Safety Information

    2.1 WARNINGS It is important to read this manual carefully before using the Microstar Microscope. ECLERIS will not be held responsible for damage that can occur to the device due to not following the instructions or not observing the following criteria: ...
  • Page 6: Description

    USER MANUAL / MANUAL DE USUARIO – MICROSTAR  Do not open the device. Maintenance and repairs must be done by ECLERIS or an authorized company by the fabricant. ECLERIS will not be responsible for the consequences of repairs or maintenance carried out by unauthorized personnel.
  • Page 7: Floor Stand Microscope (Om-200F)

    USER MANUAL / MANUAL DE USUARIO – MICROSTAR Pantographic arm Code Description Binocular rotation length 23.03” (585mm) Straight 32,09” (815mm) OM-200F Floor-stand 45º Inclined 180° Inclinable 3.3.1 FLOOR STAND MICROSCOPE (OM-200F) Components: 1. Optical Head 6. Lock brake 2. HD beam Splitter 7.
  • Page 8: Optical Head

    USER MANUAL / MANUAL DE USUARIO – MICROSTAR 10. Progressive brake 12. Column 11. Progressive brake 13. Microscope Base 3.3.2 OPTICAL HEAD Magnification selector 5. Light on/off 2. Micro focus adjustment 6. Light intensity (increase) 3. Light filter selector 7. Light intensity (decrease) 4.
  • Page 9: Installation And Operation

    USER MANUAL / MANUAL DE USUARIO – MICROSTAR 4. INSTALLATION AND OPERATION 4.1 BASE INSTALLATION To assemble the microscope base, 4 wheels and 4 counterweights must be fitted. The wheels are assembled by screwing the bolt into the ends of the base as indicated in the following figure.
  • Page 10: Column Installation

    USER MANUAL / MANUAL DE USUARIO – MICROSTAR 4.2 COLUMN INSTALLATION Once the wheels and the base cover are placed, the column must be mounted on the base. Put the column in the central axis of the base as indicated in the following figure: Tighten the set screw with an Allen wrench.
  • Page 11: Forearm-Column Installation

    USER MANUAL / MANUAL DE USUARIO – MICROSTAR The column has to be mounted in the right position, as indicated in the following figure: Column Top Column Bottom 4.3 FOREARM-COLUMN INSTALLATION Place the forearm on top of the column, then tighten the screw in order to solidly attach the forearm to the column.
  • Page 12: Optical Head-Arm Installation

    USER MANUAL / MANUAL DE USUARIO – MICROSTAR 4.4 OPTICAL HEAD-ARM INSTALLATION The optical head is mounted to the arm through a joint elbow of 60º, as shown in the picture below. Embellishing screw (E) Tensor screw (D) Clamping screw (C) Fine washer (B) Security screw (F) Thick washer (A)
  • Page 13: Binocular-Splitter Installation

    USER MANUAL / MANUAL DE USUARIO – MICROSTAR Place the thick washer (A) on the 60º joint elbow. Then install the elbow on the microscope arm. Continue placing the fine washer (B) on the arm, as the previous picture. Above it place the clamping screw (C) and adjust it the maximum.
  • Page 14 USER MANUAL / MANUAL DE USUARIO – MICROSTAR In case of installing the binocular, the maneuver must be performed by releasing the Optical head adjustment screw, so that the binocular is inserted centrally in the upper part of the head. After the joint is centered and flat against the Optical head, press the adjustment screw so that the binocular remains firmly attached to the head.
  • Page 15: Video Cables Installation

    USER MANUAL / MANUAL DE USUARIO – MICROSTAR 4.6 VIDEO CABLES INSTALLATION The splitter and the Optical head with integrated light have specific power and video connectors for normal operation. Once the splitter is mounted to the optical head, proceed to connect the video and power cables.
  • Page 16: Operation - Functions

    USER MANUAL / MANUAL DE USUARIO – MICROSTAR 5. OPERATION - FUNCTIONS 5.1 MICROSCOPE OPERATION This section describes the equipment functioning procedure. The device shall only be operated by personnel trained in the use of this type of microscopes. 1) Connect the power supply cable to the microscope and then to the mains. 2) Press the power switch on the forearm to turn the unit on (the green indicator light will turn on).
  • Page 17: Connections

    USER MANUAL / MANUAL DE USUARIO – MICROSTAR 5.2 CONNECTIONS A display monitor can be connected to the HDMI output located inside the Microstar Microscope forearm. The forearm cover must be removed to access the connectors. 5.2.1 FOREARM 1. 100 VA auxiliary power output (monitor connection)
  • Page 18: Light Source

