Labnet ProBlot 6 Instruction Manual

Hybridization ovens

Advertisement

Quick Links

Labnet ProBlot™ 6 and 12
Hybridization Ovens
Instruction Manual
Catalog Numbers:
H0600A
H600A-230V
H1200A
H1200A-230V

Advertisement

Table of Contents
loading

Summary of Contents for Labnet ProBlot 6

  • Page 1 Labnet ProBlot™ 6 and 12 Hybridization Ovens Instruction Manual Catalog Numbers: H0600A H600A-230V H1200A H1200A-230V...
  • Page 2: Table Of Contents

    6.0 Stacking Labnet ProBlot Hybridization Ovens ........
  • Page 3: Introduction

    The ProBlot 6 and 12 ovens include a drip pan for easy clean-up of accidental spills. Hybridization bottles made of high quality borosilicate glass, which provide a high level of protection against heat and thermal expansion are recommended for optimum ProBlot oven performance.
  • Page 4: Specifications

    6.0 Stacking Labnet ProBlot Hybridization Ovens The ProBlot 6 and 12 models may be stacked up to two high. You may wish to remove the adjustable feet from the upper unit when stacking. The adjustable feet can slide off the top of the lower unit and/or scratching can occur if rubber cups or matting are not used.
  • Page 5: Rotisserie

    After setting the temperature, allow at least one hour for the chamber temperature to stabilize and 24 hours for optimum stabilization. Calibration: The Labnet ProBlot hybridization oven is calibrated at 65°C at the factory. The unit can be recalibrated after the ...
  • Page 6: 10.0 Cleaning And Maintenance

    Should you have a question about the operation of the Labnet ProBlot™ hybridization oven or if service is required, contact Corning Customer Service. Do not send in a unit for service without first calling to obtain a repair authorization number. Should the unit require return for service, it should be properly packed to avoid damage.
  • Page 7: 13.0 Limited Warranty

    13.0 Limited Warranty Corning Incorporated (Corning) warrants that this product will be free from defects in material and workmanship for a period of one (1) year from date of purchase. CORNING DISCLAIMS ALL OTHER WARRANTIES WHETHER EXPRESSED OR IMPLIED, INCLUDING ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR OF FITNESS FOR A PARTICULAR PURPOSE.
  • Page 8 To request certificates, please contact us at www.labnetlink.com. Warranty/Disclaimer: Unless otherwise specified, all products are for research use or general laboratory use only.* Not intended for use in diagnostic or therapeutic procedures. Not for use in humans. These products are not intended to mitigate the presence of microorganisms on surfaces or in the environment, where such organisms can be deleterious to humans or the environment.

This manual is also suitable for:

Problot 12H0600aH600a-230vH1200aH1200a-230v

Table of Contents