Service programs

Program 3 - Error log

Program 3
Shows an error list of the last 20 error messages.
The left section shows how many error messages were logged.
The error codes are shown in the right section.
The latest error message is shown first.
Use the mAs +/- keys to display the next error messages.

Program 4 - Deletion of the error log

Program 4
Keep the kV+ key pressed for approx. 4 seconds to delete the error list.
Program 5 - changing the preferred values for kV & mAs
Program 5
After the start of the service program, "LAS VAL" is shown on the display.
The "Last Value" function can be activated with the "Collimator Lamp" button.
The standard exposure parameters can be selected via the kV+/kV- and mAs+/mAs-
keys.
Use the Collimator Lamp button to save these values and to end the program.
Program 6 - Display of the maximum values (KV & mAs)
Program 6
The maximum exposure parameters can be selected/established via the kV+/kV- and
mAs+/mAs- keys.
Collimator lamp button to save these values and to end the program.
The maximum exposure parameters can be individually programmed for all operating
modes (line operation 110 V, line operation 230 V und battery operation (Hybrid)).
Switch the (Hybrid) operating mode switch to line operation or battery operation prior to
switching the system on.
For mains operation, you must ensure that the programmed values are valid only for the
given line voltage (110V or 230V) on site.