Table of Contents INSTALLATION......................3 Mechanical installation..................3 Electrical connections..................15 SETTING UP......................17 Setting up with Satron VoAdvisor Service Software........17 Setting up with local switches..............17 Setting up with VO remote unit..............18 USER GUIDE FOR MENUS..................19 Measurement values menu................19 Menu structure....................20 DISPLAY SETUP......................20...
Do not install the sensor on top or on the bottom of the pipe. There can be air or dirt infront of the measuring lens and give false results. Minimum distance from lens to pipe wall is 50 mm. Figure 1‑2 Basic mounting dimensions Satron Instruments Inc., FI-33960 Pirkkala, Finland Tel. +358 207 464 800, www.satron.com...
Page 4
• Lock the sensor in position by screwing the hex nut. Apply 60±20 Nm torque to tight to eliminate the effect of vibration and other such factors. Do not use sealing tape etc. on threaded connection! Satron Instruments Inc., FI-33960 Pirkkala, Finland Tel. +358 207 464 800, www.satron.com...
Page 5
46 ‑ 04 FOR EHEDG CERTIFIED INSTALLATIONS. FOR INSTALLATIONS ACCORDING 3A Only the Satron VO Turbidity transmitter, models VOM All couplings that are according the 3-A appliance and VOA with G1 connection and EPDM seal are EHEDG listed in the bottom of this page.
Page 6
L‑Housing data cable 15m PVC 70000451 Extension cable +10m PVC 70000460 FUSE for L‑Housing 74212000 Suunn. Design Piirt. Drawn Seal for L‑Housing display 80017226 Tark. Insp. Hyväks. Approved Satron Instruments Inc., FI-33960 Pirkkala, Finland Tel. +358 207 464 800, www.satron.com...
Page 7
132 (+ 50 for cable) 132 (+ 50 for cable) STD. 227 (or specified) VOM/VOA with remote measuring probe and PVC M12 cable (NRT43) STD. 227 (or specified) Satron Instruments Inc., FI-33960 Pirkkala, Finland Tel. +358 207 464 800, www.satron.com...
Page 8
Piirustus n:o Drawing no Massa Weight Suunn. Suhde Design Scale Piirt. 1:10 Drawn Tark. Insp. Hyväks. Nimi Ti Approved Massa Weight PG9 std. RDU ‑ Remote Display Unit (L) T1325016 Satron Instruments Inc., FI-33960 Pirkkala, Finland Tel. +358 207 464 800, www.satron.com...
Page 9
Design Scale Tark. Insp. Piirt. Drawn Hyväks. Nimi Title Entinen Former Approved Tark. Insp. Piirustus n:o Drawing n Hyväks. Nimi Title Entinen Massa Approved Weight Piirustu Massa Weight Satron Instruments Inc., FI-33960 Pirkkala, Finland Tel. +358 207 464 800, www.satron.com...
Page 10
PVC 10meter M12 std. Suunn. Suhde Tuote Design Scale VOD with remote electronics housing with display (L housing) Piirt. 06.04.2017 Drawn Tark. Insp. Hyväks. Nimi Title Approved Satron Instruments Inc., FI-33960 Pirkkala, Finland VODNS221L0V407TN Massa Weight Tel. +358 207 464 800, www.satron.com...
Page 11
For easy installation the VOD is available with many different process connections. Such as: DN50 PN40 DN100 PN40 Tri‑Clamp DN38, DN51, DN63,5 Sandvik DN70 PN40 SMS38 SMS51 Tuchenhagen / Varivent Contact Satron for other possibilities. Satron Instruments Inc., FI-33960 Pirkkala, Finland Tel. +358 207 464 800, www.satron.com...
Page 12
VOF with remote electronics housing with display (L housing) Suunn. Suhde Tuote Product Liittyy Belon Design Scale Piirt. Drawn Tark. Insp. Hyväks. Nimi Title Entinen Former Approved Piirustus n:o Drawing no Massa Weight Satron Instruments Inc., FI-33960 Pirkkala, Finland Tel. +358 207 464 800, www.satron.com...
Page 13
Piirt. Tark. 240517 Drawn Insp. Tark. Insp. Hyväks. Approved Hyväks. Nimi Title Approved VOFNJ221NRT427B1 Massa Weight VOF B1 with remote electronics NR housing Satron Instruments Inc., FI-33960 Pirkkala, Finland Tuote Product Liittyy Belongs to Tel. +358 207 464 800, www.satron.com...
Page 14
Liittyy Belongs Design Scale Piirt. Drawn Tark. Insp. Hyväks. Nimi Title Entinen Former Suunn. Approved Design Piirt. Piirustus n:o Drawing no Drawn Massa Tark. Weight Insp. Hyväks. Approved Satron Instruments Inc., FI-33960 Pirkkala, Finland Tel. +358 207 464 800, www.satron.com...
