Test using Zinc Test Block 10.0 Long Term Storage 10.1 Laying Up 10.2 Re-commissioning Test Certificate Declaration of Conformity Buckleys (UVRAL) Ltd. Range Road, Hythe, Kent. United Kingdom. CT21 6HG. Tel:- +44 (0) 1303 260127/8 Fax:- +44 (0) 1303 262115...
This current model has been developed from the original model and on the very successful Bathycorrometer® MkV. The unit is fitted with a 3½ digit high impedance (10 megohms) LCD Volt Meter which is backlit for operation in murky conditions.
3.5 VA maximum Output voltage :- 10.5V D.C. open circuit. Output Current :- 30mA ± 2mA Charging indicator :- Green LED Buckleys (UVRAL) Ltd. Range Road, Hythe, Kent. United Kingdom. CT21 6HG. Tel:- +44 (0) 1303 260127/8 Fax:- +44 (0) 1303 262115...
The BCM is now ready for use. If required the calibration can be checked, see section on Calibration. Buckleys (UVRAL) Ltd. Range Road, Hythe, Kent. United Kingdom. CT21 6HG. Tel:- +44 (0) 1303 260127/8 Fax:- +44 (0) 1303 262115...
Electrode and the “O” Ring around the Tie Rod. Smear the “O” Ring and the thread on the new Reference Electrode with Silicone Grease. Buckleys (UVRAL) Ltd. Range Road, Hythe, Kent. United Kingdom. CT21 6HG. Tel:- +44 (0) 1303 260127/8 Fax:- +44 (0) 1303 262115...
Delrin Probe Cone onto the Tie Rod. Replace the stainless steel Probe Tip, if fitted. Safety All diving operations which involve the use of the Bathycorrometer must be carried out in accordance with the current local statutory requirements for conducting Diving Operations at Work.
Undoing the pressure housing voids the warranty. The unit should be returned to Buckleys if the reading on the BCM is more than ± 10mV out from the 1.990V test voltage. Check that it is not the BCM Checker that is at fault first.
A tarnished block will give a lower reading than clean one. Do not leave the test block in the Seawater, remove after use. Buckleys (UVRAL) Ltd. Range Road, Hythe, Kent. United Kingdom. CT21 6HG. Tel:- +44 (0) 1303 260127/8 Fax:- +44 (0) 1303 262115...
BCM and Charger should be repaired. Soak the BCM for 2 hours in a Seawater solution. Carry out calibration checks, see calibration. Buckleys (UVRAL) Ltd. Range Road, Hythe, Kent. United Kingdom. CT21 6HG. Tel:- +44 (0) 1303 260127/8 Fax:- +44 (0) 1303 262115...
Council Directive 73/23/EEC relating to safety requirements and Council Directive 89/336/EEC relating to electromagnetic compatibility. Issued on: ___________________ Authorised by : Cliff Moore Quality Manager Buckleys (UVRAL) Ltd. Range Road, Hythe, Kent. United Kingdom. CT21 6HG. Tel:- +44 (0) 1303 260127/8 Fax:- +44 (0) 1303 262115...