Bowie & Dick Program:
This program is for assessing steam penetration level. It is
strongly recommended to run this program every day before
using the device and check the integrity of the device's perfor-
mance by studying the related indicator.
If the indicator color change is not proper, you may not use the
device. You should immediately contact the after sale services
department of the company under this circumstance.
If you wish to test the steam penetration indicator with
other cycles, please remember that this test is only for cy-
cles that are designed for porous or multi-layered wrapped
items which include Universal, Prion and Gentle programs.
You may only include unwrapped or thin single layered
wrapped items in Quick B and Quick S programs. Such
loads do not require a high level of steam penetration.
Therefore do not include the steam penetration indicator
for these programs.
Vacuum Test:
You can perform a vacuum test by selecting it from the menu.
The result will be shown at the end on the screen and a report
will be printed. If Vacuum OK! is shown, the vacuum creation
inside the chamber works properly.
Autoclave
Elite
Premium B18
2000.02.00 04:05:00
020 2
074.8
0.83 b
TT:00:10
ST:00:15
Vaccum Test . . . .
Vac.Test
STRT/STOP
SET
Picture 11
Vacuum Sterilizer