    USER MANUAL / MANUAL DE USUARIO – MICROSTAR 5.3 LIGHT SOURCE The Ecleris Microstar OM-200 Microscope has a light module integrated to the microscope’s Optical head. From this module you can turn the light on or off, vary its intensity, select different diaphragm levels to reduce the illuminated field and select different light filters.
  • Page 19: Dioptric Correction Adjustment

    USER MANUAL / MANUAL DE USUARIO – MICROSTAR 5.5 DIOPTRIC CORRECTION ADJUSTMENT The Ecleris Microstar OM-200 Microscope has independent dioptric correction for both eyepieces. The adjustment is made by turning the eyepieces towards the + sign or towards the – sign.
  • Page 20: Forearm Movement

    USER MANUAL / MANUAL DE USUARIO – MICROSTAR 6.1 FOREARM MOVEMENT The forearm has a rotation system related to the column axis that allow its rotation to reach an optimum work position. The forearm can rotate 320° related to the column axis, and it has stops to avoid the accumulation of turns and winding of the cables.
  • Page 21: Rotational Movement Of The Optical Head

    USER MANUAL / MANUAL DE USUARIO – MICROSTAR 6.3 ROTATIONAL MOVEMENT OF THE OPTICAL HEAD The microscope has an elbow that positions the optical head at a fixed angle in relation to the arm. Movement can be regulated or totally blocked using the adjustment control command. If this command is completely loosened, the Optical head can be rotate freely on the end of the arm.
  • Page 22: Optical Head Rotation Axis

    USER MANUAL / MANUAL DE USUARIO – MICROSTAR 6.4 OPTICAL HEAD ROTATION AXIS For a better alignment of the optical axis in relation with the target, the Optical head can be rotated on the fork. The rotational range is 0-360 degrees. The movement can be adjusted or completely locked using the control knob.
  • Page 23: Microscope Shipment

    USER MANUAL / MANUAL DE USUARIO – MICROSTAR 6.6 MICROSCOPE SHIPMENT The Ecleris Microstar OM-200 Microscope has been designed to be moved over short distances within a doctor’s office or room. If it necessary to move the microscope over larger distances, special attention should be paid to the following cautions: ...
  • Page 24 USER MANUAL / MANUAL DE USUARIO – MICROSTAR 3. Mount the monitor holder on the back of the monitor and place the 4 Allen screws. 4. Place the column support to the end of the monitor holder. 5. Fit the Allen screw to secure the joint. Page 24 of 64 INMANUUML019 USER MANUAL - M.
  • Page 25 USER MANUAL / MANUAL DE USUARIO – MICROSTAR Page 25 of 64 INMANUUML019 USER MANUAL - M. USO OM-200 V.2023.01 ENG-ESP US...
  • Page 26: Maintenance

    USER MANUAL / MANUAL DE USUARIO – MICROSTAR 7. MAINTENANCE To maintain optimum operation of the Ecleris Microstar OM-200 Microscope, it is recommended to perform a revision of the equipment once a year. This procedure should include:  Checking turns and movements in order to evaluate the lubrication of the pieces.
  • Page 27: Electrical Connections

    USER MANUAL / MANUAL DE USUARIO – MICROSTAR 7.2 ELECTRICAL CONNECTIONS Check the condition of all components and their aging process. Search for:  Faulty, broken or cut cables.  Faulty connection between the cable and the plug.  Oxide corrosion in connections pins. 7.3 FUSE CHANGE Disconnect the power cord from the electrical network.
  • Page 28: Technical Specifications

    USER MANUAL / MANUAL DE USUARIO – MICROSTAR 4) Remove the burnt fuse and replace it with the extra fuse that is located inside the same plastic holder. Use fuses with the same electrical and safety characteristics as the original fuse, as indicated on the equipment label.
  • Page 29 USER MANUAL / MANUAL DE USUARIO – MICROSTAR Illuminated field / Diaphragm control Focal Focus lenght lens [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] POWER SUPPLY 100-240 VAC, 50 / 60 Hz FUSES 2 x T2AL250V CLASS APPLICABLE PART Does not have applicable parts DEGREE OF PROTECTION AGAINST IPX0...
  • Page 30 USER MANUAL / MANUAL DE USUARIO – MICROSTAR STORAGE AND TRANSPORTATION CONDITIONS -30 – 50 ºC TEMPERATURE 10 – 95 % relative humidity HUMIDITY 500 – 1060 hPa ATMOSPHERIC PRESSURE Page 30 of 64 INMANUUML019 USER MANUAL - M. USO OM-200 V.2023.01 ENG-ESP US...
  • Page 31: Troubleshooting