Figure 1‑18a, housing with M12‑connector, code HT Figure 1‑18c Figure 1‑18b, housing with PLUG DIN43650‑connector, Housing with PLUG DIN43650‑ and M12‑connector, test code HS connector box, code HT & HS Satron Instruments Inc., FI-33960 Pirkkala, Finland Tel. +358 207 464 800, www.satron.com...
Page 16
Installation and Setting-Up Instructions Manual Wiring Wiring Housing with display and PLUG DIN43650‑connector, code NS Housing with display and M12‑connector, code NT Wiring Remote electronics housing with display, code L Satron Instruments Inc., FI-33960 Pirkkala, Finland Tel. +358 207 464 800, www.satron.com...
2.1 Setting‑up with Satron VoAdvisor Service Software When you want to have all the operations of the Smart transmitter, we recommend the use of Satron VoAd‑ visor Service Software program. Satron Instruments Inc. will deliver you the program and HART‑modem.
2.3 Setting up with VO remote unit The Satron VO remote can be provided with a wall box which is capable of having a 20m cable between the Sensing unit and the Display unit. Inside the Display unit is a terminal where up to 3 binary inputs, 3 relay outputs and 2 analog milliamp loops can be connected.
3. USER GUIDE FOR MENUS The user interface for the series VO sensors, housing option N, consists of display and operating keys. Among other things, the user interface allows you to set process variables in the desired units on the display and to configure the sensor e.g.
3.2 Menu Structure The VO sensors are equipped with a display and with Firmware version O120601A or later, the menu is split in to a “BASIC”, and “FULL” structure. You can see the current firmware version during startup for 1 second.
Configure parameters that have an effect on the range to master. measurement. In menu 3 (PVS/CURR): Transmitter sends all pro‑ cess variables and current signal’s value. Satron Instruments Inc., FI-33960 Pirkkala, Finland Tel. +358 207 464 800, www.satron.com...
Page 22
RECIPE: UNIT: Up to 4 different recipes can be stored in the VO. Select transmitter’s unit you want to show in the The basic factory calibration is stored standard display as active recipe’s Process Value.
Page 23
Default 0°C. Set the new value and accept it with INPUT and OUTPUT relays. ENTER or press ESC if you do not want to change Satron highly recommends the use of the software the value. package VoAdvisor to alter these settings!
Page 24
TER after that will give default value 12 mA, which FUVALUES: can be changed as desired with the UP/DOWN keys. FU is short for Satron Factory Unit and can be used The last ENTER will switch the transmitter back to for diagnosing the primary measurement’s status.
Page 25
SATRON VO Installation and Setting-Up Instructions Manual Satron Instruments Inc., FI-33960 Pirkkala, Finland Tel. +358 207 464 800, www.satron.com...
1. The different recipes can be renamed and also be enabled with the binary inputs. To change a calibration Satron recommends to use the “VoAdvisor” software. It is also possible to change the calibration with the use of the sensors own user interface.
Current output 2 is galvanically isolated Maximum supply voltage 35 VDC Range 3.5...23 mA Factory setting 4...20 mA Maximum isolation voltage 100 VDC Analog output signal according to NAMUR NE 043: Satron Instruments Inc., FI-33960 Pirkkala, Finland Tel. +358 207 464 800, www.satron.com...
Hygienic installation VOM G1 with welding Not hygienic installations can be equipped with the adapter in steel pipes/tanks. FLUSHING COUPLING, and optional flushing cabinet. Item number: M1050021 Item number: M548101 Item number: M1050194‑RT Satron Instruments Inc., FI-33960 Pirkkala, Finland Tel. +358 207 464 800, www.satron.com...
• The depth of the dead leg shall be kept as shallow as possible; ≤ d1, see the EHEDG position paper. • Minimum pipe diameter: 38mm (1.5”) • Process connection direction: Vertical: upwards flow Horizontal: on the side • Cable entry direction: from below Satron Instruments Inc., FI-33960 Pirkkala, Finland Tel. +358 207 464 800, www.satron.com...
This coupling is intended to be installed on the outlet of a pipe, Clamped on the pipe wall, the sensor will be facing the flow. Item code: M1050186‑FLUSH Satron Instruments Inc., FI-33960 Pirkkala, Finland Tel. +358 207 464 800, www.satron.com...
Page 33
For drains, floor drains, wells, pits with submersible IP68 measuring head. The optional flushing coupling can be easily welded to a thread bar and held in place. Coupling order code: M1050021 Satron Instruments Inc., FI-33960 Pirkkala, Finland Tel. +358 207 464 800, www.satron.com...
