Sound Intensity Calibrator Type 4297 – User Manual
COMPLIANCE WITH STANDARDS
The CE marking is the manufacturer's declaration that the product meets the requirements of the
applicable EU directives
RCM mark indicates compliance with applicable ACMA technical standards – that is, for
telecommunications, radio communications, EMC and EME
China RoHS mark indicates compliance with administrative measures on the control of pollution
caused by electronic information products according to the Ministry of Information Industries of the
People's Republic of China
WEEE mark indicates compliance with the EU WEEE Directive
Safety
EN/IEC 61010–1: Safety requirements for electrical equipment for measurement, control and
laboratory use
ANSI/UL 61010–1: Safety requirements for electrical equipment for measurement, control and
laboratory use
EMC Emission
EN/IEC 61000–6–3: Generic emission standard for residential, commercial and light industrial
environments
EN/IEC 61000–6–4: Generic emission standard for industrial environments
CISPR 22: Radio disturbance characteristics of information technology equipment. Class B Limits
FCC Rules, Part 15: Complies with the limits for a Class B digital device
This ISM device complies with Canadian ICES–001 (standard for interference-causing equipment)
EMC Immunity
EN/IEC 61000–6–1: Generic standards – Immunity for residential, commercial and light industrial
environments
EN/IEC 61000–6–2: Generic standards – Immunity for industrial environments
EN/IEC 61326: Electrical equipment for measurement, control and laboratory use – EMC
requirements
EN/IEC 60942: Sound Calibrators – Amendment 1
Temperature
IEC 60068–2–1 & IEC 60068–2–2: Environmental Testing. Cold and Dry Heat.
Operating Temperature: – 10 to + 50 °C (14 to 122 °F)
Storage Temperature: –25 to +70 °C (–13 to 158 °F)
IEC 60068–2–14: Change of Temperature: – 10 to + 50 °C (2 cycles, 1 ° C/min.)
IEC 60068–2–3: Damp Heat: 90% RH (non-condensing at 40 °C (104 °F))
Humidity
Mechanical
Non-operating:
IEC 60068–2–6: Vibration: 0.3 mm, 20 m/s
IEC 60068–2–27: Shock: 1000 m/s
IEC 60068–2–29: Bump: 1000 bumps at 250 m/s