    If the problem cannot be solved, please contact your local distributor or the closet ECLERIS office and request technical service. Page 31 of 64 INMANUUML019 USER MANUAL - M. USO OM-200 V.2023.01 ENG-ESP US...
  • Page 32: Technical Service

    Problem´s description 4) Send the Microscope for repair or replacement. The ECLERIS Microstar OM-200 Microscope shall be sent in its original Packaging so that it is not damaged during transportation. ECLERIS shall not be liable for damages that may occur during transportation, caused by deficient packaging.
  • Page 33 La información contenida en este manual puede ser modificada o ampliada sin notificación previa, sin que ello represente responsabilidad alguna para ECLERIS. ECLERIS USA CMC Medical Devices &...
  • Page 34 MANUAL DE USUARIO / USER MANUAL – MICROSTAR TABLA DE CONTENIDO SIMBOLOGIA ........................36 INFORMACION DE SEGURIDAD ..................37 ADVERTENCIAS ........................ 37 DESCRIPCION ........................38 USO PREVISTO ......................... 38 USUARIO PREVISTO ......................38 DESCRIPCION Y COMPONENTES ................... 38 3.3.1 MICROSCOPIO CON ESTATIVO DE PIE (OM-200F) ..........39 3.3.2 CABEZAL ........................
  • Page 35 MANUAL DE USUARIO / USER MANUAL – MICROSTAR 7.1.1 LIMPIEZA DE SUPERFICIES EXTERIORES .............. 58 7.1.2 DESINFECCIÓN ......................58 7.1.3 LIMPIEZA DE OBJETIVOS Y OCULARES ..............58 CONEXIONES ELÉCTRICAS .................... 58 CAMBIO DE FUSIBLE ......................59 ESPECIFICACIONES TÉCNICAS ..................60 SOLUCIÓN DE PROBLEMAS ....................
  • Page 36: Simbologia

    MANUAL DE USUARIO / USER MANUAL – MICROSTAR 1. SIMBOLOGIA Representante Advertencia Autorizado en la Comunidad Europea Precaución Marcado CE Es obligatorio consultar los Arriba documentos acompañantes Tensión alterna Fusible Puesta a Tierra de Protección Salida de Video Tierra de Equipotencialidad Congelar Imagen Identificación del fabricante Cámara de Video...
  • Page 37: Informacion De Seguridad

    Lea cuidadosamente todo el manual antes de utilizar el dispositivo. Ante cualquier duda sobre el contenido de este manual, la utilización o seguridad del dispositivo, solicitamos comunicarse con su distribuidor local o con la oficina de ECLERIS más cercana. 2.1 ADVERTENCIAS Es importante leer cuidadosamente este manual antes de utilizar el Microscopio Microstar.
  • Page 38: Descripcion

     No abra el equipo. El mantenimiento y las reparaciones deben ser realizados por ECLERIS o una empresa autorizada por ésta. ECLERIS no se responsabiliza por las consecuencias de reparaciones o mantenimientos efectuados por personas no autorizadas.
  • Page 39: Microscopio Con Estativo De Pie (Om-200F)

    MANUAL DE USUARIO / USER MANUAL – MICROSTAR Lea cuidadosamente todo el manual antes de utilizar el microscopio. Los componentes accesorios del microscopio OM-200 son los siguientes:  Soporte para monitor  Sistema de video  Cable especial que contiene 1 ficha HDMI macho para conectar al sistema de video, y 2 fichas de alimentación DC-PLUG para alimentar el sistema de video y la luz integrada.
  • Page 40: Cabezal

    MANUAL DE USUARIO / USER MANUAL – MICROSTAR 3.3.2 CABEZAL 1. Selector de Magnificador 5. Encendido/Apagado Luz 2. Ajuste micro enfoque 6. Intensidad de Luz (incrementar) 3. Selector de Filtro de Luz 7. Intensidad de Luz (decrementar) 4. Selector de Diafragma Página 40 de 64 INMANUUML019 USER MANUAL - M.
  • Page 41: Instalacion Y Puesta En Marcha

    MANUAL DE USUARIO / USER MANUAL – MICROSTAR 4. INSTALACION Y PUESTA EN MARCHA 4.1 INSTALACION DE LA BASE Para ensamblar la base del microscopio se deben colocar 4 ruedas y 4 contrapesos. Las ruedas se montan enroscando el perno de la misma en los extremos de la base como indica la figura siguiente.
  • Page 42: Instalacion De La Columna

