In order to permanently connect the device to the distilled
water tank, put the device in a place close to the tank. Connect
the distilled water tank with a hose to the clean water input
on the back of the device and then push the down button.
The pump will work for 3 minutes and in case distilled water is
required to remove the FE alarm, push the down button again.
Manual mode:
In case you do not wish to connect the device permanently to
the distilled water and drainage tanks, you can use a flask and
connection hose for manual water loading and draining. Please
note that you will need to use the water input on the front of
the device.
For loading the device, fill the flask with distilled water. After
connecting the flask hose to the clean water input on the front
of the device, place the flask on the top of the device. For
draining dirty water after connecting the empty flask hose to
the drainage on the front of the device, place the flask on a
surface lower than the device.
The minimum water required for a full cycle is 600 cc.
The quality of the distilled water used should be according
to table 3.