8.3 Log page
8.3.1 Making the connection
Make the connection to the device with the following steps
•
Connect the device to the PC with a USB cable (with Standard-A and Mini-B plugs), or with HART
•
Select the port (COM1..COM256) with Port selection box
•
COM ports that are currently present in the system are shown with character "*", e.g. "COM1 *"
Select USB or HART with Connection selection
•
•
Note: HART communication is about 100 - 130 times slower than USB
Connect with Connect button
•
If the device is found
•
The log settings are automatically read from the device
•
Various functions on the page are enabled
•
Device type and firmware versions are shown on the Device line
• Connect button is replaced with Disconnect button.
If the device is not found, an error message is displayed.
Satron Instruments Inc., FI-33960 Pirkkala, Finland
Tel. +358 207 464 800, www.satron.com