    MANUAL DE USUARIO / USER MANUAL – MICROSTAR Una vez colocados los pesos, se debe montar el cobertor plástico sobre la base de la columna como indica la siguiente figura: 4.2 INSTALACION DE LA COLUMNA La columna se debe instalar sobre la base, una vez que están las ruedas y el cobertor colocados. Para instalar la columna se debe colocar la misma sobre el eje central de la base como indica la siguiente figura: Ajustar el tornillo de fijación con una llave Allen.
  • Page 43: Instalacion Del Antebrazo A La Columna

    MANUAL DE USUARIO / USER MANUAL – MICROSTAR La columna tiene que ser montada en la posición correcta, respetando la parte superior e inferior de la columna, como se indica en la siguiente figura: Columna Parte Superior Columna Parte Inferior 4.3 INSTALACION DEL ANTEBRAZO A LA COLUMNA Colocar el antebrazo en la parte superior de la columna, y luego fijar el tornillo para que el antebrazo quede fijado sólidamente a la columna.
  • Page 44: Instalacion Del Cabezal A Brazo

    MANUAL DE USUARIO / USER MANUAL – MICROSTAR 4.4 INSTALACION DEL CABEZAL A BRAZO El cabezal está montado al brazo mediante un codo de articulación de 60º como se indica en la imagen a continuación Página 44 de 64 INMANUUML019 USER MANUAL - M. USO OM-200 V.2023.01 ENG-ESP US...
  • Page 45: Instalacion Del Binocular / Sistema De Video

    MANUAL DE USUARIO / USER MANUAL – MICROSTAR Coloque la arandela gruesa (A) en el perno del angulo de 60º. Luego coloque el angulo de 60º al brazo del microscopio. Continúe colocando la arandela fina (B) en el brazo, como se ve en la imagen anterior.
  • Page 46 MANUAL DE USUARIO / USER MANUAL – MICROSTAR En caso de la instalación del binocular, la maniobra se debe realizar liberando el tornillo de ajuste del cabezal, de modo que el binocular se introduzca de forma centrada en la parte superior del cabezal.
  • Page 47: Instalacion Cables De Video

    MANUAL DE USUARIO / USER MANUAL – MICROSTAR 4.6 INSTALACION CABLES DE VIDEO El sistema de video y el cabezal con luz integrada, disponen de conectores de alimentación y video específicos para su normal funcionamiento. Una vez montado el sistema de video al cabezal, proceda a instalar los cables de video y alimentación.
  • Page 48: Operación - Funciones

    MANUAL DE USUARIO / USER MANUAL – MICROSTAR 5. OPERACIÓN – FUNCIONES 5.1 OPERACIÓN DEL MICROSCOPIO En esta sección se detalla el funcionamiento del equipo. Este equipo debe ser operado sólo por personal entrenado en la utilización de este tipo de microscopios.
  • Page 49: Conexiones

    MANUAL DE USUARIO / USER MANUAL – MICROSTAR 5) Apague el interruptor de encendido (posición hacia afuera). 5.2 CONEXIONES Un monitor de visualización puede conectarse a la salida HDMI que se aloja en el antebrazo del Microscopio Microstar. Para acceder al conector debe retirarse la tapa del antebrazo. 5.2.1 ANTEBRAZO 1.
  • Page 50: Fuente De Iluminacion

    MANUAL DE USUARIO / USER MANUAL – MICROSTAR 5.3 FUENTE DE ILUMINACION El Microscopio Ecleris Microstar OM-200 dispone de un módulo de luz integrada al cabezal del microscopio. Desde este módulo se puede apagar o encender la luz, variar la intensidad de la misma, seleccionar diferentes niveles de diafragma para disminuir el campo iluminado y seleccionar diferentes filtros de luz.
  • Page 51: Ajuste De La Correccion Dioptrica

    55 y 75 mm. 5.5 AJUSTE DE LA CORRECCION DIOPTRICA El Microscopio Ecleris Microstar OM-200 posee una corrección dióptrica independiente para ambos oculares. El ajuste se realiza girando los oculares hacia el signo + o el signo –.
  • Page 52: Movimiento Del Antebrazo

    MANUAL DE USUARIO / USER MANUAL – MICROSTAR 6.1 MOVIMIENTO DEL ANTEBRAZO El antebrazo tiene un sistema de rotación sobre el eje de la columna que permite rotar la posición para ajustar una posición de trabajo óptima. El antebrazo puede rotar 320º respecto al eje de la columna, y dispone de topes para evitar acumular giros y enroscar los cables.
  • Page 53: Movimiento Rotacional Del Cabezal