Page 34
The standard VOMxxxxxNRT4xxG1 is installed to the pipe assembly. The pipe assembly is held in place by a clamp and mounted to a wall or bracket. Pipe assembly order code: contact Satron Satron Instruments Inc., FI-33960 Pirkkala, Finland Tel. +358 207 464 800, www.satron.com...
Page 35
Installation example VOMxxxxNRT2xxH1 For small drains, floor drains and open tube collectors. The sensor is dropped in the drain. No flushing possible, so its important that lens is kept clean. Satron Instruments Inc., FI-33960 Pirkkala, Finland Tel. +358 207 464 800, www.satron.com...
If the automation system has its own load resistor, make sure to remove the already installed 250ohm load resistor from the flushing unit (if present between pins 7 and 10). Resistor is not included in every model version. Satron Instruments Inc., FI-33960 Pirkkala, Finland Tel. +358 207 464 800, www.satron.com...
2. Start the VoAdvisor software. Check in the ‘Start’ window that the VoAdvisor software version is 115 or later. In older versions the flushing configuration is not supported. 3. Click ‘Start communication!’ -> ‘Configuration’ -> ‘Input / Output settings’. Satron Instruments Inc., FI-33960 Pirkkala, Finland Tel. +358 207 464 800, www.satron.com...
Page 39
Here the timings can be set for how long the flushing should take place, (standard 3 seconds). And the interval between the flushing cycles (Standard 1 minute). Output State when flushing: This will allow freezing the output signal of the optical transmitter while flushing. (Standard ON) Satron Instruments Inc., FI-33960 Pirkkala, Finland Tel. +358 207 464 800, www.satron.com...
SATRON VO Installation and Setting-Up Instructions Manual 7.2 Changing the flushing parameters with the use of the VO/VC integrated display Ensure that the transmitter is powered on and all cables are connected, see Appendix. (For VO: press first “ESC” for 3 seconds to enter “SYSTCONF”) Select “I/O CONF”...
8 SI LOG ADVISOR ‑ QUICK GUIDE 8.1 Introduction SI Log Advisor is a tool for reading and analyzing log information stored in Satron devices. At this time event and data log feature is implemented in Satron VO sensors. The main functions of the software are: •...
Device type and firmware versions are shown on the Device line • Connect button is replaced with Disconnect button. If the device is not found, an error message is displayed. Satron Instruments Inc., FI-33960 Pirkkala, Finland Tel. +358 207 464 800, www.satron.com...
Page 43
1 s, 2 s, 5 s, 10 s, 30 s, 1 min, 2 min, 5 min, 10 min, 30 min, 1 hour, 2 hour, 6 hour 12 hour, 1 day Note: with device firmware versions O120516A (VO) and later, this is “Medium speed” storage interval.
Page 44
HART error (invalid HART message received) real-time-clock event Other other event Details… Opens Details page (device firmware versions O120516A (VO) and later only) Read Reads current log configuration (automatically performed with Connect) Write Writes modified log configuration to the device Note: changes take effect immediately (no restart required) Satron Instruments Inc., FI-33960 Pirkkala, Finland...
Page 45
The maximum number of logged signals is 24 • The maximum number of logged signals with “Fast” interval is 8 Estimated time Shows estimated data storage time with current log settings. Satron Instruments Inc., FI-33960 Pirkkala, Finland Tel. +358 207 464 800, www.satron.com...
Page 46
Read Reads either events or data, or both. Note: reading events/data using HART connection requires firmware version - O120327A or later (VO), or - V120308A or later (V1K) Events and Data Read either events or data, or both (at least one has to be selected).
Page 47
Open SI file (*.sif) opens previously stored log data file (*.sif ) Save SI file (*.sif) stores log data to file (*.sif = Satron Instruments file format) Export data to text file writes data to text file in CSV format (comma-separated values)
10 % (default), 5 %, 2 %, 1 %, 0.5 %, 0.2 %, 0.1 %, 0.05 %, 0.02 %, 0.01 % • With Fit to window button the timescale can be adjusted automatically so that all events/data fits to the window Satron Instruments Inc., FI-33960 Pirkkala, Finland Tel. +358 207 464 800, www.satron.com...
Page 49
When Track is enabled, the preview window and cursor are automatically moved to show the last log item (oscilloscope-like function). When Track is disabled, the user can freely move within the time space. Note: Online mode is available only with USB, not HART. Satron Instruments Inc., FI-33960 Pirkkala, Finland Tel. +358 207 464 800, www.satron.com...
Need help?
Do you have a question about the VO and is the answer not in the manual?