D.O.R.C. MICRO DIATHERMY SYSTEM - Service Manual Section 1 GENERAL The Dutch Ophthalmic Research Center (D.O.R.C.) is an innovative company that is involved in research, development and manufacturing of high quality instruments and equipment for ophthalmic surgery. All D.O.R.C. products have been clinically tested before they were released to the ophthalmic surgeon.
D.O.R.C. MICRO DIATHERMY SYSTEM - Service Manual Section 2 SPECIFICATIONS GENERAL: Apparatus MICRO DIATHERMY SYSTEM Type Type 1100 Weight 5,5 Kg Dimensions 13 x 22,5 x 29 cm. MAINS INPUT: Mains supply 100-110 VAC (50/60 Hz) 230-240 VAC (50/60 Hz)
Page 5
D.O.R.C. MICRO DIATHERMY SYSTEM - Service Manual AMBIENT CONDITIONS: Operating temperature 15-40EC Humidity 85% Max. The D.O.R.C. Diathermy system complies with the safety standards as described in the international norm IEC601-1, Type BF, Class 1 3/17 30200300E...
D.O.R.C. MICRO DIATHERMY SYSTEM - Service Manual Section 3 WARNINGS A warning: Indicates a potentially harmful situation to yourself or others. Electric shock hazard The Diathermy system contains high voltage circuits. After performing any repair, calibration procedure performs a final electrical safety check and leakage current test.
D.O.R.C. MICRO DIATHERMY SYSTEM - Service Manual Section 4 CAUTIONS A caution: Indicates a condition that may result in equipment damage or malfunction. Electrostatic discharge through the printed circuit boards will damage the components of the Diathermy system. Handle all circuit boards (replacements and defective) by their non- conductive edges and use anti-static containers when transporting them.
D.O.R.C. MICRO DIATHERMY SYSTEM - Service Manual Section 5 SHORT INSTRUCTIONS NOMENCLATURE OF FRONT PANEL. 1. Power button. 2. Preset regulator. 3. RF Output connector. 4. RF Output indicator Fig. 2 Diathermy unit - Front Panel NOMENCLATURE OF REAR PANEL. 1.
Replacement parts. may be obtained from the: Dutch Ophthalmic Research Center International b.v. Scheijdelveweg 2 3214 VN Zuidland The Netherlands Phone : (+31) (0) 181 45 80 80 : (+31) (0) 181 45 80 90 E-mail : mailto@dorc.nl 12/17 30200300E...
Page 15
D.O.R.C. MICRO DIATHERMY SYSTEM - Service Manual Introduction and applicability of this manual. The service manual give the information required to maintain and repair the Micro Diathermy system . In order to understand this manual it is necessary to have a complete understanding of the functions of the Micro Diathermy system.
D.O.R.C. MICRO DIATHERMY SYSTEM - Service Manual Section 9 SERVICING AND ADJUSTMENT Recommended tools and test equipment. ITEM SPECIFICATION 1. Auto-Ranging Multimeter Fluke Model 77 or equivalent. 2. Scopemeter Fluke model 97 or equivalent. 3. Miniature Inspection Light 4. Inspection mirror 5.
Page 17
D.O.R.C. MICRO DIATHERMY SYSTEM - Service Manual After previous adjustments are performed, check the following: - Preset position “1" =>the RF output has a value of maximum 1.5 W If applicable, re-tune the RF output with P2. Repeat if necessary adjustment procedure for an optimum result. Fig.
D.O.R.C. MICRO DIATHERMY SYSTEM - Service Manual Section 10 SPARE PARTS MANUAL Fig. 10 Diathermy unit - Inside Pos. Description Partnr. Output connector VNR0322 Preset potmeter VNR0550 Main-switch VNR0092 Coax connector VNR0878 Diathermy Module Power transformer VNR0169 16/17 30200300E...
Page 19
D.O.R.C. MICRO DIATHERMY SYSTEM - Service Manual APPENDIX A After Service Test Procedure The following test should be performed after any service to ensure proper operation of the system. Check if the repaired module meets the specifications as described in the Service- or Instruction Manual.