    MANUAL DE USUARIO / USER MANUAL – MICROSTAR 6.3 MOVIMIENTO ROTACIONAL DEL CABEZAL Este movimiento permite la ubicación del microscopio sobre cualquier lado del paciente sin necesidad de moverlo. El microscopio posee un codo que posiciona al cabezal a un ángulo fijo de 60°...
  • Page 54: Movimiento De Giro Del Cabezal

    MANUAL DE USUARIO / USER MANUAL – MICROSTAR 6.4 MOVIMIENTO DE GIRO DEL CABEZAL El cabezal puede girar sobre el eje de la horquilla, permitiendo un mejor posicionamiento de la dirección del centro óptico, en relación al objetivo observado. El rango de movimiento permitido es de 0-360º.
  • Page 55: Transporte Del Microscopio

    MANUAL DE USUARIO / USER MANUAL – MICROSTAR 6.6 TRANSPORTE DEL MICROSCOPIO El Microscopio Ecleris Microstar OM-200 ha sido diseñado para ser trasladado cortas distancias dentro del recinto médico. Si es necesario transportarlo grandes distancias, debe prestarse especial atención a las siguientes precauciones: ...
  • Page 56 MANUAL DE USUARIO / USER MANUAL – MICROSTAR 3. Monte en la parte trasera del monitor el soporte del monitor y coloque los 4 tornillos Allen. 4. Inserte la columna soporte al extremo del soporte del monitor. 5. Asegure la unión colocando el tornillo Allen. Página 56 de 64 INMANUUML019 USER MANUAL - M.
  • Page 57 MANUAL DE USUARIO / USER MANUAL – MICROSTAR Página 57 de 64 INMANUUML019 USER MANUAL - M. USO OM-200 V.2023.01 ENG-ESP US...
  • Page 58: Mantenimiento

    MANUAL DE USUARIO / USER MANUAL – MICROSTAR 7. MANTENIMIENTO Para conservar el funcionamiento óptimo del Microscopio Ecleris Microstar OM-200 se recomienda realizar una revisión anual del equipo. Este procedimiento deberá incluir:  Verificar giros y movimientos para evaluar el nivel de lubricación de los mismos.
  • Page 59: Cambio De Fusible

    MANUAL DE USUARIO / USER MANUAL – MICROSTAR Verifique que no existan:  Cables en mal estado, rotos o cortados.  Mala conexión del cable con el enchufe.  Pines de conexión oxidados. 7.3 CAMBIO DE FUSIBLE Desconecte el cable de alimentación de la red eléctrica. 1) Desconecte el cable de alimentación de la red eléctrica.
  • Page 60: Especificaciones Técnicas

    MANUAL DE USUARIO / USER MANUAL – MICROSTAR 4) Retire el fusible dañado y reemplácelo por el fusible de repuesto que se encuentra en el mismo zócalo. Utilice fusibles de idénticas características eléctricas y de seguridad que el fusible original, según se indica en el rotulado del equipo. 5) Introduzca el zócalo hasta escuchar un “click”.
  • Page 61 MANUAL DE USUARIO / USER MANUAL – MICROSTAR Campo Iluminado / Nivel de Diafragma Distancia Lente Focal Focal [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] SUMINISTRO ELÉCTRICO 100-240 VAC, 50 / 60 Hz FUSIBLES 2 x T2AL250V PROTECCIÓN CONTRA CHOQUE Clase I ELECTRICO PARTE APLICABLE...
  • Page 62 MANUAL DE USUARIO / USER MANUAL – MICROSTAR CONDICIONES DE SERVICIO 10 – 40 ºC TEMPERATURA 35 – 75 % humedad relativa HUMEDAD 700 – 1060 hPa PRESION ATMOSFERICA CONDICIONES DE ALMACENAMIENTO Y TRANSPORTE -30 – 50 ºC TEMPERATURA 10 – 95 % humedad relativa HUMEDAD 500 –...
  • Page 63: Solución De Problemas

    Si no puede solucionar el inconveniente, contacte a su distribuidor local u oficina de ECLERIS más cercana y solicite servicio técnico SERVICIO TÉCNICO El Microscopio Ecleris Microstar OM-200 debe ser reparado y/o mantenido exclusivamente en instalaciones autorizadas por ECLERIS.
  • Page 64: Garantia

    Se deberá efectuar el envío con el embalaje original para evitar que se dañe el equipo durante el traslado. ECLERIS no se responsabiliza por aquellos daños o defectos originados durante el transporte que sean vinculados a un embalaje deficiente. GARANTIA ECLERIS declara que el Microscopio Ecleris Microstar OM-200 ha pasado su Control de Calidad y no evidenció...